共 50 条
Challenges in preparing soil samples and performing a reliable plutonium isotopic analysis by ICP-MS
被引:10
|作者:
Puzas, A.
[1
]
Genys, P.
[2
]
Remeikis, V.
[1
]
Druteikiene, R.
[1
]
机构:
[1] Ctr Phys Sci & Technol, LT-02300 Vilnius, Lithuania
[2] Vilnius State Univ, LT-10222 Vilnius, Lithuania
关键词:
Plutonium isotopes;
ICP-MS;
Uranium hydride;
Lead chloride;
Polyatomic interferences;
PLASMA-MASS SPECTROMETRY;
RATIO;
D O I:
10.1007/s10967-014-3411-8
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
Soil samples were prepared by performing ion exchange and extraction chromatography separation techniques and measured in the low resolution mode (m/Delta m = 300) using a standard nebulizer (1 mL/min, U-type, "Glass Expansion") and a cyclonic 50 mL spray chamber. With this configuration if plutonium concentration is in a sub-ppt range, (UH+)-U-238-H-1 must be taken into the account; however, taking into the account (UHH+)-U-238-H-1-H-1 is not necessary. In all cases it is strongly recommended to avoid using hydrochloric acid as a final stabilizing matrix and use the ultrapure nitric acid as it creates the lowest background at 239 and 240 m/z values.
引用
收藏
页码:751 / 759
页数:9
相关论文