Optimization of exposure parameters for cone beam computed tomography sialography

被引:22
作者
Jadu, F. M. [1 ]
Hill, M. L. [2 ]
Yaffe, M. J. [2 ]
Lam, E. W. N. [1 ]
机构
[1] Univ Toronto, Fac Dent, Toronto, ON, Canada
[2] Sunnybrook Hlth Sci Ctr, Toronto, ON M4N 3M5, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
cone beam computed tomography; sialography; IMAGE QUALITY; CONTRAST MATERIAL; ROW CT; LIVER;
D O I
10.1259/dmfr/81159071
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: The assessment of image quality is a crucial step in the development of a new imaging protocol. Having proposed and reported on a preliminary protocol for sialography using cone beam CT (CBCT), the purpose of this study was to further optimize this protocol by maximizing the image signal difference-to-noise ratio (SDNR) and to relate these new data to previously published dosimetric data for CBCT sialography. Methods: An imaging phantom was constructed using samples with different concentrations of iodine and a water-immersed mandible. The CB MercuRay (Hitachi Medical Systems, Tokyo, Japan) was used to image the phantom using different peak kilovoltage (kVp) and milliamperage (mA) settings. SDNR was then calculated using the raw images based on mean pixel values (MPV) measured in selected regions of interest (ROI). Finally, a figure of merit (FOM) was calculated to examine the trade-off between image SDNR and effective radiation dose. Results: The SDNR demonstrated an expected increase as the kVp increased from 60 to 120. Also, images made with the higher mA setting (15) had greater SDNR. The iodine concentration also influenced the image quality such that SDNR increased with increased amounts of iodine. The calculated FOM was greatest for the technique using 80 kVp, with equivalent results for 10 mA and 15 mA. Conclusion: An optimized protocol for CBCT sialography using CB MercuRay entails a 6 inch field of view with 80 kVp and 10 mA. Dentomaxillofacial Radiology (2011) 40, 362-368. doi: 10.1259/dmfr/81159071
引用
收藏
页码:362 / 368
页数:7
相关论文
共 15 条
[1]   Aortic and hepatic enhancement and tumor-to-liver contrast: Analysis of the effect of different concentrations of contrast material at multi-detector row helical CV [J].
Awai, K ;
Takada, K ;
Onishi, H ;
Hori, S .
RADIOLOGY, 2002, 224 (03) :757-763
[2]   Artifacts in CT: Recognition and avoidance [J].
Barrett, JF ;
Keat, N .
RADIOGRAPHICS, 2004, 24 (06) :1679-1691
[3]   Intravenous contrast material administration at helical 16-detector row CT coronary angiography: Effect of iodine concentration on vascular attenuation [J].
Cademartiri, F ;
Mollet, NR ;
van der Lugt, A ;
McFadden, EP ;
Stijnen, T ;
de Feyter, PJ ;
Krestin, GP .
RADIOLOGY, 2005, 236 (02) :661-665
[4]  
Cademartiri Filippo, 2006, AJR Am J Roentgenol, V187, pW430, DOI 10.2214/AJR.05.1406
[5]  
*FARL INC, THEFR FIG MER
[6]   A solid iodinated phantom material for use in tomographic x-ray imaging [J].
Hill, Melissa L. ;
Mainprize, James G. ;
Mawdsley, Gordon E. ;
Yaffe, Martin J. .
MEDICAL PHYSICS, 2009, 36 (10) :4409-4420
[7]   A comparative study of the effective radiation doses from cone beam computed tomography and plain radiography for sialography [J].
Jadu, F. ;
Yaffe, M. J. ;
Lam, E. W. N. .
DENTOMAXILLOFACIAL RADIOLOGY, 2010, 39 (05) :257-263
[8]   Image quality produced by different cone-beam computed tomography settings [J].
Kwong, Jeffrey C. ;
Palomo, J. Martin ;
Landers, Michael A. ;
Figueroa, Alex ;
Hans, Mark G. .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2008, 133 (02) :317-327
[9]   A comparative evaluation of Cone Beam Computed Tomography (CBCT) and Multi-Slice CT (MSCT) Part I. On subjective image quality [J].
Liang, Xin ;
Jacobs, Reinhilde ;
Hassan, Bassam ;
Li, Limin ;
Pauwels, Ruben ;
Corpas, Livia ;
Souza, Paulo Couto ;
Martens, Wendy ;
Shahbazian, Maryam ;
Alonso, Arie ;
Lambrichts, Ivo .
EUROPEAN JOURNAL OF RADIOLOGY, 2010, 75 (02) :265-269
[10]   Measurement of image quality in diagnostic radiology [J].
Martin, CJ ;
Sharp, PF ;
Sutton, DG .
APPLIED RADIATION AND ISOTOPES, 1999, 50 (01) :21-38