Development of an In Vitro Ocular Platform to Test Contact Lenses

被引:14
作者
Phan, Chau-Minh [1 ]
Walther, Hendrik [1 ]
Gao, Huayi [2 ]
Rossy, Jordan [1 ]
Subbaraman, Lakshman N. [1 ]
Jones, Lyndon [1 ]
机构
[1] Univ Waterloo, Sch Optometry & Vis Sci, Waterloo, ON N2L 3G1, Canada
[2] Medella Hlth, Kitchener, ON, Canada
来源
JOVE-JOURNAL OF VISUALIZED EXPERIMENTS | 2016年 / 110期
基金
加拿大自然科学与工程研究理事会;
关键词
Bioengineering; Issue; 110; eye model; ocular; in vitro; contact lenses; tear deposition; drug delivery; agar model; ophthalmic materials; ex vivo cornea; SILICONE-HYDROGEL; TEAR FLOW; LIPID DEPOSITION; RELEASE; DEXAMETHASONE; LYSOZYME; IMPACT;
D O I
10.3791/53907
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Currently, in vitro evaluations of contact lenses (CLs) for drug delivery are typically performed in large volume vials,(1-6) which fail to mimic physiological tear volumes.(7) The traditional model also lacks the natural tear flow component and the blinking reflex, both of which are defining factors of the ocular environment. The development of a novel model is described in this study, which consists of a unique 2-piece design, eyeball and eyelid piece, capable of mimicking physiological tear volume. The models are created from 3-D printed molds (Polytetrafluoroethylene or Teflon molds), which can be used to generate eye models from various polymers, such as polydimethylsiloxane (PDMS) and agar. Further modifications to the eye pieces, such as the integration of an explanted human or animal cornea or human corneal construct, will permit for more complex in vitro ocular studies. A commercial microfluidic syringe pump is integrated with the platform to emulate physiological tear secretion. Air exposure and mechanical wear are achieved using two mechanical actuators, of which one moves the eyelid piece laterally, and the other moves the eyeballeyepiece circularly. The model has been used to evaluate CLs for drug delivery and deposition of tear components on CLs.
引用
收藏
页数:7
相关论文
共 27 条
[1]   Zero-order therapeutic release from imprinted hydrogel contact lenses within in vitro physiological ocular tear flow [J].
Ali, Maryarn ;
Horikawa, Shin ;
Venkatesh, Siddarth ;
Saha, Jishnu ;
Hong, Jong Wook ;
Byrne, Mark E. .
JOURNAL OF CONTROLLED RELEASE, 2007, 124 (03) :154-162
[2]  
Bajgrowicz M., 2015, INVEST OPHTHALMOL VI
[3]   Uptake and Release of Dexamethasone Phosphate From Silicone Hydrogel and Group I, II, and IV Hydrogel Contact Lenses [J].
Boone, Adrienne ;
Hui, Alex ;
Jones, Lyndon .
EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2009, 35 (05) :260-267
[4]   In vitro drug release of natamycin from β-cyclodextrin and 2-hydroxypropyl-β-cyclodextrin-functionalized contact lens materials [J].
Chau-Minh Phan ;
Subbaraman, Lakshman N. ;
Jones, Lyndon .
JOURNAL OF BIOMATERIALS SCIENCE-POLYMER EDITION, 2014, 25 (17) :1907-1919
[5]  
Coster D., 2002, CORNEA
[6]  
Faschinger C, 2010, OPHTHALMOLOGE, V107, P918, DOI 10.1007/s00347-010-2198-4
[7]  
FURUKAWA RE, 1978, AM J OPTOM PHYS OPT, V55, P69
[8]   Extraction versus In Situ Techniques for Measuring Surface-Adsorbed Lysozyme [J].
Hall, Brad ;
Phan, Chau-Minh ;
Subbaraman, Lakshman ;
Jones, Lyndon W. ;
Forrest, James .
OPTOMETRY AND VISION SCIENCE, 2014, 91 (09) :1062-1070
[9]   In Vitro and In Vivo Evaluation of Novel Ciprofloxacin-Releasing Silicone Hydrogel Contact Lenses [J].
Hui, Alex ;
Willcox, Mark ;
Jones, Lyndon .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (08) :4896-4904
[10]   Adjusting biomaterial composition to achieve controlled multiple-day release of dexamethasone from an extended-wear silicone hydrogel contact lens [J].
Kaczmarek, James C. ;
Tieppo, Arianna ;
White, Charles J. ;
Byrne, Mark E. .
JOURNAL OF BIOMATERIALS SCIENCE-POLYMER EDITION, 2014, 25 (01) :88-100