Outcomes of Solo Smart valve in a single-center experience of 270 patients

被引:1
作者
Liu, Hao [1 ,2 ]
Khani-Hanjani, Abbas [1 ]
Yang, Siyuan [1 ,3 ]
Wang, Wei [1 ]
Sidhu, Surita [4 ]
Mullen, John [1 ]
Modry, Dennis [1 ]
Wang, Shaohua [1 ]
机构
[1] Univ Alberta, Mazankowski Alberta Heart Inst, Div Cardiac Surg, Edmonton, AB, Canada
[2] Shanghai Jiao Tong Univ, Xinhua Hosp, Sch Med, Cardiothorac Dept, Shanghai, Peoples R China
[3] Guiyang Med Coll, Affiliated Hosp, Dept Cardiac Surg, Guiyang, Guizhou, Peoples R China
[4] Univ Alberta, Mazankowski Alberta Heart Inst, Dept Anesthesiol & Pain Med, Edmonton, AB, Canada
关键词
Transcatheter aortic valve replacement; Bioprosthesis; Cardiac surgical procedures; PERICARDIAL STENTLESS VALVE; AORTIC-VALVE; LONG-TERM; PROSPECTIVE MULTICENTER; PACEMAKER IMPLANTATION; CLINICAL IMPACT; FOLLOW-UP; REPLACEMENT; BIOPROSTHESIS; MISMATCH;
D O I
10.23736/S0021-9509.18.10165-0
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND: The Solo Smart pericardial aortic valve has been widely used in Europe as an option for aortic valve replacement (AVR). We are reporting early and midterm clinical outcomes of AVR with the Solo Smart valve in a single North America center. METHODS: This is a retrospective study of 270 consecutive patients who had AVR at Mazankowski Alberta Heart Institute from February 2011 to March 2015. Follow-up and echocardiographic data were collected retrospectively from electronic and paper charts. Univariate and multivariate analysis were performed to evaluate the results. RESULTS: The mean age was 71.2 +/- 10.0 years, 67.4% were male, and 793% had combined procedures. Mean STS Score was 4.18 +/- 3.91. Early mortality was 3.7% for the entire group and 0% for isolated AVR group. Mean cross-clamp time for isolated AVR and AVR with concomitant procedure was 70.8 +/- 12.7 min and 117.0+45.0 min, respectively. Permanent pacemaker implantation was necessary in 2.2% of patients. Echocardiography demonstrated a reduction in mean gradients from 40.8 +/- 17.4 mmHg to 7.6 +/- 3.7 mmHg and peak gradient from 72.5 +/- 48.8 mmHg to 15.5 +/- 7.5 mmHg. The 1-, 3-, and 5-year overall survival was 93.0%, 86.5% and 75.9%, respectively. At 5 years, freedom from valve-related death was 92.4%, freedom from structural valve deterioration and freedom from aortic valve reoperation were 96.4% and 98%. respectively. CONCLUSIONS: The Solo Smart valve is safe and has excellent hemodynamic performance. Aortic valve reoperation and rates of valve-related adverse events during midterm follow-up were low.
引用
收藏
页码:633 / 639
页数:7
相关论文
共 28 条
[1]   Guidelines for reporting mortality and morbidity after cardiac valve interventions [J].
Akins, Cary W. ;
Miller, D. Craig ;
Turina, Marko I. ;
Kouchoukos, Nicholas T. ;
Blackstone, Eugene H. ;
Grunkemeier, Gary L. ;
Takkenberg, Johanna J. M. ;
David, Tirone E. ;
Butchart, Eric G. ;
Adams, David H. ;
Shahian, David M. ;
Hagl, Siegfried ;
Mayer, John E. ;
Lytle, Bruce W. .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2008, 135 (04) :732-738
[2]  
Altintas G, 2013, TEX HEART I J, V40, P50
[3]   The Sorin Freedom SOLO stentless aortic valve: Technique of implantation and operative results in 109 patients [J].
Aymard, Thierry ;
Eckstein, Friedrich ;
Englberger, Lars ;
Stalder, Mario ;
Kadner, Alexander ;
Carrel, Thierry .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2010, 139 (03) :775-777
[4]  
Beholz S, 2010, J HEART VALVE DIS, V19, P115
[5]   Very Long-Term Outcomes of the Carpentier-Edwards Perimount Valve in Aortic Position [J].
Bourguignon, Thierry ;
Bouquiaux-Stablo, Anne-Lorraine ;
Candolfi, Pascal ;
Mirza, Alain ;
Loardi, Claudia ;
May, Marc-Antoine ;
El-Khoury, Rym ;
Marchand, Michel ;
Aupart, Michel .
ANNALS OF THORACIC SURGERY, 2015, 99 (03) :831-837
[6]   Clinical and haemodynamic outcomes in 804 patients receiving the Freedom SOLO stentless aortic valve: results from an international prospective multicentre study [J].
Grubitzsch, Herko ;
Wang, Shaohua ;
Matschke, Klaus ;
Glauber, Mattia ;
Heimansohn, David ;
Tan, Erwin ;
Francois, Katrien ;
Thalmann, Markus .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2015, 47 (03) :E97-E104
[7]   North American trial results at 1 year with the Sorin Freedom SOLO pericardial aortic valve [J].
Heimansohn, David ;
Roselli, Eric E. ;
Thourani, Vinod H. ;
Wang, Shaohua ;
Voisine, Pierre ;
Ye, Jian ;
Dabir, Reza ;
Moon, Michael .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2016, 49 (02) :493-499
[8]   Current Results of Surgical Aortic Valve Replacement: Insights From the German Aortic Valve Registry [J].
Holzhey, David ;
Mohr, Friedrich W. ;
Walther, Thomas ;
Moellmann, Helge ;
Beckmann, Andreas ;
Koetting, Joachim ;
Figulla, Hans Reiner ;
Cremer, Jochen ;
Kuck, Karl-Heinz ;
Lange, Ruediger ;
Sack, Stefan ;
Schuler, Gerhard ;
Beyersdorf, Friedhelm ;
Boehm, Michael ;
Heusch, Gerd ;
Meinertz, Thomas ;
Neumann, Till ;
Papoutsis, Konstantinos ;
Schneider, Steffen ;
Welz, Armin ;
Hamm, Christian W. .
ANNALS OF THORACIC SURGERY, 2016, 101 (02) :658-666
[9]   Single-center experience using the Freedom SOLO aortic bioprosthesis [J].
Iliopoulos, Dimitrios C. ;
Deveja, Aris Rezar ;
Androutsopoulou, Vasiliki ;
Filias, Vasilios ;
Kastelanos, Eleftherios ;
Satratzemis, Vasilios ;
Khalpey, Zain ;
Koudoumas, Dimitrios .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2013, 146 (01) :96-102
[10]   Hemodynamic performance of the Medtronic Mosaic and Perimount Magna aortic bioprostheses: five-year results of a prospectively randomized study [J].
Jose Dalmau, Maria ;
Gonzalez-Santos, Jose Maria ;
Antonio Blazquez, Jose ;
Alfonso Sastre, Jose ;
Lopez-Rodriguez, Javier ;
Bueno, Maria ;
Castano, Mario ;
Arribas, Antonio .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2011, 39 (06) :844-852