A biomechanical comparison of the Surgical Implant Generation Network (SIGN) tibial nail with the standard hollow nail

被引:12
作者
Calafi, L. A. [1 ]
Antkowiak, T. [2 ]
Curtiss, S. [2 ]
Neu, C. P. [3 ]
Moehring, D. [2 ]
机构
[1] Univ Washington, Harborview Med Ctr, Dept Orthopaed Surg, Seattle, WA 98104 USA
[2] Univ Calif Davis, Med Ctr, Dept Orthopaed Surg, Sacramento, CA 95817 USA
[3] Purdue Univ, Weldon Sch Biomed Engn, W Lafayette, IN 47907 USA
来源
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED | 2010年 / 41卷 / 07期
关键词
Intramedullary nailing; Tibia fractures; SIGN; Russell-Taylor; Mechanical stiffness; INVASIVE STABILIZATION SYSTEM; SHAFT FRACTURES; INTERNAL-FIXATION; MANAGEMENT;
D O I
10.1016/j.injury.2010.03.003
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: In developing countries, tibial shaft fractures are frequently stabilised using Surgical Implant Generation Network (SIGN) nails. Despite widespread use throughout the world, little is known regarding their biomechanical properties. This study aimed to compare the mechanical stiffness of the SIGN tibial nail with a standard hollow tibial nail. Methods: A fracture gap model was created to simulate a comminuted mid-shaft tibia fracture (AO/OTA42-C3) using synthetic composite bones. The constructs were stabilised with either a 9 mm solid SIGN nail or a 10 mm hollow Russell-Taylor nail. Both nail systems were interlocked proximally and distally. Following fixation, the specimens were loaded in axial, torsional, and cyclical axial modes to calculate construct stiffness and irreversible (plastic) deformation. Results: The mean axial stiffness for the SIGN nail constructs was 47% higher than mean stiffness for the RT nail constructs (p < 0.001). The difference in torsional stiffness was not statistically significant. However, the SIGN group demonstrated 159% more irreversible deformation than the Russell-Taylor group (p = 0.006) for the loading parameters studied. Conclusion: The SIGN tibial nail, despite its slightly smaller diameter, can provide similar construct stiffness and stability, when compared to a larger hollow nail for stabilisation of tibial shaft fractures. (C) 2010 Elsevier Ltd. All rights reserved.
引用
收藏
页码:753 / 757
页数:5
相关论文
共 22 条
[1]  
ALHO A, 1992, CLIN ORTHOP RELAT R, P243
[2]   THE BIOMECHANICAL PROPERTIES OF HOLLOW AND SOLID NAILS FOR INTERLOCKED INTRAMEDULLARY NAILING - EXPERIMENTAL STUDIES IN THE DOG [J].
BOHM, P ;
SCHERER, MA ;
FRUH, HJ ;
ASCHERL, R ;
MOTZER, E .
BIOMEDIZINISCHE TECHNIK, 1995, 40 (1-2) :24-28
[3]   TREATMENT OF TIBIAL FRACTURES BY REAMING AND INTRAMEDULLARY NAILING [J].
BONE, LB ;
JOHNSON, KD .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (06) :877-887
[4]  
Coles CP, 2000, CAN J SURG, V43, P256
[5]  
Cordey J, 1999, INJURY, V30, pS26
[6]   Reamed or unreamed nailing for closed tibial fractures - A prospective study in Tscherne C1 fractures [J].
CourtBrown, CM ;
Will, E ;
Christie, J ;
McQueen, MM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1996, 78B (04) :580-583
[7]   Mechanical validation of whole bone composite femur models [J].
Cristofolini, L ;
Viceconti, M ;
Cappello, A ;
Toni, A .
JOURNAL OF BIOMECHANICS, 1996, 29 (04) :525-535
[8]   A prospective, randomized study of intramedullary nails inserted with and without reaming for the treatment of open and closed fractures of the tibial shaft [J].
Finkemeier, CG ;
Schmidt, AH ;
Kyle, RF ;
Templeman, DC ;
Varecka, TE .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2000, 14 (03) :187-193
[9]   Structural properties of a new design of composite replicate femurs and tibias [J].
Heiner, AD ;
Brown, TD .
JOURNAL OF BIOMECHANICS, 2001, 34 (06) :773-781
[10]  
HENLEY MB, 1989, CLIN ORTHOP RELAT R, P87