3D prompt gamma imaging for proton beam range verification

被引:111
作者
Draeger, E. [1 ]
Mackin, D. [2 ]
Peterson, S. [3 ]
Chen, H. [1 ]
Avery, S. [4 ]
Beddar, S. [2 ,5 ]
Polf, J. C. [1 ]
机构
[1] Univ Maryland, Sch Med, Dept Radiat Oncol, 22 South Greene St, Baltimore, MD 21201 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, 1515 Holcombe Blvd, Houston, TX 77030 USA
[3] Univ Cape Town, Dept Phys, ZA-7700 Cape Town, South Africa
[4] Univ Penn, Dept Radiat Oncol, 3400 Civ Ctr Blvd, Philadelphia, PA 19104 USA
[5] Univ Texas Houston, Grad Sch Biomed Sci Houston, 6767 Bertner Ave, Houston, TX 77030 USA
基金
美国国家科学基金会; 美国国家卫生研究院;
关键词
Compton imaging; prompt gamma; proton therapy; range verification; COMPTON CAMERA; RAY EMISSION; THERAPY; RECONSTRUCTION; RADIOTHERAPY; ENERGY; IRRADIATION; SIMULATION; DETECTORS; RADIATION;
D O I
10.1088/1361-6560/aaa203
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We tested the ability of a single Compton camera (CC) to produce 3-dimensional (3D) images of prompt gammas (PGs) emitted during the irradiation of a tissue-equivalent plastic phantom with proton pencil beams for clinical doses delivered at clinical dose rates. PG measurements were made with a small prototype CC placed at three different locations along the proton beam path. We evaluated the ability of the CC to produce images at each location for two clinical scenarios: (1) the delivery of a single 2 Gy pencil beam from a hypo-fractionated treatment (similar to 9 x 10(8) protons), and (2) a single pencil beam from a standard treatment (similar to 1 x 10(8) protons). Additionally, the data measured at each location were combined to simulate measurements with a larger scale, clinical CC and its ability to image shifts in the Bragg peak (BP) range for both clinical scenarios. With our prototype CC, the location of the distal end of the BP could be seen with the CC placed up to 4 cm proximal or distal to the BP distal falloff. Using the data from the simulated full scale clinical CC, 3D images of the PG emission were produced with the delivery of as few as 1 x 10(8) protons, and shifts in the proton beam range as small as 2 mm could be detected for delivery of a 2 Gy spot. From these results we conclude that 3D PG imaging for proton range verification under clinical beam delivery conditions is possible with a single CC.
引用
收藏
页数:13
相关论文
共 46 条
[1]   Fast image reconstruction for Compton camera using stochastic origin ensemble approach [J].
Andreyev, Andriy ;
Sitek, Arkadiusz ;
Celler, Anna .
MEDICAL PHYSICS, 2011, 38 (01) :429-438
[2]  
[Anonymous], 2012, THESIS
[3]   GAMMA RADIATION FROM MEDIUM ENERGY PROTON BOMBARDMENT OF LITHIUM, BERYLLIUM, BORON, CARBON AND NITROGEN [J].
CLEGG, AB ;
SALMON, GL ;
SEGEL, RE ;
FOLEY, KJ .
PROCEEDINGS OF THE PHYSICAL SOCIETY OF LONDON, 1961, 78 (503) :681-&
[4]   Feasibility Studies of a New Event Selection Method to Improve Spatial Resolution of Compton Imaging for Medical Applications [J].
Draeger E. ;
Peterson S. ;
Mackin D. ;
Chen H. ;
Beddar S. ;
Polf J.C. .
IEEE Transactions on Radiation and Plasma Medical Sciences, 2017, 1 (04) :358-367
[5]   A Tracking Compton-Scattering Imaging System for Hadron Therapy Monitoring [J].
Frandes, Mirela ;
Zoglauer, Andreas ;
Maxim, Voichita ;
Prost, Remy .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2010, 57 (01) :144-150
[6]  
Gillam JE, 2011, IEEE NUCL SCI CONF R, P3508, DOI 10.1109/NSSMIC.2011.6152645
[7]   Tests of a Compton imaging prototype in a monoenergetic 4.44MeV photon field - a benchmark setup for prompt gamma-ray imaging devices [J].
Golnik, C. ;
Bemmerer, D. ;
Enghardt, W. ;
Fiedler, F. ;
Hueso-Gonzalez, F. ;
Pausch, G. ;
Roemer, K. ;
Rohling, H. ;
Schoene, S. ;
Wagner, L. ;
Kormoll, T. .
JOURNAL OF INSTRUMENTATION, 2016, 11
[8]   Proton therapy monitoring by Compton imaging: influence of the large energy spectrum of the prompt-γ radiation [J].
Hilaire, Estelle ;
Sarrut, David ;
Peyrin, Francoise ;
Maxim, Voichita .
PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (08) :3127-3147
[9]   Prompt Gamma Rays Detected with a BGO Block Compton Camera Reveal Range Deviations of Therapeutic Proton Beams [J].
Hueso-Gonzalez F. ;
Pausch G. ;
Petzoldt J. ;
Romer K.E. ;
Enghardt W. .
IEEE Transactions on Radiation and Plasma Medical Sciences, 2017, 1 (01) :76-86
[10]   Compton Camera and Prompt Gamma Ray Timing: Two Methods for In Vivo Range Assessment in Proton Therapy [J].
Hueso-Gonzalez, Fernando ;
Fiedler, Fine ;
Golnik, Christian ;
Kormoll, Thomas ;
Pausch, Guntram ;
Petzoldt, Johannes ;
Roemer, Katja E. ;
Enghardt, Wolfgang .
FRONTIERS IN ONCOLOGY, 2016, 6