Experimental Characterization of a Personal Wireless Sensor Network for the Medical X-Ray Dosimetry

被引:19
作者
Magalotti, Daniel [1 ,2 ]
Placidi, Pisana [2 ,3 ]
Dionigi, Marco [3 ]
Scorzoni, Andrea [2 ,3 ]
Servoli, Leonello [2 ]
机构
[1] Univ Modena & Reggio Emilia, I-41100 Modena, Italy
[2] Ist Nazl Fis Nucl, I-06123 Perugia, Italy
[3] Univ Perugia, I-06125 Perugia, Italy
关键词
Interventional radiology (IRad); portable instrumentation; radiation protection; wireless sensor node; X-ray dosimeter; INTERVENTIONAL RADIOLOGY; RADIATION-EXPOSURE; TESTS; STAFF;
D O I
10.1109/TIM.2016.2534661
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Wireless sensor networks are an important technology for large-scale monitoring, providing sensor measurements at high temporal and spatial resolution. In healthcare applications, a variety of system prototypes and commercial products have been designed and manufactured with the aim to provide an alternative and more efficient method for real-time operator monitoring during medical procedures. In the framework of the Real-Time Active Pixel Dosimetry project, the attention has been focused on dose monitoring of interventional radiology (IRad) operators. A sensor network has been developed, which consists of several personal sensor nodes (PSNs), monitoring the absorbed dose in different positions of the operator body (e.g., on the head or arms) and providing a measurement of the absorbed dose for each medical procedure. In this paper, we describe the first characterization of the final version of a system by measuring the radiation diagram and exposing the system to a direct X-ray beam showing a linearity of the response up to 3.3 mGy/s of dose rate. The system has been characterized during several IRad procedures when two PSNs have been simultaneously used in the network. The packet error rate of the network has been measured and the absorbed dose has been evaluated for each medical procedure, showing an uncertainty on the dose measurement lower than 10% with respect to the reference dosimetric devices.
引用
收藏
页码:2002 / 2011
页数:10
相关论文
共 26 条
[1]  
[Anonymous], 2003, Ann ICRP, V33, P1, DOI [10.1016/S0146-6453(03)00024-1, DOI 10.1016/S0146-6453(03)00024-1]
[2]   Dose Evaluation in Medical Staff during Interventional Radiology Procedures [J].
Bacchim Neto, F. A. ;
Alves, A. F. F. ;
Alvarez, M. ;
Rosa, M. E. D. ;
Macedo de Freitas, C. C. ;
Moura, R. ;
Pina, D. R. ;
Miranda, J. R. A. .
VI LATIN AMERICAN CONGRESS ON BIOMEDICAL ENGINEERING (CLAIB 2014), 2014, 49 :927-930
[3]   The Effect of Realtime Monitoring on Dose Exposure to Staff Within an Interventional Radiology Setting [J].
Baumann, Frederic ;
Katzen, Barry T. ;
Carelsen, Bart ;
Diehm, Nicolas ;
Benenati, James F. ;
Pena, Constantino S. .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2015, 38 (05) :1105-1111
[4]  
Butcher RL, 2015, J RADIOL NURS, V34, P137
[5]   Performance of several active personal dosemeters in interventional radiology and cardiology [J].
Chiriotti, S. ;
Ginjaume, M. ;
Vano, E. ;
Sanchez, R. ;
Fernandez, J. M. ;
Duch, M. A. ;
Sempau, J. .
RADIATION MEASUREMENTS, 2011, 46 (11) :1266-1270
[6]   Effect of a Real-Time Radiation Monitoring Device on Operator Radiation Exposure During Cardiac Catheterization The Radiation Reduction During Cardiac Catheterization Using Real-Time Monitoring Study [J].
Christopoulos, Georgios ;
Papayannis, Aristotelis C. ;
Alomar, Mohammed ;
Kotsia, Anna ;
Michael, Tesfaldet T. ;
Rangan, Bavana V. ;
Roesle, Michele ;
Shorrock, Deborah ;
Makke, Lorenza ;
Layne, Ronald ;
Grabarkewitz, Rebecca ;
Haagen, Donald ;
Maragkoudakis, Spyros ;
Mohammad, Atif ;
Sarode, Karan ;
Cipher, Daisha J. ;
Chambers, Charles E. ;
Banerjee, Subhash ;
Brilakis, Emmanouil S. .
CIRCULATION-CARDIOVASCULAR INTERVENTIONS, 2014, 7 (06) :744-750
[7]   ACTIVE PERSONAL DOSEMETERS IN INTERVENTIONAL RADIOLOGY: TESTS IN LABORATORY CONDITIONS AND IN HOSPITALS [J].
Clairand, I. ;
Bordy, J. -M. ;
Daures, J. ;
Debroas, J. ;
Denoziere, M. ;
Donadille, L. ;
Ginjaume, M. ;
Itie, C. ;
Koukorava, C. ;
Krim, S. ;
Lebacq, A. -L. ;
Martin, P. ;
Struelens, L. ;
Sans-Merce, M. ;
Tosic, M. ;
Vanhavere, F. .
RADIATION PROTECTION DOSIMETRY, 2011, 144 (1-4) :453-458
[8]  
Clairand I., USE ACTIVE PERSONAL
[9]   Use of a CMOS Image Sensor for an Active Personal Dosimeter in Interventional Radiology [J].
Conti, Elia ;
Placidi, Pisana ;
Biasini, Maurizio ;
Bissi, Lucia ;
Calandra, Andrea ;
Checcucci, Bruno ;
Chiocchini, Stefania ;
Cicioni, Roberto ;
Di Lorenzo, Roberto ;
Dipilato, Anna Concetta ;
Esposito, Alessandro ;
Paolucci, Massimiliano ;
Passeri, Daniele ;
Pentiricci, Andrea ;
Scorzoni, Andrea ;
Servoli, Leonello .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2013, 62 (05) :1065-1072
[10]  
Cosoli G, 2015, IEEE INT SYM MED MEA, P380, DOI 10.1109/MeMeA.2015.7145232