Successful percutaneous treatment of late outflow graft failure of the left ventricular assist device: a long-term follow-up

被引:3
作者
Zajdel, Wojciech [1 ]
Tomala, Marek [1 ]
Bryndza, Magdalena [1 ,2 ]
Krupinski, Maciej [3 ]
Kapelak, Boguslaw [2 ,4 ]
Legutko, Jacek [4 ,5 ]
Wierzbicki, Karol [2 ,4 ]
机构
[1] John Paul 2 Hosp, Dept Invas Cardiol, Krakow, Poland
[2] Jagiellonian Univ Med Coll, Inst Cardiol, Dept Cardiovasc Surg & Transplantol, Krakow, Poland
[3] John Paul 2 Hosp, Dept Radiol, Krakow, Poland
[4] John Paul 2 Hosp, Krakow, Poland
[5] Jagiellonian Univ Med Coll, Inst Cardiol, Dept Invas Cardiol, Krakow, Poland
来源
ESC HEART FAILURE | 2021年 / 8卷 / 06期
关键词
Ventricular assist device; VAD; LVAD; Outflow graft failure; DIAGNOSIS;
D O I
10.1002/ehf2.13566
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Continuous flow left ventricular assist device (LVAD) outflow graft stenosis constitutes a severe complication. Treatment options include surgical pump exchange, transcatheter procedures, or systemic thrombolysis. We present a case of a spontaneous mechanical twisting of the outflow graft at two distinct points, which was treated by a two-step percutaneous stent implantation. Self-expanding stents were used during the first procedure. We also ensured distal bilateral percutaneous neuroprotection against cerebrovascular embolism. During the second, previously unpredicted procedure, we used balloon-expandable bare-metal stents to overcome the torque of the graft, because of their higher radial force. It was assumed that the external, self-expanding layer of the stent might protect the graft from the bare-metal stents. The effects of the treatment were monitored both clinically and through computed tomography angiography. The check-up 12 months later revealed nothing of note. Interventional transcatheter procedures are a safe treatment option for outflow graft stenosis, with good long-term effects. Both self-expanding and balloon-expandable stents can be used for transcatheter intervention. Both the incidence of the graft twisting and the radial force of the implanted stent seem to be critical and robust predictors of the long-term result of percutaneous therapy.
引用
收藏
页码:5555 / 5559
页数:5
相关论文
共 8 条
  • [1] Long-Term Survival in Patients Receiving a Continuous-Flow Left Ventricular Assist Device
    Gosev, Igor
    Kiernan, Michael S.
    Eckman, Peter
    Soleimani, Behzad
    Kilic, Ahmet
    Uriel, Nir
    Rich, Johnatan D.
    Katz, Jason N.
    Cowger, Jennifer
    Lima, Brian
    McGurk, Siobhan
    Brisco-Bacik, Meredith A.
    Lee, Sanjin
    Joseph, Susan M.
    Patel, Chetan B.
    [J]. ANNALS OF THORACIC SURGERY, 2018, 105 (03) : 696 - 701
  • [2] A report on interventional acute bare-metal stenting of the HeartMate 3 left ventricular assist device twisted outflow graft
    Juraszek, Andrzej
    Szymanski, Jaroslaw
    Demkow, Marcin
    Kusmierczyk, Mariusz
    [J]. KARDIOLOGIA POLSKA, 2020, 78 (10) : 1049 - 1050
  • [3] A tale of the twist in the outflow graft: An analysis from the MOMENTUM 3 trial
    Mehra, Mandeep R.
    Salerno, Christopher
    Naka, Yoshifumi
    Uriel, Nir
    Cleveland, Joseph C., Jr.
    Horstmanshof, Douglas
    Goldstein, Daniel J.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2018, 37 (11) : 1281 - 1284
  • [4] 2016 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure
    Ponikowski, Piotr
    Voors, Adriaan A.
    Anker, Stefan D.
    Bueno, Hector
    Cleland, John G. F.
    Coats, Andrew J. S.
    Falk, Volkmar
    Ramon Gonzalez-Juanatey, Jose
    Harjola, Veli-Pekka
    Jankowska, Ewa A.
    Jessup, Mariell
    Linde, Cecilia
    Nihoyannopoulos, Petros
    Parissis, John T.
    Pieske, Burkert
    Riley, Jillian P.
    Rosano, Giuseppe M. C.
    Ruilope, Luis M.
    Ruschitzka, Frank
    Rutten, Frans H.
    van der Meer, Peter
    [J]. EUROPEAN HEART JOURNAL, 2016, 37 (27) : 2129 - U130
  • [5] Strategy for surgical correction and mitigation of outflow graft twist with a centrifugal-flow left ventricular assist system
    Potapov, Evgenij V.
    Netuka, Ivan
    Kaufmann, Friedrich
    Falk, Volkmar
    Mehra, Mandeep R.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2018, 37 (05) : 670 - 673
  • [6] Diagnosis and Treatment Algorithm for Blood Flow Obstructions in Patients With Left Ventricular Assist Device
    Scandroglio, Anna Mara
    Kaufmann, Friedrich
    Pieri, Marina
    Kretzschmar, Alexandra
    Mueller, Marcus
    Pergantis, Panagiotis
    Dreysse, Stephan
    Falk, Volkmar
    Krabatsch, Thomas
    Potapov, Evgenij V.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2016, 67 (23) : 2758 - 2768
  • [7] Interventional Treatment of LVAD Outflow Graft Stenosis by Introduction of Bare Metal Stents
    Wiedemann, Dominik
    Schloeglhofer, Thomas
    Haberl, Thomas
    Riebandt, Julia
    Dimitrov, Kamen
    Schima, Heinrich
    Kastner, Johannes
    Matzek, Wolfgang
    Laufer, Guenther
    Zimpfer, Daniel
    [J]. ASAIO JOURNAL, 2018, 64 (01) : E3 - E7
  • [8] Survival outcomes of stenting outflow graft stenosis in continuous-flow left ventricular assist devices: a systematic review
    Wood, Chelsey T.
    O'Malley, Thomas J.
    Maynes, Elizabeth J.
    Vishnevsky, Alec
    Morris, Rohinton J.
    Samuels, Louis E.
    Massey, H. Todd
    Tchantchaleishvili, Vakhtang
    [J]. HEART FAILURE REVIEWS, 2020, 25 (06) : 985 - 992