Four-dimensional magnetic resonance imaging (4D-MRI) using image-based respiratory surrogate: A feasibility study

被引:158
|
作者
Cai, Jing [1 ]
Chang, Zheng [1 ]
Wang, Zhiheng [1 ]
Segars, William Paul [2 ]
Yin, Fang-Fang [1 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiat Oncol, Durham, NC 27710 USA
[2] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
关键词
4D-MRI; 4D-CT; tumor motion; respiratory motion; cine-MRI; COMPUTED-TOMOGRAPHY; MOTION; RADIOTHERAPY; XCAT; MOBILITY; PHANTOM;
D O I
10.1118/1.3658737
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Four-dimensional computed tomography (4D-CT) has been widely used in radiation therapy to assess patient-specific breathing motion for determining individual safety margins. However, it has two major drawbacks: low soft-tissue contrast and an excessive imaging dose to the patient. This research aimed to develop a clinically feasible four-dimensional magnetic resonance imaging (4D-MRI) technique to overcome these limitations. Methods: The proposed 4D-MRI technique was achieved by continuously acquiring axial images throughout the breathing cycle using fast 2D cine-MR imaging, and then retrospectively sorting the images by respiratory phase. The key component of the technique was the use of body area (BA) of the axial MR images as an internal respiratory surrogate to extract the breathing signal. The validation of the BA surrogate was performed using 4D-CT images of 12 cancer patients by comparing the respiratory phases determined using the BA method to those determined clinically using the Real-time position management (RPM) system. The feasibility of the 4D-MRI technique was tested on a dynamic motion phantom, the 4D extended Cardiac Torso (XCAT) digital phantom, and two healthy human subjects. Results: Respiratory phases determined from the BA matched closely to those determined from the RPM: mean (+/- SD) difference in phase: -3.9% (+/- 6.4%); mean (+/- SD) absolute difference in phase: 10.40% (+/- 3.3%); mean (+/- SD) correlation coefficient: 0.93 (+/- 0.04). In the motion phantom study, 4D-MRI clearly showed the sinusoidal motion of the phantom; image artifacts observed were minimal to none. Motion trajectories measured from 4D-MRI and 2D cine-MRI (used as a reference) matched excellently: the mean (6SD) absolute difference in motion amplitude: -0.3 (+/- 0.5) mm. In the 4D-XCAT phantom study, the simulated "4D-MRI" images showed good consistency with the original 4D-XCAT phantom images. The motion trajectory of the hypothesized "tumor" matched excellently between the two, with a mean (+/- SD) absolute difference in motion amplitude of 0.5 (+/- 0.4) mm. 4D-MRI was able to reveal the respiratory motion of internal organs in both human subjects; superior-inferior (SI) maximum motion of the left kidney of Subject #1 and the diaphragm of Subject #2 measured from 4D-MRI was 0.88 and 1.32 cm, respectively. Conclusions: Preliminary results of our study demonstrated the feasibility of a novel retrospective 4D-MRI technique that uses body area as a respiratory surrogate. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3658737]
引用
收藏
页码:6384 / 6394
页数:11
相关论文
共 32 条
  • [21] A post-processing method based on interphase motion correction and averaging to improve image quality of 4D magnetic resonance imaging: a clinical feasibility study
    Deng, Zixin
    Pang, Jianing
    Lao, Yi
    Bi, Xiaoming
    Wang, Guan
    Chen, Yuhua
    Fenchel, Matthias
    Tuli, Richard
    Li, Debiao
    Yang, Wensha
    Fan, Zhaoyang
    BRITISH JOURNAL OF RADIOLOGY, 2019, 92 (1095)
  • [22] Four-Dimensional Magnetic Resonance Imaging With 3-Dimensional Radial Sampling and Self-Gating-Based K-Space Sorting: Early Clinical Experience on Pancreatic Cancer Patients
    Yang, Wensha
    Fan, Zhaoyang
    Tuli, Richard
    Deng, Zixin
    Pang, Jianing
    Wachsman, Ashley
    Reznik, Robert
    Sandler, Howard
    Li, Debiao
    Fraass, Benedick A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (05): : 1136 - 1143
  • [23] Original Four-dimensional computed tomography (4D-CT) for preoperative parathyroid localization: A good study but are we using it?
    Kuo, Lindsay E.
    Bird, Sarah H.
    Lubitz, Carrie C.
    Pandian, T. K.
    Parangi, Sareh
    Stephen, Antonia E.
    AMERICAN JOURNAL OF SURGERY, 2022, 223 (04) : 694 - 698
  • [24] Comparison of gross target volumes based on four-dimensional CT, positron emission tomography-computed tomography, and magnetic resonance imaging in thoracic esophageal cancer
    Li, Huimin
    Li, Fengxiang
    Li, Jianbin
    Zhu, Youzhe
    Zhang, Yingjie
    Guo, Yanluan
    Xu, Min
    Shao, Qian
    Liu, Xijun
    CANCER MEDICINE, 2020, 9 (15): : 5353 - 5361
  • [25] Porcine lung phantom-based validation of estimated 4D-MRI using orthogonal cine imaging for low-field MR-Linacs
    Rabe, Moritz
    Paganelli, Chiara
    Riboldi, Marco
    Bondesson, David
    Joerg Schneider, Moritz
    Chmielewski, Thomas
    Baroni, Guido
    Dinkel, Julien
    Reiner, Michael
    Landry, Guillaume
    Parodi, Katia
    Belka, Claus
    Kamp, Florian
    Kurz, Christopher
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (05)
  • [26] Four-dimensional computed tomography-based respiratory-gated whole-abdominal intensity-modulated radiation therapy for ovarian cancer: a feasibility study
    Garsa, A. A.
    Andrade, R. S.
    Heron, D. E.
    Beriwal, S.
    Kim, H.
    Brandner, E.
    Kuo, G.
    Chen, H.
    Gerszten, K.
    Yue, J. N.
    Huq, M. S.
    Lee, J.
    Lalonde, R.
    Wu, A.
    INTERNATIONAL JOURNAL OF GYNECOLOGICAL CANCER, 2007, 17 (01) : 55 - 60
  • [27] Adaptive 4D MR Imaging using navigator-based respiratory signal for MRI-Guided therapy
    Tokuda, Junichi
    Morikawa, Shigehiro
    Haque, Hasnine A.
    Tsukamoto, Tetsuji
    Matsumiya, Kiyoshi
    Liao, Hongen
    Masamune, Ken
    Dohi, Takeyoshi
    MAGNETIC RESONANCE IN MEDICINE, 2008, 59 (05) : 1051 - 1061
  • [28] Flow dynamics of type II endoleaks can determine sac expansion after endovascular aneurysm repair using four-dimensional flow-sensitive magnetic resonance imaging analysis
    Katahashi, Kazuto
    Sano, Masaki
    Takehara, Yasuo
    Inuzuka, Kazunori
    Sugiyama, Masataka
    Alley, Marcus T.
    Takeuchi, Hiroya
    Unno, Naoki
    JOURNAL OF VASCULAR SURGERY, 2019, 70 (01) : 107 - +
  • [29] Three-dimensional physics-based registration of pelvic system using 2D dynamic magnetic resonance imaging slices
    Courtecuisse, H.
    Jiang, Z.
    Mayeur, O.
    Witz, J. F.
    Lecomte-Grosbras, P.
    Cosson, M.
    Brieu, M.
    Cotin, S.
    STRAIN, 2020, 56 (03)
  • [30] Three-dimensional regions-of-interest-based intra-operative four-dimensional soft tissue perfusion imaging using a standard x-ray system with no gantry rotation: A simulation study for a proof of concept
    Taguchi, Katsuyuki
    Sauer, Thomas J.
    Segars, W. Paul
    Frey, Eric C.
    Xu, Jingyan
    Liapi, Eleni
    Stayman, J. Webster
    Hong, Kelvin
    Hui, Ferdinand K.
    Unberath, Mathias
    Du, Yong
    MEDICAL PHYSICS, 2020, 47 (12) : 6087 - 6102