Inspection of endoscope instrument channels after reprocessing using a prototype borescope

被引:74
作者
Thaker, Adarsh M. [1 ]
Kim, Stephen [1 ]
Sedarat, Alireza [1 ]
Watson, Rabindra R. [1 ]
Muthusamy, V. Raman [1 ]
机构
[1] Univ Calif Los Angeles, David Geffen Sch Med, Vatche & Tamar Manoukian Div Digest Dis, Los Angeles, CA 90095 USA
关键词
MICROBIAL-CONTAMINATION; DUODENOSCOPES; DISINFECTION; INFECTION;
D O I
10.1016/j.gie.2018.04.2366
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Background and Aims: Visual inspection of the instrument channel has been proposed as a quality assurance step during endoscope reprocessing. However, the nature and severity of findings in a broad array of endoscopes (gastroscopes, colonoscopes, duodenoscopes, and echoendoscopes) after systemic implementation of an inspection protocol remain unknown. In addition, a study using borescope inspection in upper endoscopes and colonoscopes raised concerns about persistent simethicone residue despite full reprocessing. Methods: A pilot inspection study using a prototype borescope (SteriCam Inspection Scope; SteriView Inc, San Rafael, Calif) was performed on routinely used endoscopes after high-level disinfection, manual forced-air dry of the instrument channel, and overnight vertical storage. Video recordings 1.5 to 2 minutes in duration were reviewed for visible moisture, debris, discoloration, scratches, channel shredding (scratches that result in strips or filaments of the channel lining protruding into the lumen), and visible evidence of biofilm or simethicone residue. Results: A total of 97 inspections of 59 endoscopes were reviewed. The most common finding was scratches, seen in 51 devices (86%). Channel shredding was found in 35 devices (59%). Intrachannel debris was identified in 22 (23%) of the 97 inspections. No moisture was seen (0%) in the 74 inspections performed after forced-air dry and overnight vertical storage compared with moisture in 5 of 18 inspections (28%) performed after storage alone. No visual evidence of biofilm or simethicone residue was discovered despite its frequent use in our unit. Conclusion: Internal defects of the instrument channel appear to occur frequently. Manual forced-air drying of the channel appears to be highly effective in eliminating moisture compared with overnight hang drying alone. Video inspection of the endoscope channel may be useful to audit reprocessing performance and to identify damaged endoscopes.
引用
收藏
页码:612 / 619
页数:8
相关论文
共 20 条
[1]   IN-HOSPITAL EVALUATION OF CONTAMINATION OF DUODENOSCOPES - A QUANTITATIVE ASSESSMENT OF THE EFFECT OF DRYING [J].
ALFA, MJ ;
SITTER, DL .
JOURNAL OF HOSPITAL INFECTION, 1991, 19 (02) :89-98
[2]   Comparison of Two Culture Methods for Use in Assessing Microbial Contamination of Duodenoscopes [J].
Gazdik, Michaela A. ;
Coombs, Jana ;
Burke, John P. ;
Lopansri, Bert K. .
JOURNAL OF CLINICAL MICROBIOLOGY, 2016, 54 (02) :312-316
[3]   Persistent residual contamination in endoscope channels; a fluorescence epimicroscopy study [J].
Herve, Rodolphe C. ;
Keevil, Charles W. .
ENDOSCOPY, 2016, 48 (07) :609-616
[4]   Superbugs on Duodenoscopes: the Challenge of Cleaning and Disinfection of Reusable Devices [J].
Humphries, Romney M. ;
McDonnell, Gerald .
JOURNAL OF CLINICAL MICROBIOLOGY, 2015, 53 (10) :3118-3125
[5]   Current Practice of Duodenoscope Reprocessing [J].
Kim S. ;
Muthusamy V.R. .
Current Gastroenterology Reports, 2016, 18 (10)
[6]   Endoscope drying and its pitfalls [J].
Kovaleva, J. .
JOURNAL OF HOSPITAL INFECTION, 2017, 97 (04) :319-328
[7]   Transmission of Infection by Flexible Gastrointestinal Endoscopy and Bronchoscopy [J].
Kovaleva, Julia ;
Peters, Frans T. M. ;
van der Mei, Henny C. ;
Degener, John E. .
CLINICAL MICROBIOLOGY REVIEWS, 2013, 26 (02) :231-254
[8]   Inconsistencies in endoscope-reprocessing and infection-control guidelines: The importance of endoscope drying [J].
Muscarella, Lawrence F. .
AMERICAN JOURNAL OF GASTROENTEROLOGY, 2006, 101 (09) :2147-2154
[9]   Impact of ethylene oxide gas sterilization of duodenoscopes after a carbapenem-resistant Enterobacteriaceae outbreak [J].
Naryzhny, Igor ;
Silas, Dean ;
Chi, Kenneth .
GASTROINTESTINAL ENDOSCOPY, 2016, 84 (02) :259-262
[10]   Effectiveness of current disinfection procedures against biofilm on contaminated GI endoscopes [J].
Neves, Marcelo S. ;
da Silva, Marlei Gomes ;
Ventura, Grasiella M. ;
Cortes, Patricia Barbur ;
Duarte, Rafael Silva ;
de Souza, Heitor S. .
GASTROINTESTINAL ENDOSCOPY, 2016, 83 (05) :944-953