A rare complication of carotid artery stenting: Displacement of marker ring causing locking of stent and incomplete stent expansion

被引:4
作者
Akgul, E. [1 ]
Asungur, E. H.
Korur, K.
Aikirnbaev, K.
Yaliniz, H.
机构
[1] Cukurova Univ, Fac Med, Dept Radiol, TR-01330 Adana, Turkey
[2] Cukurova Univ, Fac Med, Dept Cardiovasc Surg, TR-01330 Adana, Turkey
关键词
D O I
10.3174/ajnr.A0562
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
We present a rare case of a complication of placement of a carotid artery stent represented by partial opening of a carotid Wallstent caused by displacement of its metal ring marker, which thus hindered complete expansion of the stent. An intraluminally locked carotid stent necessitated referral of the patient for urgent carotid endarterectomy. A possible reason of this unusual complication can be a manufacturing defect, which, to our knowledge, was not documented previously in open public data bases or on the Internet.
引用
收藏
页码:1403 / 1404
页数:2
相关论文
共 10 条
[1]   Complication with an embolic protection device during carotid angioplasty [J].
Cardaioli, P ;
Giordan, M ;
Panfili, M ;
Chioin, R .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 62 (02) :234-236
[2]  
Castriota F, 2002, J ENDOVASC THER, V9, P786, DOI 10.1583/1545-1550(2002)009<0786:IOCPDO>2.0.CO
[3]  
2
[4]   Percutaneous carotid artery stenting: a strategy in evolution [J].
Kovacic, JC ;
Roy, PR ;
Baron, DW ;
Gunalingam, B ;
Muller, DWM .
INTERNAL MEDICINE JOURNAL, 2005, 35 (03) :143-150
[5]   Carotid artery stenting [J].
Liistro, F ;
Di Mario, C .
HEART, 2003, 89 (08) :944-948
[6]  
Mathew Boban, 2005, Cardiol Rev, V13, P197, DOI 10.1097/01.crd.0000137257.00371.a6
[7]   Carotid artery angioplasty with stenting and postprocedure hypotension [J].
Park, B ;
Shapiro, D ;
Dahn, M ;
Arici, M .
AMERICAN JOURNAL OF SURGERY, 2005, 190 (05) :691-695
[8]  
Rabe Kasja, 2004, J Interv Cardiol, V17, P417, DOI 10.1111/j.1540-8183.2004.04085.x
[9]   Favorable early outcome of carotid artery stenting without protection devices [J].
Sztriha, LK ;
Vörös, E ;
Sas, K ;
Szentgyörgyi, R ;
Pócsik, A ;
Barzó, P ;
Szikra, P ;
Makai, A ;
Szólics, A ;
Elek, P ;
Rudas, L ;
Vécsei, L .
STROKE, 2004, 35 (12) :2862-2866
[10]  
2006, MED NEWS TODAY 1101