Measurement of Hemoglobin A1c from Filter Papers for Population-Based Studies

被引:23
作者
Egier, David A. [1 ]
Keys, Judy L. [1 ]
Hall, S. Kim [1 ]
McQueen, Matthew J. [1 ,2 ,3 ]
机构
[1] Hamilton Gen Hosp, CRCTL, Hamilton, ON L8L 2X2, Canada
[2] Populat Hlth Res Inst, Hamilton, ON, Canada
[3] McMaster Univ, Dept Pathol & Mol Med, Hamilton, ON, Canada
关键词
RAD VARIANT-II; HBA(1C); COMPLICATIONS; MANAGEMENT; SAMPLES; BLOOD; TIME; KIT; GHB;
D O I
10.1373/clinchem.2010.156380
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
BACKGROUND: Stability and transport challenges make hemoglobin (Hb) A(1c) measurement from EDTA whole blood (WB) inconvenient and costly for large-scale population studies. This study investigated Hb A(1c) measurement from WB blotted on filter paper (FP) in a Level I National Glycohemoglobin Standardization Program (NGSP)-accredited laboratory. METHODS: Three Bio-Rad Variant (TM) II HPLC instruments and WB and FP specimens were used. Precision, accuracy, linearity, and readable total area of the 6.5min (beta-thalassemia method) Variant II HbA(2)/HbA(1c) Dual Program were assessed. Hb A(1c) stability was measured using in-house FP QC samples. The INTERHEART (a study of the effect of potentially modifiable risk factors associated with myocardial infarction in 52 countries) and CURE (Clopidogrel in Unstable Angina to Prevent Recurrent Events) studies provided chromatographs for morphometric analyses and interoperator variability experiments. Statistical analyses were performed to assess long-term sample stability, WB vs FP agreement, and significance of Hb A(1c) peak integration. RESULTS: Intra-and interassay CVs were <= 2.00%. Total area counts between 0.8 and 5.5 X 10(6) mu V/s produced accurate Hb A(1c) results. The regression equation for agreement between WB(x) and FP(y) was as follows: y = 0.933x + 0.4 (n = 85). FP QC samples stored at -70 degrees C and tested over approximately 3 years yielded CVs of 1.72%-2.73% and regression equations with slopes of -1.08 X 10(-4) to 7.81 X 10-(4). The CURE study, with better preanalytical preparation, achieved a 97% reportable rate, and the reportable rate of the INTERHEART study was 85%. CONCLUSIONS: The FP collection method described provided accurate, robust, and reproducible measurement of Hb A(1c) using the Bio-Rad Variant II HPLC autoanalyzer when FP specimens were prepared according to standardized protocols, and analyses were performed in an NGSP-certified laboratory, supporting the use of FP collection cards in large multinational studies. (C) 2011 American Association for Clinical Chemistry
引用
收藏
页码:577 / 585
页数:9
相关论文
共 31 条
[1]  
Anjali, 2007, J Assoc Physicians India, V55, P115
[2]  
[Anonymous], IDF DIAB ATL
[3]  
Chandalia HB, 2002, CURR SCI INDIA, V83, P1522
[4]   The effect of intensive glycemic treatment on coronary artery calcification in type 1 diabetic participants of the diabetes control and complications Trial/Epidemiology of diabetes interventions and complications (DCCT/EDIC) study [J].
Cleary, Patricia A. ;
Orchard, Trevor J. ;
Genuth, Saul ;
Wong, Nathan D. ;
Detrano, Robert ;
Backlund, Jye-Yu C. ;
Zinman, Bernard ;
Jacobson, Alan ;
Sun, Wanjie ;
Lachin, John M. ;
Nathan, David M. .
DIABETES, 2006, 55 (12) :3556-3565
[5]  
Eckfeldt JH, 1997, CLIN CHEM, V43, P1811
[6]   RED-CELL AGE-RELATED-CHANGES OF HEMOGLOBINS AIA+B AND AIC IN NORMAL AND DIABETIC SUBJECTS [J].
FITZGIBBONS, JF ;
KOLER, RD ;
JONES, RT .
JOURNAL OF CLINICAL INVESTIGATION, 1976, 58 (04) :820-824
[7]   HbA1c measurements from dried blood spots: validation and patient satisfaction [J].
Fokkema, M. Rebecca ;
Bakker, Andries J. ;
de Boer, Fokje ;
Kooistra, Jeltsje ;
de Vries, Sifra ;
Wolthuis, Albert .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2009, 47 (10) :1259-1264
[8]   Tests of glycemia in diabetes [J].
Goldstein, DE ;
Little, RR ;
Lorenz, RA ;
Malone, JI ;
Nathan, D ;
Peterson, CM ;
Sacks, DB .
DIABETES CARE, 2004, 27 (07) :1761-1773
[9]   Tentative identification of hemoglobin variants in the Bio-Rad VARIANT II Hb A1C method [J].
Higgins, TN ;
Ridley, B .
CLINICAL BIOCHEMISTRY, 2005, 38 (03) :272-277
[10]   Analytical evaluation of the Bio-Rad Variant II automated HbA1c analyzer [J].
Higgins, TN ;
Blakney, GB ;
Dayton, J .
CLINICAL BIOCHEMISTRY, 2001, 34 (05) :361-365