MICRO ELECTRICAL IMPEDANCE SPECTROSCOPY (μEIS) WITH CELL TRAPS IN VARIOUS SIZE FOR DIFFERENTIATION BETWEEN NORMAL AND CANCEROUS HUMAN UROTHELIAL CELLS

被引:0
作者
Park, Y. [1 ]
Kim, H. [1 ]
Yun, J. [1 ]
Kang, G. [1 ]
Seo, S. [2 ]
Park, C. [1 ]
Yang, J. [1 ]
Chung, E. [1 ,2 ]
Lee, J. -H. [1 ,2 ]
机构
[1] GIST, Dept Med Syst Engn, Gwangju, South Korea
[2] GIST, Sch Mechatron, Gwangju, South Korea
来源
2015 TRANSDUCERS - 2015 18TH INTERNATIONAL CONFERENCE ON SOLID-STATE SENSORS, ACTUATORS AND MICROSYSTEMS (TRANSDUCERS) | 2015年
关键词
Electrical impedance; photolithography; urothelial cell; sensing electrodes; slanted sidewalls; BLADDER-CANCER; DIAGNOSIS;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper describes a novel micro electrical impedance spectroscopy (mu EIS) device with cell traps, which can discriminate between normal and cancerous human urothelial cells. Three-dimensional traps with electrodes were employed for effective capture of target cells. This allows cells to be automatically captured, while the electrodes on the slanted area of the trap accurately characterize the electrical impedance of the cells. The device is equipped with five traps of various dimensions in a single device, which can capture various types of cells regardless of their size and deformability. The mu EIS device accurately distinguished cancerous human urothelial cells (TCCSUP) from normal cells (SV-HUC-1); the average differences in the phase angle and resistance between the target cells were 1.8 degrees at 119 kHz and 416.1 Omega at 109 kHz, respectively.
引用
收藏
页码:355 / 358
页数:4
相关论文
共 16 条
  • [1] Aa M. V., 2008, BRIT J UROLOGY INT, V101, P1106
  • [3] The diagnosis and staging of bladder cancer: From RBCs to TURs
    Carmack, AJK
    Soloway, MS
    [J]. UROLOGY, 2006, 67 (3A) : 3 - 8
  • [4] Classification of cell types using a microfluidic device for mechanical and electrical measurement on single cells
    Chen, Jian
    Zheng, Yi
    Tan, Qingyuan
    Shojaei-Baghini, Ehsan
    Zhang, Yan Liang
    Li, Jason
    Prasad, Preethy
    You, Lidan
    Wu, Xiao Yu
    Sun, Yu
    [J]. LAB ON A CHIP, 2011, 11 (18) : 3174 - 3181
  • [5] Debanjan D., 2014, ROYAL SOC CHEM, V4, P18178
  • [6] Clinically relevant reduction in risk of recurrence of superficial bladder cancer using 5-aminolevulinic acid-induced fluorescence diagnosis: 8-year results of prospective randomized study
    Denzinger, Stefan
    Burger, Maximilian
    Walter, Bernhard
    Knuechel, Ruth
    Roessler, Wolfgang
    Wieland, Wolf F.
    Filbeck, Thomas
    [J]. UROLOGY, 2007, 69 (04) : 675 - 679
  • [7] Electric impedance microflow cytometry for characterization of cell disease states
    Du, E.
    Ha, Sungjae
    Diez-Silva, Monica
    Dao, Ming
    Suresh, Subra
    Chandrakasan, Anantha P.
    [J]. LAB ON A CHIP, 2013, 13 (19) : 3903 - 3909
  • [8] Jemal A, 2011, CA-CANCER J CLIN, V61, P69
  • [9] Discrimination between the human prostate normal cell and cancer cell by using a novel electrical impedance spectroscopy controlling the cross-sectional area of a microfluidic channel
    Kang, Giseok
    Kim, Young-jae
    Moon, Hong-sang
    Lee, Jeong-Woo
    Yoo, Tag-Keun
    Park, Kwangsung
    Lee, Jong-Hyun
    [J]. BIOMICROFLUIDICS, 2013, 7 (04):
  • [10] Differentiation Between Normal and Cancerous Cells at the Single Cell Level Using 3-D Electrode Electrical Impedance Spectroscopy
    Kang, Giseok
    Yoo, Sung Keun
    Kim, Hyoung-Ihl
    Lee, Jong-Hyun
    [J]. IEEE SENSORS JOURNAL, 2012, 12 (05) : 1084 - 1089