Determination of 99Tc in fresh water using TRU resin by ICP-MS

被引:16
|
作者
Guerin, Nicolas [1 ]
Riopel, Remi [1 ]
Kramer-Tremblay, Sheila [1 ]
de Silva, Nimal [1 ]
Cornett, Jack [1 ]
Dai, Xiongxin [1 ]
机构
[1] CNL, Radiol Protect Res & Instrumentat Branch, Chalk River, ON K0J 1J0, Canada
关键词
Technetium-99; Rhenium; Inductively coupled plasma mass spectrometry; TRU resin; Anion exchange resin; Environmental water samples; ENVIRONMENTAL-SAMPLES; VAPOR-PRESSURES; TECHNETIUM; RHENIUM; SEPARATION; EXTRACTION; HEPTOXIDE; LOBSTERS; TC;
D O I
10.1016/j.aca.2017.08.013
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Technetium-99 (Tc-99) determination at trace level by inductively coupled plasma mass spectrometry (ICP-MS) is challenging because there is no readily available appropriate Tc isotopic tracer. A new method using Re as a recovery tracer to determine Tc-99 in fresh water samples, which does not require any evaporation step, was developed. Tc(VII) and Re(VII) were pre-concentrated on a small anion exchange resin (AER) cartridge from one litre of water sample. They were then efficiently eluted from the AER using a potassium permanganate (KMnO4) solution. After the reduction of KMnO4 in 2 M sulfuric acid solution, the sample was passed through a small TRU resin cartridge. Tc(VII) and Re(VII) retained on the TRU resin were eluted using near boiling water, which can be directly used for the ICP-MS measurement. The results for method optimisation, validation and application were reported. Crown Copyright (C) 2017 Published by Elsevier B.V. All rights reserved.
引用
收藏
页码:114 / 120
页数:7
相关论文
共 50 条
  • [1] 99Tc detection in water samples by ICP-MS
    Mas, JL
    García-León, M
    Bolívar, JP
    RADIOCHIMICA ACTA, 2004, 92 (01) : 39 - 46
  • [2] Determination of 99Tc in environmental samples and depleted uranium penetrators using ICP-MS
    Sahli, Hans
    Rollin, Stefan
    Alvarado, Jose A. Corcho
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2017, 311 (03) : 1633 - 1642
  • [3] Determination of 99Tc in environmental samples and depleted uranium penetrators using ICP-MS
    Hans Sahli
    Stefan Röllin
    José A. Corcho Alvarado
    Journal of Radioanalytical and Nuclear Chemistry, 2017, 311 : 1633 - 1642
  • [4] Capillary electrophoresis as sample introduction system for highly sensitive and interference-free determination of 99Tc by ICP-MS
    Zhou, Dan
    Bao, Liangjin
    Long, Haoqi
    Zhou, Duo
    Xu, Yuwei
    Wang, Bo
    Xia, Chuanqin
    Xian, Liang
    Zheng, Chengbin
    CHINESE CHEMICAL LETTERS, 2025, 36 (04)
  • [5] Automation of 99Tc extraction by LOV prior ICP-MS detection: Application to environmental samples
    Rodriguez, Rogelio
    Leal, Luz
    Miranda, Silvia
    Ferrer, Laura
    Avivar, Jessica
    Garcia, Ariel
    Cerda, Victor
    TALANTA, 2015, 133 : 88 - 93
  • [6] A comparison study on the use of Dowex 1 and TEVA-resin in determination of 99Tc in environmental and nuclear coolant samples in a SIA system with ICP-MS detection
    Kolacinska, Kamila
    Samczynski, Zbigniew
    Dudek, Jakub
    Bojanowska-Czajka, Anna
    Trojanowicz, Marek
    TALANTA, 2018, 184 : 527 - 536
  • [7] Preconcentration and separation of 99Tc in groundwater by using TEVA resin
    Zhang, Zi-lu
    Zhou, Guo-qing
    Lin, Jian-feng
    Ma, Yue
    Yi, Xiao-wei
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2017, 314 (01) : 161 - 166
  • [8] Accurate technetium-99 determination using the combination of TEVA resin pretreatment and ICP-MS measurement and its influence on the Tc-99/Cs-137 scaling factor calculation
    Chen, Lian-Song
    Wang, Tsing-Hai
    Hsieh, Yi-Kong
    Jian, Li-Wei
    Chen, Wie-Han
    Tsai, Tsuey-Lin
    Wang, Chu-Fang
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2014, 299 (03) : 1883 - 1889
  • [9] Preconcentration and separation of 99Tc in groundwater by using TEVA resin
    Zi-lu Zhang
    Guo-qing Zhou
    Jian-feng Lin
    Yue Ma
    Xiao-wei Yi
    Journal of Radioanalytical and Nuclear Chemistry, 2017, 314 : 161 - 166
  • [10] Rapid determination of 99Tc in water samples using Ti(OH)3-TcO2 co-precipitation and TK200 resin by liquid scintillation counting
    Yuan, Ni
    An, Quan
    Xing, Shan
    Dai, Xiongxin
    Hou, Xiaolin
    Yang, Yonggang
    Ma, Yan
    JOURNAL OF ENVIRONMENTAL RADIOACTIVITY, 2022, 251-252