Uncertainty estimation in 60Co gamma-ray dosimetry at JAERI involving a two-way dose intercomparison study with NPL in the dose range 1-50 kGy

被引:9
作者
Kojima, T
Tachibana, H
Haneda, N
Kawashima, I
Sharpe, PHG
机构
[1] Japan Atom Energy Res Inst, Takasaki, Gumma 3701293, Japan
[2] Radiat Applicat Dev Assoc, Takasaki, Gumma 3701207, Japan
[3] Natl Phys Lab, Teddington TW11 0LW, Middx, England
关键词
Co-60; gamma-ray; absorbed dose; uncertainty; calibration; alanine dosimeters; dichromate dosimeters; dose intercomparison;
D O I
10.1016/S0969-806X(98)00282-5
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The uncertainties in Co-60 gamma-ray dosimetry at JAERI have been estimated by reviewing irradiation and calibration procedures. The uncertainty in dose delivered by the JAERI calibration system has been estimated to be 2.2% (2 sigma) and the uncertainty in dose measurement using JAERI alanine/ESR transfer dosimeters has been estimated to be 3.4% (2 sigma), A two-way dose intercomparison study between NPL and JAERI employing alanine-PS, alanine-paraffin, and dichromate dosimeters as transfer dosimeters was performed to confirm the estimated uncertainty over the dose range 1-50 kGy. Dose intercomparison results using both alanine and dichromate dosimeters show agreement within about 2%, which gives confidence in the estimated uncertainty in Co-60 gamma-ray dosimetry at JAERI. (C) 1999 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:619 / 626
页数:8
相关论文
共 11 条
[1]   FADING CHARACTERISTICS OF IRRADIATED ALANINE PELLETS - THE IMPORTANCE OF PREIRRADATION CONDITIONING [J].
ARBER, JM ;
SHARPE, PHG .
APPLIED RADIATION AND ISOTOPES, 1993, 44 (1-2) :19-22
[2]  
*ASTM, 1997, E140095A ASTM STAND, P729
[3]   PHOTON MASS ATTENUATION AND ENERGY-ABSORPTION COEFFICIENTS FROM 1 KEV TO 20 MEV [J].
HUBBELL, JH .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1982, 33 (11) :1269-1290
[4]  
KOJIMA T, 1992, APPL RADIAT ISOTOPES, V43, P863, DOI 10.1016/0883-2889(92)90147-7
[5]  
KOJIMA T, 1989, APPL RADIAT ISOTOPES, V40, P851, DOI 10.1016/0883-2889(89)90006-3
[6]  
MEHTA KK, 1992, IAEA SSDL NEWSLETT, V31, P31
[7]   Uncertainty of dose measurement in radiation processing [J].
Miller, A .
RADIATION PHYSICS AND CHEMISTRY, 1996, 47 (03) :479-482
[8]   CALCULATION OF PHOTON MASS ENERGY-TRANSFER AND MASS ENERGY-ABSORPTION COEFFICIENTS [J].
SELTZER, SM .
RADIATION RESEARCH, 1993, 136 (02) :147-170
[9]   THE RELATIVE RESPONSE OF FRICKE, DICHROMATE AND ALANINE DOSIMETERS TO CO-60 AND HIGH-ENERGY ELECTRON-BEAM RADIATION [J].
SHARPE, PHG ;
BURNS, DT .
RADIATION PHYSICS AND CHEMISTRY, 1995, 46 (4-6) :1273-1277
[10]   ACIDIC AQUEOUS DICHROMATE SOLUTIONS AS REFERENCE DOSIMETERS IN THE 10-40 KGY RANGE [J].
SHARPE, PHG ;
BARRETT, JH ;
BERKLEY, AM .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1985, 36 (08) :647-652