Reduction of total knee replacement wear with vitamin E blended highly cross-linked ultra-high molecular weight polyethylene

被引:24
作者
Vaidya, C.
Alvarez, E.
Vinciguerra, J. [2 ]
Bruce, D. A.
DesJardins, J. D. [1 ]
机构
[1] Clemson Univ, Dept Bioengn, Clemson, SC 29634 USA
[2] DJO Surg, Austin, TX USA
关键词
wear; TKR; highly cross-linked UHMWPE with vitamin E; force-controlled simulator; GUR; 1020; MECHANICAL-PROPERTIES; JOINT REPLACEMENTS; IRRADIATED UHMWPE; ALPHA-TOCOPHEROL; IN-VITRO; RESISTANCE; OXIDATION; SIMULATOR; BEHAVIOR; LINKING;
D O I
10.1243/09544119JEIM774
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Ultra-high molecular weight polyethylene (UHMWPE) is a common bearing component in total knee replacement (TKR) implants, and its susceptibility to wear continues to be the long-term limiting factor in the life of these implants. This study hypothesized that in TKR systems, a highly cross-linked (HXL) UHMWPE blended with vitamin E will result in reduced wear as compared to a direct compression-moulded (DCM) UHMWPE. A wear simulation study was conducted using an asymmetric lateral pivoting '3D Knee' design to compare the two inserts. The highly cross-linked UHMWPE was aged prior to the testing and force-controlled wear testing was carried out for 5 million cycles using a load-controlled ISO-14243 standard at a frequency of 1 Hz on both groups. Gravimetric measurements of DCM UHMWPE (4.4 +/- 3.0 mg/million cycles) and HXL UHMWPE with vitamin E (1.9 +/- 1.9 mg/million cycles) showed significant statistical differences (p < 0.01) between the wear rates. Wear modes and surface roughness for both groups revealed no significant dissimilarities.
引用
收藏
页码:1 / 7
页数:7
相关论文
共 23 条
  • [1] Wear behaviour of cross-linked polyethylene assessed in vitro under severe conditions
    Affatato, S
    Bersaglia, G
    Rocchi, M
    Taddei, P
    Fagnano, C
    Toni, A
    [J]. BIOMATERIALS, 2005, 26 (16) : 3259 - 3267
  • [2] *ASTM, 1996, ASTM STP, V1272, P163
  • [3] Costa L., 2006, J BONE JOINT SURG BR, V88-B, P46
  • [4] A review of metallic, ceramic and surface-treated metals used for bearing surfaces in human joint replacements
    Dearnley, PA
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1999, 213 (H2) : 107 - 135
  • [5] The use of a force-controlled dynamic knee simulator to quantify the mechanical performance of total knee replacement designs during functional activity
    DesJardins, JD
    Walker, PS
    Haider, H
    Perry, J
    [J]. JOURNAL OF BIOMECHANICS, 2000, 33 (10) : 1231 - 1242
  • [6] DESJARDINS JD, 2007, P 53 ANN M ORTH RES, P1785
  • [7] A direct comparison of patient and force-controlled simulator total knee replacement kinematics
    DesJardins, John D.
    Banks, Scott A.
    Benson, Lisa C.
    Pace, Thomas
    LaBerge, Martine
    [J]. JOURNAL OF BIOMECHANICS, 2007, 40 (15) : 3458 - 3466
  • [8] GREER KW, 1998, T 44 ORTHOP RES SOC, V23, P52
  • [9] HASTINGS RS, 1999, T SOC BIOMATER, V22, P328
  • [10] *INT ORG STAND, 1991, 142431 ISOCD