A case example and literature review of catastrophic wear before catastrophic failure: identification of trunnionosis and metallosis in metal-on-polyethylene hip arthroplasty prior to frank failure or fracture

被引:8
作者
Hunter, Allison M. [1 ]
Hsu, Alan [1 ]
Shah, Ashish [1 ]
Naranje, Sameer M. [1 ]
机构
[1] Univ Alabama Birmingham, Dept Orthoped Surg, Birmingham, AL 35210 USA
关键词
Arthroplasty; Metallosis; Metal-on-polyethylene; Metal-on-metal; Trunnionosis;
D O I
10.1007/s00590-018-2333-5
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Though rare, there are documented failures of femoral prosthesis due to corrosion of the head-neck interface in total hip arthroplasty (THA), a phenomenon known as trunnionosis. This wear can result in metallosis, whereby metal debris scatters the surrounding soft tissues. We present on a 58-year-old female who presented with increase in hip and back pain 10 years following right THA using a metal-on-polyethylene construct with a large femoral head (44 mm). Aspiration withdrew metallic fluid, and intraoperative findings showed corrosion of the head-neck taper with surrounding metallosis and pseudocapsule formation. Despite advances in THA design, corrosion and wear between components still exists and may be cause for failure. We present on both the subtle clinical findings and the recommended workup when suspicion is high for trunnionosis, metallosis, or wear, ideally with identification prior to catastrophic failure such as component dislocation or fracture as previously reported.
引用
收藏
页码:711 / 715
页数:5
相关论文
共 16 条
[1]  
Arvinte Dan, 2015, JBJS Case Connect, V5, pe35, DOI 10.2106/JBJS.CC.N.00146
[2]   Gross Trunnion Failure After Primary Total Hip Arthroplasty [J].
Banerjee, Samik ;
Cherian, Jeffrey J. ;
Bono, James V. ;
Kurtz, Steven M. ;
Geesink, Rudolph ;
Meneghini, R. Michael ;
Delanois, Ronald E. ;
Mont, Michael A. .
JOURNAL OF ARTHROPLASTY, 2015, 30 (04) :641-648
[3]   Adverse Local Tissue Reaction Arising from Corrosion at the Femoral Neck-Body Junction in a Dual-Taper Stem with a Cobalt-Chromium Modular Neck [J].
Cooper, H. John ;
Urban, Robert M. ;
Wixson, Richard L. ;
Meneghini, R. Michael ;
Jacobs, Joshua J. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2013, 95A (10) :865-872
[4]   Corrosion at the Head-Neck Taper as a Cause for Adverse Local Tissue Reactions After Total Hip Arthroplasty [J].
Cooper, H. John ;
Della Valle, Craig J. ;
Berger, Richard A. ;
Tetreault, Matthew ;
Paprosky, Wayne G. ;
Sporer, Scott M. ;
Jacobs, Joshua J. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2012, 94A (18) :1655-1661
[5]   Why do some titanium-alloy total hip arthroplasty modular necks fail? [J].
Fokter, Samo K. ;
Molicnik, Andrej ;
Kavalar, Rajko ;
Pelicon, Primoz ;
Rudolf, Rebeka ;
Gubeljak, Nenad .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2017, 69 :107-114
[6]   Case Report: Femoral Neuropathy Secondary to Total Hip Arthroplasty Wear Debris [J].
Fokter, Samo K. ;
Repse-Fokter, Alenka ;
Takac, Iztok .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2009, 467 (11) :3032-3035
[7]   In vitro corrosion testing of modular hip tapers [J].
Goldberg, JR ;
Gilbert, JL .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2003, 64B (02) :78-93
[8]   Ten-Year Cross-Sectional Study of Mechanically Assisted Crevice Corrosion in 1352 Consecutive Patients With Metal-on-Polyethylene Total Hip Arthroplasty [J].
Hussey, Daniel K. ;
McGrory, Brian J. .
JOURNAL OF ARTHROPLASTY, 2017, 32 (08) :2546-2551
[9]   Catastrophic Femoral Head-Stem Trunnion Dissociation Secondary to Corrosion [J].
Ko, Laura Matsen ;
Chen, Antonia F. ;
Deirmengian, Gregory K. ;
Hozack, William J. ;
Sharkey, Peter F. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2016, 98 (16) :1400-1404
[10]   Quality of life and functionality after total hip arthroplasty: a long-term follow-up study [J].
Mariconda, Massimo ;
Galasso, Olimpio ;
Costa, Giovan Giuseppe ;
Recano, Pasquale ;
Cerbasi, Simone .
BMC MUSCULOSKELETAL DISORDERS, 2011, 12