Feasibility study of multi-purpose quality assurance phantom for pretreatment verification of volumetric modulated arc therapy

被引:6
作者
Won, H. S. [1 ,2 ]
Chung, J. B. [1 ]
Eom, K. Y. [1 ]
Hwang, D. G. [2 ]
Kang, S. W. [3 ,4 ]
Suh, T. S. [3 ,4 ]
机构
[1] Seoul Natl Univ, Dept Radiat Oncol, Bungdang Hosp, Seongnam, South Korea
[2] Sangji Univ, Dept Biomed Engn, Wonju, South Korea
[3] Catholic Univ Korea, Coll Med, Dept Biomed Engn, Seoul, South Korea
[4] Catholic Univ Korea, Coll Med, Res Inst Biomed Engn, Seoul, South Korea
来源
INTERNATIONAL JOURNAL OF RADIATION RESEARCH | 2018年 / 16卷 / 03期
关键词
Volumetric modulated arc therapy; phantom; absolute dosimetry; relative dosimetry; POINT DOSE MEASUREMENT; PINPOINT ION-CHAMBER; HIGH-ENERGY PHOTON; RADIATION-THERAPY; WATER PHANTOM; EBT2; FILM; IMRT; DOSIMETRY; RADIOTHERAPY; BEAMS;
D O I
10.18869/acadpub.ijrr.16.2.279
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The purpose of this study was to evaluate the feasibility of a multi-purpose quality assurance (QA) phantom for pretreatment verification of volumetric modulated arc therapy (VMAT). Materials and Methods: The QA phantom was constructed with polymethyl methacrylate (PMMA) to perform relative dosimetry using EBT3 film and MapCHECK, as well as absolute dosimetry using an ionization chamber. The QA phantom was constructed to perform relative dosimetry using EBT3 film and MapCHECK, as well as the absolute dosimetry using ionization chamber. In order to verify the pretreatment plans, 25 patients treated with VMAT were selected. The pretreatment plans were calculated in the Eclipse treatment planning system using the Acuros XB dose calculation algorithm and CT images for the QA phantom, with the same beam setup and monitor units (MUs) as those for patient treatment. All plans were delivered to the Varian TrueBeam accelerator equipped with a high-definition multi-leaf collimator. Results: The multi-purpose QA phantom is developed for convenient VMAT dose verification. By using the QA phantom, all 25 cases passed +/- 3% acceptability criteria in absolute dosimetry with an ionization chamber for pretreatment verification. The relative dosimetry using EBT3 film and MapCHECK system also showed high agreement of more than 90% for 2%/2-mm and 3%/3-mm criteria. Conclusion The results of this study demonstrated the good multi-purpose capabilities of the phantom for the absolute and relative dosimetry. Therefore, the developed multi-purpose QA phantom was applied in our institution for routine VMRT dose verification.
引用
收藏
页码:279 / 287
页数:9
相关论文
共 38 条
[1]   Response to high-energy photons of PTW31014 PinPoint ion chamber with a central aluminum electrode [J].
Agostinelli, S. ;
Garelli, S. ;
Piergentili, M. ;
Foppiano, F. .
MEDICAL PHYSICS, 2008, 35 (07) :3293-3301
[2]   AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams [J].
Almond, PR ;
Biggs, PJ ;
Coursey, BM ;
Hanson, WF ;
Huq, MS ;
Nath, R ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (09) :1847-1870
[3]   Planning, delivery, and quality assurance of intensity-modulated radiotherapy using dynamic multileaf collimator: A strategy for large-scale implementation for the treatment of carcinoma of the prostate [J].
Burman, C ;
Chui, CS ;
Kutcher, G ;
Leibel, S ;
Zelefsky, M ;
LoSasso, T ;
Spirou, S ;
Wu, QW ;
Yang, J ;
Stein, J ;
Mohan, R ;
Fuks, Z ;
Ling, CC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 39 (04) :863-873
[4]   Evaluation of the magnitude of EBT Gafchromic film polarization effects [J].
Butson, M. J. ;
Cheung, T. ;
Yu, P. K. N. .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2009, 32 (01) :21-25
[5]   3D DVH-based metric analysis versus per-beam planar analysis in IMRT pretreatment verification [J].
Carrasco, Pablo ;
Jornet, Nuria ;
Latorre, Artur ;
Eudaldo, Teresa ;
Ruiz, Agusti ;
Ribas, Montserrat .
MEDICAL PHYSICS, 2012, 39 (08) :5040-5049
[6]  
Chang Kyung Hwan, 2016, [PROGRESS in MEDICAL PHYSICS, 의학물리], V27, P111
[7]   Comparison of Dosimetric Performance among Commercial Quality Assurance Systems for Verifying Pretreatment Plans of Stereotactic Body Radiotherapy Using Flattening-Filter-Free Beams [J].
Chung, Jin-Beom ;
Kang, Sang-Won ;
Eom, Keun-Yong ;
Song, Changhoon ;
Choi, Kyoung-Sik ;
Suh, Tae Suk .
JOURNAL OF KOREAN MEDICAL SCIENCE, 2016, 31 (11) :1742-1748
[8]   Absorption spectra time evolution of EBT-2 model GAFCHROMIC™ film [J].
Devic, Slobodan ;
Aldelaijan, Saad ;
Mohammed, Huriyyah ;
Tomic, Nada ;
Liang, Li-Heng ;
DeBlois, Francois ;
Seuntjens, Jan .
MEDICAL PHYSICS, 2010, 37 (05) :2207-2214
[9]   Patient-specific point dose measurement for IMRT monitor unit verification [J].
Dong, L ;
Antolak, J ;
Salehpour, M ;
Forster, K ;
O'Neill, L ;
Kendall, R ;
Rosen, I .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (03) :867-877
[10]   IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 [J].
Ezzell, Gary A. ;
Burmeister, Jay W. ;
Dogan, Nesrin ;
LoSasso, Thomas J. ;
Mechalakos, James G. ;
Mihailidis, Dimitris ;
Molineu, Andrea ;
Palta, Jatinder R. ;
Ramsey, Chester R. ;
Salter, Bill J. ;
Shi, Jie ;
Xia, Ping ;
Yue, Ning J. ;
Xiao, Ying .
MEDICAL PHYSICS, 2009, 36 (11) :5359-5373