Validation of the dosimetric and geometric accuracy of MR-only treatment planning solution for prostate cancer radiotherapy

被引:2
|
作者
Posiewnik, Michal [1 ]
Piotrowski, Tomasz [2 ,3 ]
机构
[1] Gdynia Oncol Ctr, Dept Med Phys, Gdynia, Poland
[2] Poznan Univ Med Sci, Dept Electroradiol, 15 Garbary St, PL-61866 Poznan, Poland
[3] Greater Poland Canc Ctr, Dept Med Phys, Poznan, Poland
来源
WSPOLCZESNA ONKOLOGIA-CONTEMPORARY ONCOLOGY | 2021年 / 25卷 / 04期
关键词
MR-only planning; prostate cancer; fiducial registration error; synthetic CT; MAGNETIC-RESONANCE; RADIATION-THERAPY; DOSE CALCULATIONS; CT; WORKFLOW;
D O I
10.5114/wo.2021.112518
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Introduction: The aim was to validate the dosimetric and geometric accuracy of radiotherapy treatment plans for prostate cancer based on magnetic resonance (MR) imaging only and a solution based on computed tomography (CT) supported by MR imaging. Material and methods: We used CT and MR images of ten prostate cancer patients implanted with three fiducial markers (FM) in the prostate gland. Rigid registration based on FM was performed to assess the fusion accuracy between MR and CT images. The differences between prostate contours (clinical target volume - CTV) on CT (CTV CT) and MR (CTV MR) images were scored using the Dice similarity coefficient and directly comparing the outlined volumes. The volumetric modulated arc therapy plans were designed and optimised on synthetic CT (sCT) to obtain the dose distribution for the MR-only solution. In the next step, the sCT images were replaced by conventional CT images and the plans were recalculated. The doses obtained on sCT and CT were compared by direct dose subtraction and the gamma method. Results: The averaged fiducial registration error was equal to 0.5 mm. All CTV CT volumes were significantly bigger than corresponding CTV delineated on MR images (p = 0.005). The direct dose comparison shows that for 97.1% of patients' bodies, the differences were smaller than 0.1%. The average gamma passing rates were higher than 0.970. Conclusions: MR imaging allows for a more precise delineation of the prostate compared to CT imaging. The workflow of plan preparation based on MR and CT is burdened with an FM registration error that is eliminated by an MR-only solution with no compromise on dose distribution.
引用
收藏
页码:241 / 245
页数:5
相关论文
共 50 条
  • [21] Investigating the generalisation of an atlas-based synthetic-CT algorithm to another centre and MR scanner for prostate MR-only radiotherapy
    Wyatt, Jonathan J.
    Dowling, Jason A.
    Kelly, Charles G.
    McKenna, Jill
    Johnstone, Emily
    Speight, Richard
    Henry, Ann
    Greer, Peter B.
    McCallum, Hazel M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (24): : N548 - N560
  • [22] Are cone beam CT image matching skills transferrable from planning CT to planning MRI for MR-only prostate radiotherapy?
    Brooks, Rachel L.
    McCallum, Hazel M.
    Pearson, Rachel A.
    Pilling, Karen
    Wyatt, Jonathan
    BRITISH JOURNAL OF RADIOLOGY, 2021, 94 (1123):
  • [23] Dosimetric evaluation of an atlas-based synthetic CT generation approach for MR-only radiotherapy of pelvis anatomy
    Farjam, Reza
    Tyag, Neelam
    Deasy, Joseph O.
    Hun, Margie A.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (01): : 101 - 109
  • [24] Brain and Head-and-Neck MRI in Immobilization Mask: A Practical Solution for MR-Only Radiotherapy
    Mandija, Stefano
    D'Agata, Federico
    Navest, Robin J. M.
    Sbrizzi, Alessandro
    Tijssen, Rob H. N.
    Philippens, Marielle E. P.
    Raaijmakers, Cornelis P. J.
    Seravalli, Enrica
    Verhoeff, Joost J. C.
    Lagendijk, Jan J. W.
    van den Berg, Cornelis A. T.
    FRONTIERS IN ONCOLOGY, 2019, 9
  • [25] Assessment of dosimetric impact of system specific geometric distortion in an MRI only based radiotherapy workflow for prostate
    Gustafsson, C.
    Nordstrom, F.
    Persson, E.
    Brynolfsson, J.
    Olsson, L. E.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (08): : 2976 - 2989
  • [26] Comparison of four synthetic CT generators for brain and prostate MR-only workflow in radiotherapy
    Autret, Damien
    Guillerminet, Camille
    Roussel, Alban
    Cossec-Kerloc'h, Erwan
    Dufreneix, Stephane
    RADIATION ONCOLOGY, 2023, 18 (01)
  • [27] Dosimetric evaluation of synthetic CT image generated using a neural network for MR-only brain radiotherapy
    Tang, Bin
    Wu, Fan
    Fu, Yuchuan
    Wang, Xianliang
    Wang, Pei
    Orlandini, Lucia Clara
    Li, Jie
    Hou, Qing
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (03): : 55 - 62
  • [28] Comparison of four synthetic CT generators for brain and prostate MR-only workflow in radiotherapy
    Damien Autret
    Camille Guillerminet
    Alban Roussel
    Erwan Cossec-Kerloc’h
    Stéphane Dufreneix
    Radiation Oncology, 18 (1)
  • [29] Dosimetric Accuracy of MR-Guided Online Adaptive Planning for Nasopharyngeal Carcinoma Radiotherapy on 1.5 T MR-Linac
    Ding, Shouliang
    Liu, Hongdong
    Li, Yongbao
    Wang, Bin
    Li, Rui
    Huang, Xiaoyan
    FRONTIERS IN ONCOLOGY, 2022, 12
  • [30] Optimisation and validation of an integrated magnetic resonance imaging-only radiotherapy planning solution
    O'Connor, Laura M.
    Skehan, Kate
    Choi, Jae H.
    Simpson, John
    Martin, Jarad
    Warren-Forward, Helen
    Dowling, Jason
    Greer, Peter
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2021, 20 : 34 - 39