High-efficiency liquid chromatography-mass spectrometry separations with 50 mm, 250 mm, and 1 m long polymer-based monolithic capillary columns for the characterization of complex proteolytic digests

被引:61
作者
Eeltink, Sebastiaan [1 ,2 ]
Dolman, Sebastiaan [2 ]
Detobel, Frederik [1 ]
Swart, Remco [2 ]
Ursem, Mario [2 ]
Schoenmakers, Peter J. [3 ]
机构
[1] Vrije Univ Brussel, Dept Chem Engn, B-1050 Brussels, Belgium
[2] Dionex, NL-1046 AA Amsterdam, Netherlands
[3] Univ Amsterdam, Vant Hoff Inst Mol Sci, NL-1018 WV Amsterdam, Netherlands
关键词
Proteomics; High-performance liquid chromatography; Nano-LC; Mass spectrometry; STATIONARY PHASES; PEAK-CAPACITY; FABRICATION; PARTICLES; ROD;
D O I
10.1016/j.chroma.2010.03.037
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In this study, high-efficiency LC-MS/MS separations of complex proteolytic digests are demonstrated using 50 mill. 250 min. and 1 m long poly(styrene-co-divinylbenzene) monolithic capillary columns The chromatographic performance of the 50 and 250 mm monoliths was compared at the same gradient steepness for gradient durations between 5 and 150 mm. The maximum peak capacity of 400 obtained with a 50mm column, increased to 485 when using the 250 mm long column and scaling the gradient duration according column length With a 5-fold increase in column length only a 20% increase in peak capacity was observed, which could be explained by the larger macropore size of the 250 mm long monolith When taking into account the total analysis time, including the dwell time, gradient time and column equilibration time, the 50 mm long monolith yielded better peptide separations than the 250 mm long monolithic column for gradient times below 80 min (n(c) =370). For more demanding separation the 250 mm long monolith provided the highest peak production rate and consequently higher sequence coverage For the analysis of a proteolytic digest of Escherichia coli proteins a monolithic capillary column of 1 m in length was used, yielding a peak capacity of 1038 when applying a 600 min gradient (C) 2010 Elsevier B V All rights reserved
引用
收藏
页码:6610 / 6615
页数:6
相关论文
共 35 条
[1]   Towards the column bed stabilization of columns in capillary electroendosmotic chromatography - Immobilization of microparticulate silica columns to a continuous bed [J].
Adam, T ;
Unger, KK ;
Dittmann, MM ;
Rozing, GP .
JOURNAL OF CHROMATOGRAPHY A, 2000, 887 (1-2) :327-337
[2]   Silica-based monoliths for capillary electrochromatography: Methods of fabrication and their applications in analytical separations [J].
Allen, D ;
El Rassi, Z .
ELECTROPHORESIS, 2003, 24 (22-23) :3962-3976
[3]   Aspects of column fabrication for packed capillary electrochromatography [J].
Angus, PDA ;
Demarest, CW ;
Catalano, T ;
Stobaugh, JF .
JOURNAL OF CHROMATOGRAPHY A, 2000, 887 (1-2) :347-365
[4]   New synthetic ways for the preparation of high-performance liquid chromatography supports [J].
Buchmeiser, MR .
JOURNAL OF CHROMATOGRAPHY A, 2001, 918 (02) :233-266
[5]   Cancer proteomics by quantitative shotgun proteomics [J].
Chen, Emily I. ;
Yates, John R., III .
MOLECULAR ONCOLOGY, 2007, 1 (02) :144-159
[6]   Pillar-structured microchannell's for on-chip liquid chromatography: Evaluation of the permeability and separation performance [J].
De Pra, Mauro ;
De Malsche, Wim ;
Desmet, Gert ;
Schoenmakers, Peter J. ;
Kok, Wim Th. .
JOURNAL OF SEPARATION SCIENCE, 2007, 30 (10) :1453-1460
[7]   Data analysis and bioinformatics tools for tandem mass spectrometry in proteomics [J].
Deutsch, Eric W. ;
Lam, Henry ;
Aebersold, Ruedi .
PHYSIOLOGICAL GENOMICS, 2008, 33 (01) :18-25
[8]   Reversed-phase liquid chromatographic separation of complex samples by optimizing temperature and gradient time I. Peak capacity limitations [J].
Dolan, JW ;
Snyder, LR ;
Djordjevic, NM ;
Hill, DW ;
Waeghe, TJ .
JOURNAL OF CHROMATOGRAPHY A, 1999, 857 (1-2) :1-20
[9]   Performance limits of monolithic and packed capillary columns in high-performance liquid chromatography and capillary electrochromatography [J].
Eeltink, S ;
Desmet, G ;
Vivó-Truyols, G ;
Rozing, GP ;
Schoenmakers, PJ ;
Kok, WT .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1104 (1-2) :256-262
[10]   Tailoring the morphology of methacrylate ester-based monoliths for optimum efficiency in liquid chromatography [J].
Eeltink, S ;
Herrero-Martinez, JM ;
Rozing, GP ;
Schoenmakers, PJ ;
Kok, WT .
ANALYTICAL CHEMISTRY, 2005, 77 (22) :7342-7347