A designed setup of low-priced in-house goniometer/tensiometer

被引:2
作者
Ahmed, Zeeshan [1 ]
Sarode, Ajinkya [2 ]
Tripathi, Abhishek Kumar [3 ]
Parida, Satyajeet [3 ]
Singh, V. K. [4 ]
机构
[1] Aditya Engn Coll, Dept Mech Engn, Surampalem 53343, Andhra Pradesh, India
[2] Indian Inst Technol Gandhinagar, Dept Mech Engn, Palaj 382355, Gujarat, India
[3] Aditya Engn Coll, Dept Min Engn, Surampalem 53343, Andhra Pradesh, India
[4] Space Applicat Ctr, Thermal Engn Div, Ahmadabad 380015, Gujarat, India
来源
OPTIK | 2022年 / 258卷
关键词
Goniometer; Surface tension; Contact angle; Droplet; Tensiometer; CONTACT-ANGLE MEASUREMENT; SURFACE-TENSION;
D O I
10.1016/j.ijleo.2022.168783
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
In many educational institutes and industries, the contact angle and surface tension analysis are done to measure the quality of a solid surface and of a liquid. In order to determine these, the sessile and pendant droplet method are the most widely used metrics, which are conventionally obtained using commercial goniometer/tensiometer. In view of the growing concern regarding compact and inexpensive but accurate experimental devices, we have developed a low-priced inhouse goniometer/tensiometer. This device has the benefit of easiness, compactness, and movability over the conventional measurement instruments. In this study, we have used the commercial Drop image advanced as well as the Drop analysis plugin in ImageJ with high quality sharp edges images to compute static contact angle and surface tension values. Detailed procedure to measure them have been discussed in this work. To validate the apparatus, various water-based solutions have been considered. The surface tension values obtained using the apparatus are in good agreement with those reported in the literature. However, the apparatus needs to validate for highly wettable and volatile fluids. The efficacious manifestation of this device to determine the contact angle is a compelling improvement in the area as it disrupts the trendof a bench designated setup which is heavy, bulky and expensive and are not suitable for field work. The standard threaded syringe-based low-priced goniometer/tensiometer fabricated is affordable, being less than 40 USD or 3000 INR. Variety of water-based solutions like Millipore water, SDS solution at CMC * (8.1 mM), SDS solution at CMC/2, SDS solution at 2xCMC have been considered for validation and compared with surface tension values. Such a low-cost goniometer prototype can serve the technological needs of scientific research labs, in imparting practical learning about repeatability and highly beneficial for industrial applications.
引用
收藏
页数:18
相关论文
共 20 条
[1]   Multi-wavelength photodetectors based on porous spin-coated and compact RF-sputtered NiO films grown over Si substrate: Effect of surface morphology [J].
Ahmed, Anas A. ;
Hashim, M. R. ;
Qahtan, Talal F. ;
Rashid, Marzaini .
OPTIK, 2022, 255
[2]  
[Anonymous], 2002, ENCY SURFACE COLLOID
[3]  
Arashiro E. Y., 1999, MAT RES, V2, P23, DOI [10.1590/S1516-14391999000100005, DOI 10.1590/S1516-14391999000100005]
[4]   Measurement of surface and interfacial tension using pendant drop tensiometry [J].
Berry, Joseph D. ;
Neeson, Michael J. ;
Dagastine, Raymond R. ;
Chan, Derek Y. C. ;
Tabor, Rico F. .
JOURNAL OF COLLOID AND INTERFACE SCIENCE, 2015, 454 :226-237
[5]   Symmetric and antisymmetric solitons in the fractional nonlinear schro•dinger equation with saturable nonlinearity and PT-symmetric potential: Stability and dynamics [J].
Bo, Wen-Bo ;
Liu, Wei ;
Wang, Yue-Yue .
OPTIK, 2022, 255
[6]   Liquid bridge as a tunable-focus cylindrical liquid lens [J].
Chen, H. ;
Tabatabaei, N. ;
Amirfazli, A. .
APPLIED PHYSICS LETTERS, 2017, 110 (04)
[7]  
Daerr A., 2016, J OPEN RES SOFTWARE, V4, DOI DOI 10.5334/JORS.97
[8]   Contact angle measurements by axisymmetric drop shape analysis and an automated polynomial fit program [J].
del Río, OI ;
Kwok, DY ;
Wu, R ;
Alvarez, JM ;
Neumann, AW .
COLLOIDS AND SURFACES A-PHYSICOCHEMICAL AND ENGINEERING ASPECTS, 1998, 143 (2-3) :197-210
[9]   Surface tension in human pathophysiology and its application as a medical diagnostic tool [J].
Fathi-Azarbayjani, Anahita ;
Jouyban, Abolghasem .
BIOIMPACTS, 2015, 5 (01) :29-44
[10]   A precise goniometer/tensiometer using a low cost single-board computer [J].
Favier, Benoit ;
Chamakos, Nikolaos T. ;
Papathanasiou, Athanasios G. .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2017, 28 (12)