Novel detachable stents for the treatment of benign esophageal strictures

被引:5
作者
Shang, Liang [1 ]
Pei, Qing-Shan [2 ]
Xu, Dan [3 ]
Liu, Ji-Yong [2 ,4 ]
Liu, Jin [2 ,4 ]
机构
[1] Shandong Univ, Shandong Prov Hosp, Dept Gastrointestinal Surg, Jinan 250021, Shandong, Peoples R China
[2] Shandong Univ, Shandong Prov Hosp, Dept Gastroenterol, 324 Jingwu Rd, Jinan 250021, Shandong, Peoples R China
[3] Shandong Univ, Shandong Prov Hosp, Endoscopy Room, Jinan 250021, Shandong, Peoples R China
[4] Shandong Univ, Shandong Prov Hosp, Lab Translat Med, Jinan 250021, Shandong, Peoples R China
关键词
esophageal stricture; stents; endoscopic procedure; animal experimentation; BIODEGRADABLE STENTS; ENDOSCOPIC TREATMENT; METAL STENTS; PLACEMENT; PREVENTION; MANAGEMENT; STENOSIS; REMOVAL;
D O I
10.3892/etm.2019.8190
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
The radial force of esophageal stents may not completely change during extraction and therefore, the procedure of stent removal may cause tissue damage. The present study reports the manufacture of 2 novel detachable stents, which were designed to reduce tissue damage through their capacity to be taken or fall apart prior to removal and evaluated the supporting properties of these stents and the extent of local mucosal injury during their removal. The stents were manufactured by braiding, heat-setting, coating and connecting. The properties of the stents were evaluated by determining the following parameters: Expansion point, softening point, stent flexibility, radial compression ratio and radial force. A total of 18 rabbits with induced esophageal stricture were randomly assigned to 3 groups as follows: Detachable stent (DS) group, biodegradable stent (BS) group and control group. The stricture rate, complications, survival, degradation and stent removal were observed over 8 weeks. The stents of the DS and BS groups provided a similar supporting effect. The stricture rate, incidence of complications and survival were also similar between the 2 groups, while significant differences were noted between the DS and control groups and between the BS and control groups. In the BS group, the stents were degraded and moved to the stomach within 7 weeks (2 in 6 weeks and 3 in 7 weeks). The debris was extracted using biopsy forceps. In the DS group, all stents were easy to remove and 2 cases exhibited minor hemorrhage. In conclusion, the 2 types of novel detachable stent provided an equally efficient supporting effect in vitro and in vivo and may reduce the incidence of secondary injury during stent removal.
引用
收藏
页码:115 / 122
页数:8
相关论文
共 32 条
[1]   Management of persistent gastroesophageal anastomotic strictures with removable self-expandable polyester silicon-covered (Polyflex) stents: an alternative to serial dilation [J].
Barthel, James S. ;
Kelley, Scott T. ;
Mapman, Jason B. .
GASTROINTESTINAL ENDOSCOPY, 2008, 67 (03) :546-552
[2]   Removable Self-Expanding Plastic Esophageal Stent as a Continuous, Non-Permanent Dilator in Treating Refractory Benign Esophageal Strictures: A Prospective Two-Center Study [J].
Dua, Kulwinder S. ;
Vleggaar, Frank P. ;
Santharam, Rajesh ;
Siersema, Peter D. .
AMERICAN JOURNAL OF GASTROENTEROLOGY, 2008, 103 (12) :2988-2994
[3]   Tissue hyperplasia following placement of a biodegradable stent for a refractory esophageal stricture: treatment with argon plasma coagulation [J].
Dumoulin, F. L. ;
Plassmann, D. .
ENDOSCOPY, 2012, 44 :E356-E357
[4]   Treatment of esophageal leaks, fistulae, and perforations with temporary stents: evaluation of efficacy, adverse events, and factors associated with successful outcomes [J].
El Hajj, Ihab I. ;
Imperiale, Thomas F. ;
Rex, Douglas K. ;
Ballard, Darren ;
Kesler, Kenneth A. ;
Birdas, Thomas J. ;
Fatima, Hala ;
Kessler, William R. ;
DeWitt, John M. .
GASTROINTESTINAL ENDOSCOPY, 2014, 79 (04) :589-598
[5]   Novel Removable Internally Fully Covered Self-Expanding Metal Esophageal Stent: Feasibility, Technique of Removal, and Tissue Response in Humans [J].
Eloubeidi, Mohamad A. ;
Lopes, Tercio L. .
AMERICAN JOURNAL OF GASTROENTEROLOGY, 2009, 104 (06) :1374-1381
[6]   Severe hyperplastic tissue stenosis of a novel biodegradable esophageal stent and subsequent successful management with high-pressure balloon dilation [J].
Hair, C. S. ;
Devonshire, D. A. .
ENDOSCOPY, 2010, 42 :E132-E133
[7]   Traversing Difficult Esophageal Strictures from the Retrograde Approach [J].
Hasan, Muhammad ;
Maple, John T. .
TECHNIQUES IN GASTROINTESTINAL ENDOSCOPY, 2008, 10 (04) :149-154
[8]   Stent-in-Stent Technique for Removal of Embedded Esophageal Self-Expanding Metal Stents [J].
Hirdes, M. M. C. ;
Siersema, P. D. ;
Houben, M. H. M. G. ;
Weusten, B. L. A. M. ;
Vleggaar, F. P. .
AMERICAN JOURNAL OF GASTROENTEROLOGY, 2011, 106 (02) :286-293
[9]   Management of esophageal stricture after complete circular endoscopic submucosal dissection for superficial esophageal squamous cell carcinoma [J].
Isomoto, Hajime ;
Yamaguchi, Naoyuki ;
Nakayama, Toshiyuki ;
Hayashi, Tomayoshi ;
Nishiyama, Hitoshi ;
Ohnita, Ken ;
Takeshima, Fuminao ;
Shikuwa, Saburo ;
Kohno, Shigeru ;
Nakao, Kazuhiko .
BMC GASTROENTEROLOGY, 2011, 11
[10]   Temporary metallic stent placement in the treatment of refractory benign esophageal strictures: results and factors associated with outcome in 55 patients [J].
Kim, Jin Hyoung ;
Song, Ho-Young ;
Choi, Eugene K. ;
Kim, Kyung Rae ;
Shin, Ji Hoon ;
Lim, Jin-Oh .
EUROPEAN RADIOLOGY, 2009, 19 (02) :384-390