摘要
Purpose: In this study, the influence of the cardiac cycle on the amount of label produced by a velocity-selective (VSASL) and acceleration-selective arterial spin labeling (AccASL) module was investigated. Methods: A short-PLD sequence was developed where a single VSASL- or AccASL-module was preceded by pCASL labeling to isolate the arterial blood pool. ASL subtraction was performed with label/control images with similar cardiac phase and time-of-measurement, followed by retrospective binning in 10 cardiac phase bins. ASL signal variation over the heart cycle was evaluated and tested for significance using a permutation test. Results: VSASL and AccASL showed significant arterial signal fluctuations over the cardiac cycle of up to similar to 36% and similar to 64%, respectively, mainly in areas containing large arteries. pCASL also showed significant signal fluctuations, of up to similar to 25% in arteries. Raw label/control images confirmed that the observed signal fluctuations were caused by the amount of label produced during the cardiac cycle, rather than inflow-effects, because the raw images did not all show equal cardiac phase dependence. No significant effects of the cardiac cycle were found on the gray matter ASL-signal. Conclusion: Significant influence of the cardiac cycle on the generated label was found for spatially nonselective ASL-sequences. Hence, to become independent of the cardiac cycle, sufficient averages need to be taken. Alternatively, these findings could be highly interesting for the purpose of quantifying pulsatility more distally in the vascular tree.
引用
收藏
页码:872 / 882
页数:11
相关论文
共 19 条
[1]   Assessment of blood flow velocity and pulsatility in cerebral perforating arteries with 7-T quantitative flow MRI [J].
Bouvy, W. H. ;
Geurts, L. J. ;
Kuijf, H. J. ;
Luijten, P. R. ;
Kappelle, L. J. ;
Biessels, G. J. ;
Zwanenburg, J. J. M. .
NMR IN BIOMEDICINE, 2016, 29 (09) :1295-1304
[2]   Measuring the labeling efficiency of pseudocontinuous arterial spin labeling [J].
Chen, Zhensen ;
Zhang, Xingxing ;
Yuan, Chun ;
Zhao, Xihai ;
van Osch, Matthias J. P. .
MAGNETIC RESONANCE IN MEDICINE, 2017, 77 (05) :1841-1852
[3]   Localization of cardiac-induced signal change in fMRI [J].
Dagli, MS ;
Ingeholm, JE ;
Haxby, JV .
NEUROIMAGE, 1999, 9 (04) :407-415
[4]   Timing dependence of peripheral pulse-wave-triggered pulsed arterial spin labeling [J].
Fushimi, Yasutaka ;
Okada, Tomohisa ;
Yamamoto, Akira ;
Kanagaki, Mitsunori ;
Fujimoto, Koji ;
Togashi, Kaori .
NMR IN BIOMEDICINE, 2013, 26 (11) :1527-1533
[5]   Pulsed star labeling of arterial regions (PULSAR): A robust regional perfusion technique for high field imaging [J].
Golay, X ;
Petersen, ET ;
Hui, F .
MAGNETIC RESONANCE IN MEDICINE, 2005, 53 (01) :15-21
[6]   Cardiorespiratory noise correction improves the ASL signal [J].
Hassanpour, Mahlega S. ;
Luo, Qingfei ;
Simmons, W. Kyle ;
Feinstein, Justin S. ;
Paulus, Martin P. ;
Luh, Wen-Ming ;
Bodurka, Jerzy ;
Khalsa, Sahib S. .
HUMAN BRAIN MAPPING, 2018, 39 (06) :2353-2367
[7]   Improved sensitivity and temporal resolution in perfusion FMRI using velocity selective inversion ASL [J].
Hernandez-Garcia, Luis ;
Nielsen, Jon-Fredrik ;
Noll, Douglas C. .
MAGNETIC RESONANCE IN MEDICINE, 2019, 81 (02) :1004-1015
[8]   Cardiac-triggered pseudo-continuous arterial-spin-labeling: A cost-effective scheme to further enhance the reliability of arterial-spin-labeling MRI [J].
Li, Yang ;
Mao, Deng ;
Li, Zhiqiang ;
Schaer, Michael ;
Pillai, Jay J. ;
Pipe, James G. ;
Lu, Hanzhang .
MAGNETIC RESONANCE IN MEDICINE, 2018, 80 (03) :969-975
[9]   Determination of Spin Compartment in Arterial Spin Labeling MRI [J].
Liu, Peiying ;
Uh, Jinsoo ;
Lu, Hanzhang .
MAGNETIC RESONANCE IN MEDICINE, 2011, 65 (01) :120-U3
[10]   Effects of central arterial aging on the structure and function of the peripheral vasculature: implications for end-organ damage [J].
Mitchell, Gary F. .
JOURNAL OF APPLIED PHYSIOLOGY, 2008, 105 (05) :1652-1660