Comprehensive Quality Control of the ITG 68Ge/68Ga Generator and Synthesis of 68Ga-DOTATOC and 68Ga-PSMA-HBED-CC for Clinical Imaging

被引:40
作者
Amor-Coarasa, Alejandro [1 ]
Sehoendorf, Megan [1 ]
Meckel, Marian [2 ]
Vallabhajosula, Shankar [1 ]
Babich, John W. [1 ]
机构
[1] Weill Cornell Med, Dept Radiol, Belfer Res Bldg,Rm 1600,413 East 69th St, New York, NY 10021 USA
[2] Isotope Technol Garching GmbH, Garching, Germany
关键词
Ga-68; generator; Ga-68-PSMA-HBED-CC; Ga-68-DOTATOC; quality control;
D O I
10.2967/jnumed.115.171249
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A good-manufacturing-practices (GMP) Ge-68/Ga-68 generator that uses modified dodecyl-3,4,5-trihydroxybenzoate hydrophobically bound to a octadecyl silica resin (C-18) as an adsorbent has been developed that allows for dilute HCl (0.05N) to efficiently elute metal-impurity-free Ga-68(3+) ready for peptide labeling. We characterized the performance of this generator system over a year in conjunction with the production of Ga-68-labeled DOTATOC and Glu-NH-CO-NH-Lys(Ahx)-HBED-CC (PSMA-HBED-CC) intended for clinical studies and established protocols for batch release. Methods: A 2,040-MBq self-shielded Ge-68/Ga-68 generator provided metal-free (GaCl3)-Ga-68 ready for peptide labeling in the fluidic labeling module after elution with 4 mL of 0.05N HCl. The compact system was readily housed in a laminar flow cabinet allowing an ISO class-5 environment. Ga-68 labeling of peptides using GMP kits was performed in 15-20 min, and the total production time was 45-50 min. Batch release quality control specifications were established to meet investigational new drug submission and institutional review board approval standards. Results: Over a period of 12 mo, Ga-68 elution yields from the generator averaged 80% (range, 72.0%-95.1%), and Ge-68 breakthrough was less than 0.006%, initially decreasing with time to 0.001% (expressed as pertentage of Ge-68 activity present in the generator at the time of elution), a unique characteristic of this generator. The radiochemical purity of both Ga-68-DOTATOC and Ga-68-PSMA-HBED-CC determined by high-performance liquid chromatography analysis was greater than 98%, with a minimum specific activity of 12.6 and 42 GBq/pmol, respectively. The radionuclidic (Ge-68) impurity was 0.00001% or less (under the detection limit). Final sterile, pyrogen-free formulation was provided in physiologic saline with 5%-7% ethanol. Conclusion: The GMP-certified Ge-68/Ga-68 generator system was studied for a year. The generator system is contained within the fluidic labeling module, and it is compact, self-shielded, and easy to operate using simple manual techniques. The system provides radiolabeled peptides with high (>98%) radiochemical purity and greater than 80% radiochemical yield. The Ge-68 levels in the final drug products were under the detection limits at all times. Ga-68-DOTATOC and Ga-68-PSMA-HBED-CC investigational radiopharmaceuticals are currently being studied clinically under investigational new drug (IND) applications submitted to the U.S. Food and Drug Administration.
引用
收藏
页码:1402 / 1405
页数:4
相关论文
共 14 条
[1]  
Alejandro Amor Coarasa Seza Gulec and Anthony McGoron, 2012, J NUCL MED S1, V53, P1742
[2]  
Amor-Coarasa Alejandro, 2014, Int J Mol Imaging, V2014, P269365, DOI 10.1155/2014/269365
[3]  
ANGER HO, 1964, J NUCL MED, V5, P515
[4]   Validation of 68Ge/68Ga generator processing by chemical purification for routine clinical application of 68Ga-DOTATOC [J].
Asti, Mattia ;
De Pietri, Giovanni ;
Fraternali, Alessandro ;
Grassi, Elisa ;
Sghedoni, Roberto ;
Floroni, Federica ;
Roesch, Frank ;
Versari, Annibale ;
Salvo, Diana .
NUCLEAR MEDICINE AND BIOLOGY, 2008, 35 (06) :721-724
[5]  
GLEASON G I, 1960, Int J Appl Radiat Isot, V8, P90, DOI 10.1016/0020-708X(60)90052-1
[6]   AN IMPROVED GALLIUM-68 COW [J].
GREENE, MW ;
TUCKER, WD .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1961, 12 (1-2) :62-&
[7]  
LOCH C, 1980, J NUCL MED, V21, P171
[8]  
Razbash AA, 2005, 5 INT C IS 5ICI APR
[9]  
Stang LG, 1958, RADIOISOTOPES SCI RE, P50
[10]   Characterization of SnO2-based 68Ge/68Ga generators and 68Ga-DOTATATE preparations: radionuclide purity, radiochemical yield and long-term constancy [J].
Sudbrock, Ferdinand ;
Fischer, Thomas ;
Zimmermanns, Beate ;
Guliyev, Mehrab ;
Dietlein, Markus ;
Drzezga, Alexander ;
Schomaecker, Klaus .