A dosimetric analysis of the rectal doses in intracavitary brachytherapy of carcinoma cervix: A prospective study from a single institute

被引:0
作者
Suryadevara, Aparna [1 ]
Talluri, And Kumar [1 ]
Alluri, Krishnam Raju [1 ]
Vasundhara, E. [1 ]
Kumar, Sudhakar [1 ]
Sresty, N. V. N. Madhusudhana [1 ]
机构
[1] Basavatarakam Indo Amer Canc Hosp & Res Inst, Dept Radiat Oncol, Hyderabad, Telangana, India
关键词
Henschke applicator; ICRU 38 rectal point; modified ICRU rectal point; rectal dose; rectal wire; RADIATION-THERAPY; UTERINE CERVIX; COMPLICATIONS; PREDICTION; CANCER;
D O I
10.4103/jmp.JMP_50_19
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Carcinoma cervix is a common gynecologic malignancy in India and is treated with radical chemoradiation where intracavitary brachytherapy (ICR) is an integral part. In ICR of cervix, the two-dimensional (2D) point-based dosimetry cervix is the most common method used in high-volume centers with rectal dose calculation at modified ICRU rectal point with rectal wire placement. The rectal dose measurement using this method underestimates the dose to the rectum, and rectal dose also varies with the type of applicator used. The aim of our study is to compare the rectal dose calculated by ICRU 38 method versus rectal dose calculated by the rectal wire method using Henschke applicator. Materials and Methods: This is a single-institute, dosimetric comparison study done prospectively. Fifty patients were planned for ICR after 2D orthogonal radiograph-based, computer planning by iridium 192 high-dose rate remote afterloading technique after placing the appropriate Henschke applicator. The vaginal packing was done using sterile gauze with contrast material for defining the ICRU 38 rectal point, and a rectal wire was placed for the modified ICRU rectal point. Rectal doses were calculated by both the methods and compared. Results: The modified ICRU rectal point recorded a lower rectal dose (mean of 25%) compared to ICRU 38 rectal point in the study patients. There were ten patients (20%) with either too much or too little contrast material which made the visualization of the rectal point and radiation planning difficult. P value by paired t-test method was 0.0001, which was statistically significant. Conclusion: The modified ICRU rectal point is easier to visualize than ICRU 38 method (100% vs. 80%) for dosimetry, but it underestimates the rectal doses when compared to ICRU 38 rectal point. There needs to be a correction factor applied (25% in our study for Henschke applicator) when evaluating the rectal doses calculated by rectal wire method, to reduce the rectal toxicity.
引用
收藏
页码:287 / 291
页数:5
相关论文
共 16 条
[1]  
Amin M.B., 2017, AJCC CANC STAGING MA, VVIII ed.
[2]   Brachytherapy in the treatment of cervical cancer: a review [J].
Banerjee, Robyn ;
Kamrava, Mitchell .
INTERNATIONAL JOURNAL OF WOMENS HEALTH, 2014, 6 :555-564
[3]   Impact on treatment outcome and late effects of customized treatment planning in cervix carcinomas: Baseline results to compare new strategies [J].
Barillot, I ;
Horiot, JC ;
Maingon, P ;
Truc, G ;
Chaplain, G ;
Comte, J ;
Brenier, JP .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 48 (01) :189-200
[4]   Global estimates of cancer prevalence for 27 sites in the adult population in 2008 [J].
Bray, Freddie ;
Ren, Jian-Song ;
Masuyer, Eric ;
Ferlay, Jacques .
INTERNATIONAL JOURNAL OF CANCER, 2013, 132 (05) :1133-1145
[5]   The prediction of late rectal complications following the treatment of uterine cervical cancer by high-dose-rate brachytherapy [J].
Chen, SW ;
Liang, JA ;
Yang, SN ;
Liu, RT ;
Lin, FJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 47 (04) :955-961
[6]   The prediction of late rectal complications in patients treated with high dose-rate brachytherapy for carcinoma of the cervix [J].
Clark, BG ;
Souhami, L ;
Roman, TN ;
Chappell, R ;
Evans, MDC ;
Fowler, JF .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (05) :989-993
[7]  
Ferlay J., 2013, GLOBOCAN 2012 CANC I
[8]  
Halperin E. C., 2013, PEREZ BRADYS PRINCIP
[9]  
Huh SJ, 2003, STRAHLENTHER ONKOL, V179, P191, DOI 10.1007/s00066-003-1015-2
[10]   Comparisons between the KKU-Model and Conventional Rectal Tubes as Markers for Checking Rectal Doses during Intracavitary Brachytherapy of Cervical Cancer [J].
Padoongcharoen, Prawat ;
Krusun, Srichai ;
Palusuk, Voranipit ;
Pesee, Montien ;
Supaadirek, Chunsri ;
Thamronganantasakul, Komsan .