Retrospective evaluation of pediatric cranio-spinal axis irradiation plans with the Hi-ART tomotherapy system

被引:19
作者
Penagaricano, Jose A. [1 ]
Yan, Yulong [1 ]
Corry, Peter [1 ]
Moros, Eduardo [1 ]
Ratanatharathorn, Vaneerat [1 ]
机构
[1] Univ Arkansas Med Sci Hosp, Dept Radiat Oncol, Little Rock, AR 72205 USA
关键词
cranio-spinal irradiation; helical tornotherapy; beam-on time;
D O I
10.1177/153303460700600413
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Helical tornotherapy (HT) can be used for the delivery of cranio-spinal axis irradiation (CSAI) without the need for beam matching of conventional linac-based external beam irradiation. The aim of this study is to retrospectively evaluate HT plans used for treatment in nine patients treated with CSAI. Helical tornotherapy cranio-spinal axis irradiation (HT-CSAI) plans were created for each patient. Average length along the cranio-spinal axis of the PTV was 65.6 cm with a range between 53 and 74 cm. Treatment planning optimization and plan evaluation parameters were obtained from the HT planning station for each of the nine patients. PTV coverage by the 95% isodose surface ranged between 98.0 to 100.0% for all nine patients. The clinically acceptable dose variation within the PTV or tolerance range was between 0.7 and 2.5% for all nine patients. Doses to the organs at risk were clinically acceptable. An increasing length along the longitudinal axis of the PTV did not consistently increase the beam-on time indicating that using a larger jaw width had a greater impact on treatment time. With a larger jaw width it is possible to substantially reduce the normalized beam-on treatment time without compromising plan quality and sparing of organs at risk. By using a larger jaw width or lower modulation factor or both, normalized beam-on times were decreased by up to 61% as compared to the other evaluated treatment plans. From the nine cases reported in this study the minimum beam-on time was achieved with a jaw width of 5.0 cm, pitch of 0.287 and a modulation factor of 2.0. Large and long cylindrical volumes can be effectively treated with helical tornotherapy with both clinically acceptable dose distribution and beam-on time.
引用
收藏
页码:355 / 360
页数:6
相关论文
共 20 条
[1]   Specification of dose delivery in radiation therapy - Recommendations by the Nordic Association of Clinical Physics (NACP) [J].
Aaltonen, P ;
Brahme, A ;
Lax, I ;
Levernes, S ;
Naslund, I ;
Reitan, JB ;
Turesson, I .
ACTA ONCOLOGICA, 1997, 36 :1-32
[2]   Integral radiation dose to normal structures with conformal external beam radiation:: In regards to Aoyama et al. (Int J Radiat Oncol Biol Phys 2006;64:962-967) -: In response to Dr. Penagaricano [J].
Aoyama, Hidefumi ;
Mackie, Thomas R. ;
Bentzen, Soren M. ;
Olivera, Gustavo H. ;
Westerly, David C. ;
Mehta, Minesh P. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 65 (04) :1274-1275
[3]   Helical tomotherapy for craniospinal radiation [J].
Bauman, G ;
Yartsev, S ;
Coad, T ;
Fisher, B ;
Kron, T .
BRITISH JOURNAL OF RADIOLOGY, 2005, 78 (930) :548-552
[4]   Optimization of helical tomotherapy treatment plans for prostate cancer [J].
Grigorov, G ;
Kron, T ;
Wong, E ;
Chen, J ;
Sollazzo, J ;
Rodrigues, G .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (13) :1933-1943
[5]   Radiation-induced second cancers: The impact of 3D-CRT and IMRT [J].
Hall, EJ ;
Wuu, CS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (01) :83-88
[6]   Feasibility study of helical tomotherapy for total body or total marrow irradiation [J].
Hui, SK ;
Kapatoes, J ;
Fowler, J ;
Henderson, D ;
Olivera, G ;
Manon, RR ;
Gerbi, B ;
Mackie, TR ;
Welsh, JS .
MEDICAL PHYSICS, 2005, 32 (10) :3214-3224
[7]   A feasible method for clinical delivery verification and dose reconstruction in tomotherapy [J].
Kapatoes, JM ;
Olivera, GH ;
Ruchala, KJ ;
Smilowitz, JB ;
Reckwerdt, PJ ;
Mackie, TR .
MEDICAL PHYSICS, 2001, 28 (04) :528-542
[8]   Delivery verification in sequential and helical tomotherapy [J].
Kapatoes, JM ;
Olivera, GH ;
Reckwerdt, PJ ;
Fitchard, EE ;
Schloesser, EA ;
Mackie, TR .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (07) :1815-1841
[9]   The helical tomotherapy thread effect [J].
Kissick, MW ;
Fenwick, J ;
James, JA ;
Jerai, R ;
Kapatoes, JM ;
Keller, H ;
Mackie, TR ;
Olivera, G ;
Soisson, ET .
MEDICAL PHYSICS, 2005, 32 (05) :1414-1423
[10]  
Low D A, 2001, Med Dosim, V26, P79, DOI 10.1016/S0958-3947(00)00064-9