Novel use of a modified Amplatzer Vascular Plug® to occlude a patent ductus arteriosus in two patients

被引:21
作者
Glatz, Andrew C. [1 ]
Petit, Christopher J. [1 ]
Gillespie, Matthew J. [1 ]
机构
[1] Univ Penn, Childrens Hosp Philadelphia, Sch Med, Div Cardiol, Philadelphia, PA 19104 USA
关键词
cardiac catheterization; congenital heart disease; embolization coils; device closure; transcatheter occlusion;
D O I
10.1002/ccd.21546
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We describe two cases in which a patent ductus arteriosus (PDA) was successfully occluded percutaneously, using the novel approach of prefilling an Amplatzer Vascular Plug (R) with embolization coils to decrease the risk of residual shunting through the device. One patient was a small premature neonate, and the other was a child with an aneurismal, tubular PDA with no aortic ampulla. These are situations in which more popular PDA devices such as the Amplatzer Duct Occlude (R) may be contraindicated. (c) 2008 Wiley-Liss, Inc.
引用
收藏
页码:82 / 86
页数:5
相关论文
共 24 条
[1]   The Amplatzer duct occluder: Experience in 209 patients [J].
Bilkis, AA ;
Alwi, M ;
Hasri, S ;
Haifa, AL ;
Geetha, K ;
Rehman, MA ;
Hasanah, I .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (01) :258-261
[2]   Not so fast with that novel use: Does AVP = PDA? [J].
Cheatham, JP .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2005, 65 (04) :581-583
[3]  
Ebeid M R, 2001, J Interv Cardiol, V14, P33
[4]   Challenges encountered during closure of patent ductus arteriosus [J].
Ewert, P .
PEDIATRIC CARDIOLOGY, 2005, 26 (03) :224-229
[5]  
Faella HJ, 2000, CATHETER CARDIO INTE, V51, P50, DOI 10.1002/1522-726X(200009)51:1<50::AID-CCD11>3.0.CO
[6]  
2-6
[7]   Transcatheter closure of persistent ductus arteriosus in infants using the Amplatzer duct occluder [J].
Fischer, G ;
Stieh, J ;
Uebing, A ;
Grabitz, R ;
Kramer, HH .
HEART, 2001, 86 (04) :444-447
[8]   Transcatheter closure of the patent ductus arteriosus [J].
Grifka, RG .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 61 (04) :554-570
[9]   Transcatheter closure of large patent ductus arteriosus (>=4 mm) with multiple Gianturco coils: Immediate and mid-term results [J].
Hijazi, ZM ;
Geggel, RL .
HEART, 1996, 76 (06) :536-540
[10]   Evaluation of the AMPLATZER vascular plug for embolization of peripheral vascular malformations associated with congenital heart disease [J].
Hill, SL ;
Hijazi, ZM ;
Hellenbrand, WE ;
Cheatham, JP .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2006, 67 (01) :113-119