Analysis of Anatomic periarticular tibial plate fit on normal adults

被引:36
作者
Goyal, Kanu S.
Skalak, Anthony S.
Marcus, Randall E.
Vallier, Heather A.
Cooperman, Daniel R.
机构
[1] Case Western Reserve Univ, Sch Med, Univ Hosp, Med Ctr,Dept Orthopaed, Cleveland, OH 44106 USA
[2] Case Western Reserve Univ, Sch Med, MetroHlth Med Ctr, Dept Orthopaed, Cleveland, OH 44106 USA
关键词
D O I
10.1097/BLO.0b013e31809e6e45
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Implant manufacturers are producing anatomically contoured periarticular plates to improve the treatment of proximal tibia fractures. We assessed the accuracy of the designation anatomic. We applied eight-hole medial and lateral anatomically contoured periarticular plates to 101 cadaveric tibiae. The tibiae and the plate fits were mapped, quantified, and analyzed using a MicroScribe (R) G2LX digitizer, Rhinoceros (R) software, and MATLAB (R) software. By corresponding the clinical appearance of good fit with our digital findings, we created numerical criteria for plate fit in three planes: coronal (volume of free space between the plate and bone), sagittal (alignment with the tibial plateau and shaft), and axial (match in curvature between the proximal horizontal part of the plate and the tibial plateau). An anatomic fit should mirror the shape of the tibia in all three planes, and only four medial and four lateral plate fits qualified. Recognizing and understanding the substantial variations in fit that exist between anatomically contoured plates and the tibia may help lead to a more stable fixation and prevent malreduction of the fracture and/or soft tissue impingement.
引用
收藏
页码:245 / 257
页数:13
相关论文
共 36 条
[1]   Limb bone bilateral asymmetry: variability and commonality among modern humans [J].
Auerbach, BM ;
Ruff, CB .
JOURNAL OF HUMAN EVOLUTION, 2006, 50 (02) :203-218
[2]   Complications associated with internal fixation of high-energy bicondylar tibial plateau fractures utilizing a two-incision technique [J].
Barei, DP ;
Nork, SE ;
Mills, WJ ;
Henley, MB ;
Benirschke, SK .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (10) :649-657
[3]  
Bartel D.L., 2006, Orthopaedic Biomechanics
[4]  
Benirschke S K, 1992, J Orthop Trauma, V6, P78
[5]  
BOLHOFNER BR, 1995, CLIN ORTHOP RELAT R, P75
[6]   Compartment pressures after submuscular fixation of proximal tibia fractures [J].
Cole, PA ;
Zlowodzki, M ;
Kregor, PJ .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2003, 34 :S43-S46
[7]  
Duwelius PJ, 1997, CLIN ORTHOP RELAT R, P47
[8]  
ECKHOFF DG, 1994, CLIN ORTHOP RELAT R, P42
[9]   A RETROSPECTIVE ANALYSIS OF PLATE CONTOURING IN THE TIBIA USING THE CONVENTIONAL 4.5 (NARROW) DYNAMIC COMPRESSION PLATE [J].
FRANKLE, MA ;
CORDEY, J ;
FRANKLE, MD ;
BAUMGART, F ;
PERREN, S .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1994, 8 (01) :59-63
[10]  
Gaudinez RF, 1996, CLIN ORTHOP RELAT R, P203