Design and realization of high resolution optical micro-scanning thermal microscope imaging system

被引:5
作者
Gao, MeiJing [1 ]
Tan, AiLing [1 ]
Zhang, BoZhi [1 ]
Wang, LiuZhu [1 ]
Li, ShiYu [1 ]
Chang, QiuYue [1 ]
Peng, ChunYang [1 ]
机构
[1] Yanshan Univ, Sch Informat Sci & Engn, Key Lab Special Fiber & Fiber Sensor Hebei Prov, Qinhuangdao 066004, Hebei, Peoples R China
关键词
Micro-scanning error; Optical micro-scanning thermal microscope imaging system; High resolution; Oversampling reconstruction; REGISTRATION;
D O I
10.1016/j.infrared.2018.10.013
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In order to meet the needs of subtle thermal analysis, We developed a thermal microscope imaging system based on uncooled focal plane detector in 2008. But for more sophisticated observations, the higher spatial resolution of thermal microscope imaging system is needed. Optical micro-scanning technology is one of the effective techniques to solve this problem. It can improve the systems spatial resolution based on existing imaging devices. Early we have developed an optical micro-scanning thermal microscope imaging system. But the system has a large micro-scanning error and requires high accuracy on the mechanical manufacturing and optical processing, which is not conducive to product. In order to reduce the micro-scanning error of developed optical microscanning thermal microscope imaging system and improve the spatial resolution of the system, a high precision micro-scanning component design method was proposed. The working principle of optical flat plate rotation micro-scanning is analyzed. The parameters and processing tolerances of micro-scanner are given. The integrated design and processing are realized with the existing uncooled thermal microscope imaging system. Results of standard oversample reconstruction for the real thermal microscope images have shown that the system is designed successfully and achieves higher resolution. It can be applied to high resolution microthermal analysis.
引用
收藏
页码:46 / 52
页数:7
相关论文
共 18 条
[1]   Image registration for a UV-Visible dual-band imaging system [J].
Chen, Tao ;
Yuan, Shuang ;
Li, Jianping ;
Xing, Sheng ;
Zhang, Honglong ;
Dong, Yuming ;
Chen, Liangpei ;
Liu, Peng ;
Jiao, Guohua .
OPTICS AND LASERS IN ENGINEERING, 2018, 105 :209-218
[2]  
Gao M.-J., 2007, SPIE, V6621, P1
[3]   High-resolution over-sampling reconstruction algorithm for a microscanning thermal microscope imaging system [J].
Gao, Meijing ;
Wang, Jingyuan ;
Xu, Wei ;
Guan, Congrong .
INFRARED PHYSICS & TECHNOLOGY, 2016, 76 :661-666
[4]  
Gao Meijing, 2009, Chinese Journal of Scientific Instrument, V30, P1037
[5]  
[高美静 GAO Meijing], 2007, [北京理工大学学报, Journal of beijing institute of technology], V27, P50
[6]  
Gao Meijing, 2009, T BEIJING I TECHNOL, V29, P705
[7]   ALIASING REDUCTION IN STARING INFRARED IMAGERS UTILIZING SUBPIXEL TECHNIQUES [J].
GILLETT, JC ;
STADTMILLER, TM ;
HARDIE, RC .
OPTICAL ENGINEERING, 1995, 34 (11) :3130-3137
[8]   Thermal microdevices for biological and biomedical applications [J].
Jain, Ankur ;
Goodson, Kenneth E. .
JOURNAL OF THERMAL BIOLOGY, 2011, 36 (04) :209-218
[9]   Comparison of image registration methods for composing spectral retinal images [J].
Laaksonen, Lauri ;
Claridge, Ela ;
Falt, Pauli ;
Hauta-Kasari, Markku ;
Uusitalo, Hannu ;
Lensu, Lasse .
BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2017, 36 :234-245
[10]  
Morikawa J., 2011, P SOC PHOTO-OPT INS, V8013, P1