The Influence of a Positioning Hex on Abutment Rotation in Tapered Internal Implants: A 3D Finite Element Model Study

被引:5
作者
Hyun, Dong-Geun [1 ,2 ]
Kwon, Ho-Beom [1 ,2 ]
Lim, Young-Jun [1 ,2 ]
Koak, Jai-Young [1 ,2 ]
Kim, Myung-Joo [1 ,2 ]
机构
[1] Seoul Natl Univ, Sch Dent, Dept Prosthodont, 101 Daehak Ro, Seoul 110749, South Korea
[2] Seoul Natl Univ, Sch Dent, Dent Res Inst, Seoul, South Korea
关键词
3D finite element model; implant rotation; positioning hex; tapered internal implant; von Mises stress; STRESS-DISTRIBUTION; BONE TRABECULAE; CONNECTION; COMPLICATIONS; PROSTHESES; STABILITY; SYSTEMS; DESIGNS;
D O I
10.11607/jomi.7673
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The purpose of this study was to investigate the resistance against rotation of a two-piece abutment with or without the presence of a positioning hex in tapered internal connection implant systems. Materials and Methods: A tapered internal connection implant with an 11-degree internal oblique angle was placed into a constructed three-dimensional (3D) bone model. Two types of abutments were compared: hex-type abutments with a positioning hex and two-piece round-type nonhex abutments. Vertical compressions equivalent to the screw tightening torque of 30 Ncm and 40 Ncm were applied to the abutment screws. A horizontal load of 150 N was applied to the superstructure to create moments. The contact area between the implant and abutment was compared and analyzed. The rotational displacement of the abutments was investigated as well. Results: When a preload equivalent to 40 Ncm of tightening torque was applied, the round-type abutment exhibited a larger contact area than the hex-type abutment by approximately 2.5% to 3.5% in contrast to the internal surface area of the implant. When the screw tightening torque of 30 Ncm was applied, the rotational displacement of the hex-type abutment was approximately 0.54 degrees and the round-type was approximately 1.16 degrees under clockwise moment loading. Conclusion: Within the limitations of this study, it can be assumed that the hexagonal design of the abutment acts as rotational resistance against moments.
引用
收藏
页码:281 / 288
页数:8
相关论文
共 42 条
[1]   A 15-YEAR STUDY OF OSSEOINTEGRATED IMPLANTS IN THE TREATMENT OF THE EDENTULOUS JAW [J].
ADELL, R ;
LEKHOLM, U ;
ROCKLER, B ;
BRANEMARK, PI .
INTERNATIONAL JOURNAL OF ORAL SURGERY, 1981, 10 (06) :387-416
[2]   Evaluation of the mechanical characteristics of the implant-abutment complex of a reduced-diameter morse-taper implant -: A nonlinear finite element stress analysis [J].
Akça, K ;
Çehreli, MC ;
Iplikçioglu, H .
CLINICAL ORAL IMPLANTS RESEARCH, 2003, 14 (04) :444-454
[3]   Influence of occlusal forces on stress distribution in preloaded dental implant screws [J].
Alkan, I ;
Sertgöz, A ;
Ekici, E .
JOURNAL OF PROSTHETIC DENTISTRY, 2004, 91 (04) :319-325
[4]   A jaw model for the study of the mandibular flexure taking into account the anisotropy of the bone [J].
Alvarez-Arenal, Angel ;
Sanchez Lasheras, Fernando ;
Martin Fernandez, Elena ;
Gonzalez, Ignacio .
MATHEMATICAL AND COMPUTER MODELLING, 2009, 50 (5-6) :695-704
[5]  
Binon PP, 1996, INT J PROSTHODONT, V9, P149
[6]   Mechanical strength and fracture point of a dental implant under certification conditions: A numerical approach by finite element analysis [J].
Castolo, Guillermo de la Rosa ;
Guevara Perez, Sonia, V ;
Arnoux, Pierre-Jean ;
Badih, Laurent ;
Bonnet, Franck ;
Behr, Michel .
JOURNAL OF PROSTHETIC DENTISTRY, 2018, 119 (04) :611-619
[7]   Stress distribution of two commercial dental implant systems: A three-dimensional finite element analysis [J].
Chang, Hao-Sheng ;
Chen, Yi-Chin ;
Hsieh, Yao-Dung ;
Hsu, Ming-Lun .
JOURNAL OF DENTAL SCIENCES, 2013, 8 (03) :261-271
[8]   Three-Dimensional Finite Element Analysis of the Stress Distribution at the Internal Implant-Abutment Connection [J].
Cho, Sung-Yong ;
Huh, Yoon-Hyuk ;
Park, Chan-Jin ;
Cho, Lee-Ra .
INTERNATIONAL JOURNAL OF PERIODONTICS & RESTORATIVE DENTISTRY, 2016, 36 (03) :E49-E58
[9]   Influences of Internal Tapered Abutment Designs on Bone Stresses Around a Dental Implant: Three-Dimensional Finite Element Method With Statistical Evaluation [J].
Chu, Chun-Ming ;
Huang, Heng-Li ;
Hsu, Jui-Ting ;
Fuh, Lih-Jyh .
JOURNAL OF PERIODONTOLOGY, 2012, 83 (01) :111-118
[10]  
English C E, 1992, Implant Dent, V1, P273, DOI 10.1097/00008505-199200140-00009