Intra-abdominal pressure: The neglected variable in selecting the ventriculoperitoneal shunt for treating hydrocephalus

被引:29
作者
Sahuquillo, Juan [1 ]
Arikan, Fuat [1 ]
Poca, Maria A. [1 ]
Noguer, Montserrat [2 ]
Martinez-Ricarte, Francisco [1 ]
机构
[1] Autonomous Univ Barcelona, Vall D Hebron Univ Hosp, Dept Neurosurg, Neurotraumatol & Neurosurg Res Unit,Vall D Hebro, E-08193 Barcelona, Spain
[2] Autonomous Univ Barcelona, Vall D Hebron Univ Hosp, Dept Anesthesiol, Neurotraumatol & Neurosurg Res Unit,Vall D Hebro, E-08193 Barcelona, Spain
关键词
body mass index; cerebrospinal fluid shunts; hydrocephalus; intra-abdominal pressure; normal pressure hydrocephalus; ventriculoperitoneal shunts;
D O I
10.1227/01.NEU.0000311071.33615.E1
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: In the selection of a ventriculoperitoneal cerebrospinal fluid shunt, the intra-abdominal pressure (IAP) is traditionally neglected as a result of the idea that its value is close to 0 mmHg. Our aim was to explore the relationship between body mass index (BMI) and IAP with the goal of providing clinically relevant data that could help neurosurgeons to estimate IAP and select the appropriate shunt for patients with hydrocephalus and especially those with normal-pressure hydrocephalus syndrome. METHODS: Sixty patients requiring the placement of a ventriculoperitoneal shunt were included in the study. We determined weight, BMI, and IAP. IAP was measured through an intraperitoneal catheter during the shunt surgery. To determine whether a linear relationship existed between quantitative variables, linear regression analysis was used. RESULTS: BMI was 28.1 +/- 4.8 kg/m(2). Eighteen patients (30%) had normal weight, 21 (35%) were moderately overweight, and 21 (35%) were obese. IAP was related to patient BMI. A significant positive linear correlation was identified between BMI and IAP (r = 0.52; P = 0.018) with a slope of 0.31 (P < 0.001) and an intercept of -5.5. CONCLUSION: In our study, we determined that IAP had a strong positive linear relationship with BMI. This correlation was independent of sex. An IAP of 0 mmHg can, therefore, only be assumed for patients with a normal BMI who are recumbent. In obese or overweight patients, neurosurgeons should take IAP into account when selecting both the most adequate differential pressure valve to be implanted and in which distal cavity to place the distal catheter to avoid shunt underdrainage induced by high IAP.
引用
收藏
页码:143 / 149
页数:7
相关论文
共 31 条
[1]   Intraabdominal pressure: A revised method for measurement (vol 186, pg 368, 1998) [J].
Cheatham, ML ;
Safcsak, K .
JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, 1998, 186 (05) :594-595
[2]   Abdominal perfusion pressure: A superior parameter in the assessment of intra-abdominal hypertension [J].
Cheatham, ML ;
White, MW ;
Sagraves, SG ;
Johnson, JL ;
Block, EFJ .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2000, 49 (04) :621-626
[3]   Hydrodynamic properties of hydrocephalus shunts: United Kingdom shunt evaluation laboratory [J].
Czosnyka, M ;
Czosnyka, Z ;
Whitehouse, H ;
Pickard, JD .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1997, 62 (01) :43-50
[4]  
CZOSNYKA M, 1995, MED DELTA VALVE MDA, P1
[5]  
Emerson H, 1911, ARCH INTERN MED, V7, P754
[6]   EFFECT OF CEREBROSPINAL-FLUID SHUNTS ON INTRACRANIAL PRESSURE AND ON CEREBROSPINAL-FLUID DYNAMICS .2. NEW TECHNIQUE OF PRESSURE MEASUREMENTS - RESULTS AND CONCEPTS .3. CONCEPT OF HYDROCEPHALUS [J].
FOX, JL ;
MCCULLOUGH, DC ;
GREEN, RC .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1973, 36 (02) :302-312
[7]  
Gallagher J J, 2000, Crit Care Nurse, V20, P87
[8]   Comparison of different methods for measuring intra-abdominal pressure [J].
Gudmundsson, FF ;
Viste, A ;
Gislason, H ;
Svanes, K .
INTENSIVE CARE MEDICINE, 2002, 28 (04) :509-514
[9]  
HALSTED CH, 2004, HARRISONS PRINCIPLES, P460
[10]  
Harrahill M, 1998, J Emerg Nurs, V24, P465