Patient doses for barium meal examination in Serbia and Montenegro and potentials for dose reduction through changes in equipment settings

被引:2
作者
Ciraj, O [1 ]
Markovic, S [1 ]
Kosutic, D [1 ]
机构
[1] VINCA, Inst Nucl Sci, Belgrade 11001, Serbia Monteneg
关键词
D O I
10.1093/rpd/nch531
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Patient doses for barium meal examination performed at three general hospitals, in Serbia and Montenegro were measured using a kerma-area product (KAP) meter. The results were analysed in order to obtain dose-related parameters. Although the observed doses were within the range reported in other studies, intra-hospital and inter-hospital dose variations were significant. Mean KAP values for total examination in three hospitals were 8.4, 24.4 and 13.9 Gy cm(2), respectively. Contribution from fluoroscopy was greater than from radiography. Factors contributing to the increased dose delivery were determined and the recommendations on radiographic techniques were made. Changes, in radiography settings allowed dose reduction up to 48% in the radiographic part of examination, that is. up to 12% in total dose without loss of image quality. In addition, fluoroscopy time was noted as the second major contributor to the dose variations. The results demonstrated the need for standardisation of practice for barium meal examination in the country.
引用
收藏
页码:158 / 163
页数:6
相关论文
共 11 条
  • [1] [Anonymous], 1992, NAT PROT PAT DOS MEA
  • [2] A comparison of analogue and digital techniques in upper gastrointestinal examinations:: absorbed dose and diagnostic quality of the images
    Axelsson, B
    Bodén, K
    Fransson, SG
    Hansson, IB
    Persliden, J
    Witt, HH
    [J]. EUROPEAN RADIOLOGY, 2000, 10 (08) : 1351 - 1354
  • [3] Do CEC guidelines under-utilise the full potential of increasing kVp as a dose-reducing tool?
    Doherty, P
    O'Leary, D
    Brennan, PC
    [J]. EUROPEAN RADIOLOGY, 2003, 13 (08) : 1992 - 1999
  • [4] *EUR COMM, 1997, J EUR COMMUN, P180
  • [5] ESTIMATION OF EFFECTIVE DOSE FROM DOSE-AREA PRODUCT MEASUREMENTS FOR BARIUM MEALS AND BARIUM ENEMAS
    HART, D
    WALL, BF
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (797) : 485 - 489
  • [6] Hart D, 2002, W14 NRPB
  • [7] MACCIA C, 1988, HEALTH PHYS, V54, P155
  • [8] Patient dose from barium procedures
    Ruiz-Cruces, R
    Ruiz, F
    Pérez-Martínez, M
    López, J
    Ausina, IT
    de los Ríos, AD
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2000, 73 (871) : 752 - 761
  • [9] Dose-area product readings for fluoroscopic and plain film examinations, including an analysis of the source of variation for barium enema examinations
    Warren-Forward, HM
    Haddaway, MJ
    Temperton, DH
    McCall, IW
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (849) : 961 - 967
  • [10] Patient doses from barium meal and barium enema examinations and potential for reduction through proper set-up of equipment
    Yakoumakis, E
    Tsalafoutas, IA
    Sandilos, P
    Koulentianos, H
    Kasfiki, A
    Vlahos, L
    Proukakis, C
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1999, 72 (854) : 173 - 178