Technical Note: Dosimetric characterization of the dynamic beam flattening MLC sequence on a ring shaped, Jawless Linear Accelerator with double stacked MLC

被引:7
作者
Bollinger, Douglas [1 ]
Laugeman, Eric [2 ]
Li, Taoran [1 ]
Hilliard, Jessica [2 ]
Heermann, Ana [2 ]
Kim, Hyun [2 ]
Hugo, Geoff [2 ]
Mutic, Sasa [2 ]
Dong, Lei [1 ]
Cai, Bin [2 ]
机构
[1] Univ Penn, Dept Radiat Oncol, Philadelphia, PA 19104 USA
[2] Washington Univ, Dept Radiat Oncol, St Louis, MO 63110 USA
关键词
dynamic beam flattening; halcyon; photon radiotherapy; PHOTON;
D O I
10.1002/mp.14001
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo characterize the dosimetric features and limitations of the dynamic beam flattening (DBF) on the Halcyon 2.0 linear accelerator (Varian Medical Systems). MethodsA predefined multi-leaf collimator (MLC) sequence was introduced and used to flatten the 6 MV flattening filter free (FFF) beam on the Halcyon 2.0. Dosimetric characterizations of the flattened beams, including beam flatness, symmetry, percent depth dose (PDD), output factor and MU linearity, were investigated. Flatness and symmetry were obtained from profile measurements with both radiographic films (EDR2) and a two dimensional ion-chamber array (IC Profiler, Sun Nuclear Corporation). MU linearity, output factors, and PDDs were measured in a water tank with a CC13 ion chamber (Scanditronix Wellhofer, Nuremburg, Germany). In addition, the effect of the DBF sequence on 3D plan quality was evaluated by creating DBF plans for a 4-field box rectum and an AP/PA spine plan. Patient specific QA was performed on these plans. ResultsAt 100 cm SSD and 10 cm depth, a flatness of <3% was observed on both transversal and radial profiles for all square field sizes >= 10 cm with DBF. For both larger and smaller field sizes the flatness showed a tendency to increase as the fields got bigger or smaller, respectively. Similar trends in flatness were observed at all depths measured. All measured output factors for square field sizes >= 5 cm were within 1% of the TPS prediction. Linearity was <= 2.02% for all measurements. For both treatment sites, the MD judged the plans created for the Halcyon without the use of DBF not to be clinically acceptable, however considered both the TrueBeam plan and the Halcyon plan with the DBF sequence to be clinically acceptable. ConclusionsThe DBF sequence on the Halcyon and its characteristics were investigated. The analysis indicates that the DBF sequence can be used on the Halcyon to generate clinically acceptable 3D treatment plans. (c) 2019 American Association of Physicists in Medicine
引用
收藏
页码:948 / 957
页数:10
相关论文
共 19 条
[1]   AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams [J].
Almond, PR ;
Biggs, PJ ;
Coursey, BM ;
Hanson, WF ;
Huq, MS ;
Nath, R ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (09) :1847-1870
[2]   Initial Clinical Experience Treating Patients with Breast Cancer on a 6-MV Flattening-Filter-Free O-Ring Linear Accelerator [J].
Barsky, Andrew R. ;
O'Grady, Fionnbarr ;
Kennedy, Christopher ;
Taunk, Neil K. ;
Dong, Lei ;
Metz, James M. ;
Li, Taoran ;
Freedman, Gary M. .
ADVANCES IN RADIATION ONCOLOGY, 2019, 4 (04) :571-578
[3]   Commissioning and dosimetric characteristics of TrueBeam system: Composite data of three TrueBeam machines [J].
Chang, Zheng ;
Wu, Qiuwen ;
Adamson, Justus ;
Ren, Lei ;
Bowsher, James ;
Yan, Hui ;
Thomas, Andrew ;
Yin, Fang-Fang .
MEDICAL PHYSICS, 2012, 39 (11) :6981-7018
[4]  
Constantin M, 2018, AAPM ANN M JUL 31 NA
[5]   Calculation of enhanced dynamic wedge factors for symmetric and asymmetric photon fields [J].
Gibbons, JP .
MEDICAL PHYSICS, 1998, 25 (08) :1411-1418
[6]   Whole Breast Irradiation with Halcyon (TM) 2.0: Workflow and Efficiency of Field-in-Field Treatment with Dynamic Beam Flattening Technique and kV Cone Beam Computed Tomography [J].
Kennedy, Chris ;
Freedman, Gary ;
Taunk, Neil ;
Scheuermann, Ryan ;
Dong, Lei ;
Metz, James M. ;
Li, Taoran .
CUREUS, 2018, 10 (10)
[7]  
Lim T, 2018, AAPM ANN M JUL 31 NA
[8]   Characterization of the HalcyonTM multileaf collimator system [J].
Lim, Tze Yee ;
Dragojevic, Irena ;
Hoffman, David ;
Flores-Martinez, Everardo ;
Kim, Gwe-Ya .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (04) :106-114
[9]   Characterizing output for the Varian enhanced dynamic wedge field [J].
Liu, CR ;
Li, ZF ;
Palta, JR .
MEDICAL PHYSICS, 1998, 25 (01) :64-70
[10]   Measuring dose distributions for enhanced dynamic wedges using a multichamber detector array [J].
Liu, HH ;
Lief, EP ;
McCullough, EC .
MEDICAL PHYSICS, 1997, 24 (09) :1515-1519