Calculation of Coincidence Summing Correction Factors for an HPGe detector using GEANT4

被引:13
作者
Giubrone, G. [1 ]
Ortiz, J. [1 ]
Gallardo, S. [2 ]
Martorell, S.
Bas, M. C. [3 ]
机构
[1] Univ Politecn Valencia, Lab Radiactividad Ambiental, E-46022 Valencia, Spain
[2] Univ Politecn Valencia, Inst Univ Seguridad Ind Radiofis & Medioambiental, Camino de Vera S-N, E-46022 Valencia, Spain
[3] Univ Politecn Valencia, Dept Estadist & Invest Operat Aplicadas & Calidad, E-46022 Valencia, Spain
关键词
Coincidence Summing; HPGe detector; Monte Carlo; GEANT4; GAMMA-RAY SPECTROMETRY; SIMULATION; CODE;
D O I
10.1016/j.jenvrad.2016.04.008
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The aim of this paper was to calculate the True Coincidence Summing Correction Factors (TSCFs) for an HPGe coaxial detector in order to correct the summing effect as a result of the presence of Y-88 and Co-60 in a multigamma source used to obtain a calibration efficiency curve. Results were obtained for three volumetric sources using the Monte Carlo toolkit, GEANT4. The first part of this paper deals with modeling the detector in order to obtain a simulated full energy peak efficiency curve. A quantitative comparison between the measured and simulated values was made across the entire energy range under study. The True Summing Correction Factors were calculated for Y-88 and Co-60 using the full peak efficiencies obtained with GEANT4. This methodology was subsequently applied to Cs-134, and presented a complex decay scheme. (C) 2016 Elsevier Ltd. All rights reserved.
引用
收藏
页码:114 / 118
页数:5
相关论文
共 50 条
[41]   Coincidence summing correction factors for 238U and 232Th decay series using the Monte Carlo method [J].
Ordonez, J. ;
Gallardo, S. ;
Ortiz, J. ;
Martorell, S. .
RADIATION PHYSICS AND CHEMISTRY, 2019, 155 :244-247
[42]   Parallel tally geometry in GEANT4 for medical dose calculation [J].
Aso, T. ;
Apostolakis, J. ;
Asai, M. ;
Cosmo, G. ;
Daquino, G. ;
Kimura, A. ;
Koi, T. ;
Sasaki, T. ;
Verderi, M. .
2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, :978-980
[43]   Corrections of HPGe detector efficiency curve due to true coincidence summing by program EFFTRAN and by Monte Carlo simulations [J].
Knezevic, J. ;
Mrdja, D. ;
Hansman, J. ;
Bikit, K. ;
Forkapic, S. ;
Bikit, I. ;
Velimirovic, D. ;
Kuzmanovic, P. .
APPLIED RADIATION AND ISOTOPES, 2022, 189
[44]   Nuclear criticality calculations using GEANT4 [J].
Seghour, A. .
ANNALS OF NUCLEAR ENERGY, 2021, 164
[45]   Calculation of Response Function for Bonner Sphere Spectrometer Based on Geant4 [J].
曹靖 ;
蒋小菲 ;
蒋春雨 ;
曹宏睿 ;
阴泽杰 .
Plasma Science and Technology, 2015, 17 (01) :80-83
[46]   Calculation of Response Function for Bonner Sphere Spectrometer Based on Geant4 [J].
Cao Jing ;
Jiang Xiaofei ;
Jiang Chunyu ;
Cao Hongrui ;
Yin Zejie .
PLASMA SCIENCE & TECHNOLOGY, 2015, 17 (01) :80-83
[47]   Beta-gamma coincidence counting efficiency and energy resolution optimization by Geant4 Monte Carlo simulations for a phoswich well detector [J].
Zhang, Weihua ;
Mekarski, Pawel ;
Ungar, Kurt .
APPLIED RADIATION AND ISOTOPES, 2010, 68 (12) :2377-2381
[48]   Evaluating cosmic-ray backgrounds in HPGe gamma spectrometry: Experimental observations and GEANT4 simulations [J].
Sun, Zaijing ;
Nangeelil, Krishnakumar Divakar ;
Pak, Sungmin .
APPLIED RADIATION AND ISOTOPES, 2025, 218
[49]   Evaluation of Effective Parameters on a Simple Muon Detector Efficiency Using GEANT4 Monte Carlo Model [J].
Z. Khalifeh ;
S. Z. Islami rad .
Physics of Particles and Nuclei Letters, 2022, 19 :235-240
[50]   Evaluation of Effective Parameters on a Simple Muon Detector Efficiency Using GEANT4 Monte Carlo Model [J].
Khalifeh, Z. ;
Rad, S. Z. Islami .
PHYSICS OF PARTICLES AND NUCLEI LETTERS, 2022, 19 (03) :235-240