Surface applicator calibration and commissioning of an electronic brachytherapy system for nonmelanoma skin cancer treatment

被引:29
作者
Rong, Yi [1 ,2 ]
Welsh, James S. [1 ,2 ]
机构
[1] Univ Wisconsin, Dept Human Oncol, Sch Med & Publ Hlth, Madison, WI 53792 USA
[2] Univ Wisconsin, Canc Ctr Riverview, Riverview Hosp Assoc, Wisconsin Rapids, WI 54494 USA
关键词
nonmelanoma skin cancer; basal cell carcinoma; squamous cell carcinoma; electronic brachytherapy; surface applicator; DOSE-RATE BRACHYTHERAPY; RADIOTHERAPY; IRRADIATION; THERAPY; CARCINOMAS; DOSIMETRY;
D O I
10.1118/1.3489379
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The Xoft Axxent (R) x-ray source has been used for treating nonmelanoma skin cancer since the surface applicators became clinically available in 2009. The authors report comprehensive calibration procedures for the electronic brachytherapy (eBx) system with the surface applicators. Methods: The Xoft miniature tube (model S700) generates 50 kVp low-energy x rays. The new surface applicators are available in four sizes of 10, 20, 35, and 50 mm in diameter. The authors' tests include measurements of dose rate, air-gap factor, output stability, depth dose verification, beam flatness and symmetry, and treatment planning with patient specific cutout factors. The TG-61 in-air method was used as a guideline for acquiring nominal dose-rate output at the skin surface. A soft x-ray parallel-plate chamber (PTW T34013) and electrometer was used for the output commissioning. GafChromic (R) EBT films were used for testing the properties of the treatment fields with the skin applicators. Solid water slabs were used to verify the depth dose and cutout factors. Patients with basal cell or squamous cell carcinoma were treated with eBx using a calibrated Xoft system with the low-energy x-ray source and the skin applicators. Results: The average nominal dose-rate output at the skin surface for the 35 mm applicator is 1.35 Gy/min with +/- 5% variation for 16 sources. The dose-rate output and stability (within +/- 5% variation) were also measured for the remaining three applicators. For the same source, the output variation is within 2%. The effective source-surface distance was calculated based on the air-gap measurements for four applicator sizes. The field flatness and symmetry are well within 5%. Percentage depth dose in water was provided by factory measurements and can be verified using solid water slabs. Treatment duration was calculated based on the nominal dose rate, the prescription fraction size, the depth dose percentage, and the cutout factor. The output factor needs to be measured for each case with varying shapes of cutouts. Conclusions: Together with TG-61, the authors' methodology provides comprehensive calibration procedures for medical physicists for using the Xoft eBx system and skin applicators for nonmelanoma skin cancer treatments. (C) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3489379]
引用
收藏
页码:5509 / 5517
页数:9
相关论文
共 23 条
[1]  
Aldrich J E, 1992, Med Dosim, V17, P69
[2]  
[Anonymous], 2008, Cancer facts and figures
[3]   DOSIMETRIC PRECISION REQUIREMENTS IN RADIATION-THERAPY [J].
BRAHME, A .
ACTA RADIOLOGICA ONCOLOGY, 1984, 23 (05) :379-391
[4]   A dosimetric comparison of Xoft Axxent Electronic Brachytherapy and iridium-192 high-dose-rate brachytherapy in the treatment of endometrial cancer [J].
Dickler, Adam ;
Kirk, Michael C. ;
Coon, Alan ;
Bernard, Damian ;
Zusag, Tom ;
Rotmensch, Jacob ;
Wazer, David E. .
BRACHYTHERAPY, 2008, 7 (04) :351-354
[5]   A dosimetric comparison of MammoSite high-dose-rate brachytherapy and Xoft Axxent electronic brachytherapy [J].
Dickler, Adam ;
Kirk, Michael C. ;
Seif, Neil ;
Griem, Katherine ;
Dowlatshahi, Kambiz ;
Francescatti, Darius ;
Abrams, Ross A. .
BRACHYTHERAPY, 2007, 6 (02) :164-168
[6]  
FREEMAN RG, 1964, CANCER-AM CANCER SOC, V17, P535, DOI 10.1002/1097-0142(196404)17:4<535::AID-CNCR2820170415>3.0.CO
[7]  
2-P
[8]  
Gauden S, 2008, BRACHYTHERAPY, V7, ps159
[9]   Treatment of skin carcinomas of the face by high-dose-rate brachytherapy and custom-made surface molds [J].
Guix, B ;
Finestres, F ;
Tello, JI ;
Palma, C ;
Martinez, A ;
Guix, JR ;
Guix, R .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 47 (01) :95-102
[10]   Chemotherapeutic agents and the skin: An update [J].
Heidary, Noushin ;
Naik, Haley ;
Burgin, Susan .
JOURNAL OF THE AMERICAN ACADEMY OF DERMATOLOGY, 2008, 58 (04) :545-570