Laparoscopic Sphincter Augmentation Device Eliminates Reflux Symptoms and Normalizes Esophageal Acid Exposure One- and 2-Year Results of a Feasibility Trial

被引:125
作者
Bonavina, Luigi [1 ]
DeMeester, Tom [2 ]
Fockens, Paul [3 ]
Dunn, Daniel [4 ]
Saino, Greta [1 ]
Bona, Davide [1 ]
Lipham, John [2 ]
Bemelman, Willem [3 ]
Ganz, Robert A. [5 ]
机构
[1] Univ Milan, Dept Surg, IRCCS Policlin San Donato, Sch Med, I-20097 Milan, Italy
[2] Univ So Calif, Keck Sch Med, Dept Surg, Los Angeles, CA 90033 USA
[3] Univ Amsterdam, Acad Med Ctr, NL-1105 AZ Amsterdam, Netherlands
[4] Abbott NW Hosp, Minneapolis, MN 55407 USA
[5] Minnesota Gastroenterol, Minneapolis, MN USA
关键词
GASTROESOPHAGEAL-REFLUX; NISSEN FUNDOPLICATION; DISEASE; FEASIBILITY; IMPEDANCE; SURGERY; GERD;
D O I
10.1097/SLA.0b013e3181fd879b
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objectives: One- and 2-year evaluation of a feasibility trial (clinicaltrials.gov registration numbers NCT01057992, NCT01058070, and 01058564) to assess the safety and efficacy of a laparoscopically implanted sphincter augmentation device for the treatment of gastroesophageal reflux disease (GERD). Methods: A sphincter augmentation device (LINX Reflux Management System; Torax Medical, Shoreview, MN), designed to prevent reflux due to abnormal opening of the lower esophageal sphincter (LES), was laparoscopically implanted at the gastroesophageal junction in 44 patients. At baseline, all patients had abnormal esophageal acid exposure on 24-hour pH monitoring and improved, but persistent, typical GERD symptoms while on acid suppression therapy with proton pump inhibitors (PPIs). The device comprises a miniature string of interlinked titanium beads, with magnetic cores, placed around the gastroesophageal junction. The magnetic bond between adjacent beads augments sphincter competence. The beads temporarily separate to accommodate a swallowed bolus, allow belching or vomiting, and reapproximate to augment the LES in the closed position. Patients were evaluated after surgery by GERD Health-Related Quality of Life symptom score, PPI usage, endoscopy, esophageal manometry, and 24-hour esophageal pH monitoring. Results: The total mean GERD Health-Related Quality of Life symptom scores improved from a mean baseline value of 25.7 to 3.8 and 2.4 at 1- and 2-year follow-up, representing an 85% and 90% reduction, respectively (P < 0.0001). Complete cessation of PPI use was reported by 90% of patients at 1 year and by 86% of patients at 2 years. Early dysphagia occurred in 43% of the patients and self-resolved by 90 days. One device was laparoscopically explanted for persistent dysphagia without disruption of the anatomy or function of the cardia. There were no device migrations, erosions, or induced mucosal injuries. At 1 and 2 years, 77% and 90% of patients had a normal esophageal acid exposure. The mean percentage time pH was less than 4 decreased from a baseline of 11.9% to 3.1% (P < 0.0001) at 1 year and to 2.4% (P < 0.0001) at 2 years. Patient satisfaction was 87% at 1 year and 86% at 2 years. Conclusions: The new laparoscopically implanted sphincter augmentation device eliminates GERD symptoms without creating undue side effects and is effective at 1 and 2 years of follow-up.
引用
收藏
页码:857 / 862
页数:6
相关论文
共 19 条
[1]  
AGA Institute, 2008, GERD PAT STUD PAT TH
[2]  
BONAVINA L, 1997, Esophagus, V10, P115
[3]   Magnetic Augmentation of the Lower Esophageal Sphincter: Results of a Feasibility Clinical Trial [J].
Bonavina, Luigi ;
Saino, Greta I. ;
Bona, Davide ;
Lipham, John ;
Ganz, Robert A. ;
Dunn, Daniel ;
DeMeester, Tom .
JOURNAL OF GASTROINTESTINAL SURGERY, 2008, 12 (12) :2133-2140
[4]   Clinical results of laparoscopic fundoplication at ten years after surgery [J].
Dallemagne, B ;
Weerts, J ;
Markiewicz, S ;
Dewandre, JM ;
Wahlen, C ;
Monami, B ;
Jehaes, C .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2006, 20 (01) :159-165
[5]  
Dean B, 2006, CLIN GASTROENTEROL H, V2, P656
[6]   CLINICAL AND INVITRO ANALYSIS OF DETERMINANTS OF GASTROESOPHAGEAL COMPETENCE - STUDY OF THE PRINCIPLES OF ANTIREFLUX SURGERY [J].
DEMEESTER, TR ;
WERNLY, JA ;
BRYANT, GH ;
LITTLE, AG ;
SKINNER, DB .
AMERICAN JOURNAL OF SURGERY, 1979, 137 (01) :39-46
[7]   Use of a magnetic sphincter for the treatment of GERD: a feasibility study [J].
Ganz, Robert A. ;
Gostout, Christopher J. ;
Grudem, Jerry ;
Swanson, William ;
Berg, Todd ;
DeMeester, Tom R. .
GASTROINTESTINAL ENDOSCOPY, 2008, 67 (02) :287-294
[8]   Laparoscopic fundoplication failures - Patterns of failure and response to fundoplication revision [J].
Hunter, JG ;
Smith, CD ;
Branum, GD ;
Waring, JP ;
Trus, TL ;
Cornwell, M ;
Galloway, K .
ANNALS OF SURGERY, 1999, 230 (04) :595-604
[9]  
Kwok H, 1999, CLIN ANAT, V12, P164, DOI 10.1002/(SICI)1098-2353(1999)12:3<164::AID-CA4>3.3.CO
[10]  
2-E