Evaluation of methods to derive blood flow velocity from 1000 fps high-speed angiographic sequences (HSA) using optical flow (OF) and computational fluid dynamics (CFD)

被引:8
作者
Shields, A. [1 ]
Nagesh, S. V. Setlur [1 ]
Ionita, C. [1 ]
Bednarek, D. R. [1 ]
Rudin, S. [1 ]
机构
[1] Univ Buffalo, Canon Stroke & Vasc Res Ctr, Buffalo, NY 14260 USA
来源
MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING | 2021年 / 11595卷
关键词
High-Speed Angiography; Neurointerventional Radiology; Photon-Counting Detectors; Neurovascular Disease; Optical Flow; Computational Fluid Dynamics; CONTRAST;
D O I
10.1117/12.2580881
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Digital Subtraction Angiography (DSA) is considered the gold standard for imaging and guiding treatment of neurovascular lesions, such as cerebral aneurysms and carotid stenoses. Though DSA can show high-resolution morphology, it remains difficult to extract temporal physiological information, because higher frame-rates are necessary to accurately quantify neurovascular flow details. Recent advances in photon-counting detector technology have led us to develop High-Speed Angiography (HSA), where X-ray images are acquired at 1000 fps for more accurate visualization and quantification of blood flow. Blood flow was imaged using HSA under constant flow conditions within various 3D printed patient-specific phantoms. Blood velocity was quantified using an open source Optical Flow algorithm, OpenOpticalFlow, to perform velocity estimation based on the spatio -temporal intensity changes of iodinated contrast wavefronts. The results of these algorithms are then compared with Computational Fluid Dynamics (CFD) simulations, using the same inlet boundary conditions and model geometries. The performance of these algorithms at lower temporal resolution was then also assessed by simulating lower frame rates from the acquired 1000 fps data. It is important to ascertain the hemodynamic effect of abnormal neurovascular conditions, as well as their effect on treatment of such conditions during the actual clinical interventional procedure. While theoretical CFD results requiring considerable computer capability are delayed for hours or more, it is expected that clinical results from multiple HSA sequences will be available almost immediately while the patient is still under treatment, and even right after flow conditions are changed beneficially by the intervention.
引用
收藏
页数:18
相关论文
共 9 条
[1]  
[Anonymous], 2017, J.Open Res. Software, V5, P29, DOI [DOI 10.5334/JORS.168, 10.5334/jors.168]
[2]   A Review of X-Ray Flow Visualization With Applications to Multiphase Flows [J].
Heindel, Theodore J. .
JOURNAL OF FLUIDS ENGINEERING-TRANSACTIONS OF THE ASME, 2011, 133 (07)
[3]   DETERMINING OPTICAL-FLOW [J].
HORN, BKP ;
SCHUNCK, BG .
ARTIFICIAL INTELLIGENCE, 1981, 17 (1-3) :185-203
[4]   The mixability of angiographic contrast with arterial blood [J].
Lieber, Baruch B. ;
Sadasivan, Chander ;
Hao, Qing ;
Seong, Jaehoon ;
Cesar, Liliana .
MEDICAL PHYSICS, 2009, 36 (11) :5064-5078
[5]   Fluid flow and optical flow [J].
Liu, Tianshu ;
Shen, Lixin .
JOURNAL OF FLUID MECHANICS, 2008, 614 :253-291
[6]   A DSA-Based Method Using Contrast-Motion Estimation for the Assessment of the Intra-Aneurysmal Flow Changes Induced by Flow-Diverter Stents [J].
Pereira, V. M. ;
Bonnefous, O. ;
Ouared, R. ;
Brina, O. ;
Stawiaski, J. ;
Aerts, H. ;
Ruijters, D. ;
Narata, A. P. ;
Bijlenga, P. ;
Schaller, K. ;
Lovblad, K. -O. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2013, 34 (04) :808-815
[7]   X-ray videodensitometric methods for blood flow and velocity measurement: A critical review of literature [J].
Shpilfoygel, SD ;
Close, RA ;
Valentino, DJ ;
Duckwiler, GR .
MEDICAL PHYSICS, 2000, 27 (09) :2008-2023
[8]   Recovering estimates of fluid flow from image sequence data [J].
Wildes, RP ;
Amabile, MJ ;
Lanzillotto, AM ;
Leu, TS .
COMPUTER VISION AND IMAGE UNDERSTANDING, 2000, 80 (02) :246-266
[9]   Divergence Compensatory Optical Flow Method for Blood Velocimetry [J].
Yang, Zifeng ;
Yu, Hongtao ;
Huang, George P. ;
Ludwig, Bryan .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2017, 139 (06)