Simple and rapid analytical method for detection of amino acids in blood using blood spot on filter paper, fast-GC/MS and isotope dilution technique

被引:22
作者
Kawana, Shuichi [1 ]
Nakagawa, Katsuhiro [1 ]
Hasegawa, Yuki [1 ]
Yamaguchi, Seiji [1 ]
机构
[1] Shimane Univ, Dept Pediat, Sch Med, Izumo, Shimane 6938501, Japan
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2010年 / 878卷 / 30期
关键词
Gas chromatograph-Mass spectrometer; Amino acids; Propyl chloroformate; Blood spot sample; CHROMATOGRAPHY-MASS-SPECTROMETRY; SYRUP-URINE-DISEASE; GC-MS ANALYSIS; GAS-CHROMATOGRAPHY; CHEMICAL-IONIZATION; INBORN-ERRORS; PLASMA AMINO; ELECTROSPRAY-IONIZATION; INTERNAL STANDARDS; DERIVATIZATION;
D O I
10.1016/j.jchromb.2010.09.015
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A simple and rapid method for quantitative analysis of amino acids, including valine (Val), leucine (Leu), isoleucine (Ile), methionine (Met) and phenylalanine (Phe), in whole blood has been developed using GC/MS. In this method, whole blood was collected using a filter paper technique, and a 1/8 in. blood spot punch was used for sample preparation. Amino acids were extracted from the sample, and the extracts were purified using cation-exchange resins. The isotope dilution method using H-2(8)-Val, H-2(3)-Leu, H-2(3)-Met and H-2(5)-Phe as internal standards was applied. Following propyl chloroformate derivatization, the derivatives were analyzed using fast-GC/MS. The extraction recoveries using these techniques ranged from 69.8% to 87.9%, and analysis time for each sample was approximately 26 min. Calibration curves at concentrations from 0.0 to 1666.7 mu mol/l for Val, Leu, Ile and Phe and from 0.0 to 333.3 mu mol/l for Met showed good linearity with regression coefficients = 1. The method detection limits for Val, Leu, Ile, Met and Phe were 24.2, 16.7, 8.7, 1.5 and 12.9 mu mol/l, respectively. This method was applied to blood spot samples obtained from patients with phenylketonuria (PKU), maple syrup urine disease (MSUD), hypermethionine and neonatal intrahepatic cholestasis caused by citrin deficiency (NICCD), and the analysis results showed that the concentrations of amino acids that characterize these diseases were increased. These results indicate that this method provides a simple and rapid procedure for precise determination of amino acids in whole blood. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:3113 / 3118
页数:6
相关论文
共 41 条
[1]   STUDY OF PLASMA FREE AMINO-ACID LEVELS .1. STUDY OF FACTORS AFFECTING VALIDITY OF AMINO-ACID ANALYSES [J].
ARMSTRONG, MD ;
STAVE, U .
METABOLISM-CLINICAL AND EXPERIMENTAL, 1973, 22 (04) :549-560
[2]  
BLAU R, 2003, PHYS GUIDE LAB DIAGN
[3]  
Calder AG, 1999, RAPID COMMUN MASS SP, V13, P2080, DOI 10.1002/(SICI)1097-0231(19991115)13:21<2080::AID-RCM755>3.0.CO
[4]  
2-O
[5]   Use of tandem mass spectrometry for multianalyte screening of dried blood specimens from newborns [J].
Chace, DH ;
Kalas, TA ;
Naylor, EW .
CLINICAL CHEMISTRY, 2003, 49 (11) :1797-1817
[6]  
Dallüge J, 2002, J SEP SCI, V25, P608, DOI 10.1002/1615-9314(20020601)25:9<608::AID-JSSC608>3.0.CO
[7]  
2-
[8]   Rapid determination of amino acids in neonatal blood samples based on derivatization with isobutyl chloroformate followed by solid-phase microextraction and gas chromatography/mass spectrometry [J].
Deng, CH ;
Li, N ;
Zhang, XM .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2004, 18 (21) :2558-2564
[9]   Diagnosis of maple syrup urine disease by determination of L-valine, L-isoleucine, L-leucine and L-phenylalanine in neonatal blood spots by gas chromatography-mass spectrometry [J].
Deng, CH ;
Deng, YH .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 792 (02) :261-268
[10]   Rapid diagnosis of phenylketonuria and other aminoacidemias by quantitative analysis of amino acids in neonatal blood spots by gas chromatography-mass spectrometry [J].
Deng, CH ;
Shang, CQ ;
Hu, YM ;
Zhang, XM .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2002, 775 (01) :115-120