Comparison of the Beacon and Quark indirect calorimetry devices to measure resting energy expenditure in ventilated ICU patients

被引:3
作者
Slingerland-Boot, H. [1 ]
Adhikari, S. [2 ]
Mensink, M. R. [2 ]
van Zanten, A. R. H. [1 ,2 ]
机构
[1] Gelderse Vallei Hosp, Dept Intens Care Med, Willy Brandtlaan 10, NL-6716 RP Ede, Netherlands
[2] Wageningen Univ & Res, Div Human Nutr & Hlth, Wageningen, Netherlands
关键词
Critically ill patients; Intensive care unit (ICU); Mechanical ventilation; Indirect calorimetry (IC); Breath-by-breath analysis; Resting energy expenditure (REE); METABOLIC-RATE; PREDICTIVE EQUATIONS; VALIDATION;
D O I
10.1016/j.clnesp.2022.01.015
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Introduction: Critically ill patients in the Intensive Care Unit (ICU) should receive nutritional support matched to their metabolic needs as both under- and overfeeding energy has been shown to increase mortality. Critical illness can significantly affect metabolism. Consequently, resting energy expenditure (REE) can vary markedly during critical illness. Therefore, indirect calorimetry to estimate REE is recommended to determine energy requirements in individual ICU patients and to guide optimal nutritional support. Currently, the Quark metabolic monitor is considered the gold standard in our ICU, but novel mechanical support devices are also equipped with indirect calorimetry functionalities. This study aimed to evaluate the performance of a currently unevaluated device. Methods: A cross-sectional analysis in mechanically ventilated patients was conducted in a mixed medical-surgical ICU. The primary outcome was a numerical and visual comparison of the performance of the Beacon indirect calorimeter to calculate REE compared to the Quark device using Bland Altman plots. Performance was evaluated using bias, precision, accuracy, and reliability. Secondary analysis included a comparison with REE estimated by predictive equations. Results: Seventy-one measurements were obtained in 27 mechanically ventilated subjects. An underestimation by the Beacon device in calculated REE of -96.2 kcal/day (4.5%) was found. There was a bias towards higher VCO2 and lower VO2 values with Beacon as compared to Quark. The reliability of the Beacon was good, with an absolute intraclass correlation coefficient of 0.897 (95%CI 0.751-0.955; p = 0.000). There was a poor correlation (<0.40) between the separate indirect calorimetry devices and most predictive equations. Only the Faisy predictive equations had good reliability (ICC 0.687, p = 0.002). Conclusions: Beacon indirect calorimetry accurately determined REE in mechanically ventilated critically ill patients compared to the gold standard in our ICU (Quark indirect calorimeter), although confidence intervals were wide. There was low bias and good reliability. On the other hand, predictive equations performed poorly compared to both devices, underestimating the true metabolic needs of mechanically ventilated ICU patients. (C) 2022 The Author(s). Published by Elsevier Ltd on behalf of European Society for Clinical Nutrition and Metabolism.
引用
收藏
页码:370 / 377
页数:8
相关论文
共 34 条
  • [11] Indirect calorimetry: A practical guide for clinicians
    Haugen, Heather A.
    Chan, Lingtak-Neander
    Li, Fanny
    [J]. NUTRITION IN CLINICAL PRACTICE, 2007, 22 (04) : 377 - 388
  • [12] Nutrition therapy in critical illness: a review of the literature for clinicians
    Lambell, Kate J.
    Tatucu-Babet, Oana A.
    Chapple, Lee-anne
    Gantner, Dashiell
    Ridley, Emma J.
    [J]. CRITICAL CARE, 2020, 24 (01)
  • [13] Are patients fed appropriately according to their caloric requirements?
    McClave, SA
    Lowen, CC
    Kleber, MJ
    Nicholson, JF
    Jimmerson, SC
    McConnell, JW
    Jung, LY
    [J]. JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 1998, 22 (06) : 375 - 381
  • [14] Comparison of equations for estimating resting metabolic rate in healthy subjects over 70 years of age
    Melzer, Katarina
    Karsegard, Wronique Laurie
    Genton, Laurence
    Kossovsky, Michel P.
    Kayser, Bengt
    Pichard, Claude
    [J]. CLINICAL NUTRITION, 2007, 26 (04) : 498 - 505
  • [15] A NEW PREDICTIVE EQUATION FOR RESTING ENERGY-EXPENDITURE IN HEALTHY-INDIVIDUALS
    MIFFLIN, MD
    STJEOR, ST
    HILL, LA
    SCOTT, BJ
    DAUGHERTY, SA
    KOH, YO
    [J]. AMERICAN JOURNAL OF CLINICAL NUTRITION, 1990, 51 (02) : 241 - 247
  • [16] Energy expenditure and indirect calorimetry in critical illness and convalescence: current evidence and practical considerations
    Moonen, Hanneke Pierre Franciscus Xaverius
    Beckers, Karin Josephina Hubertina
    van Zanten, Arthur Raymond Hubert
    [J]. JOURNAL OF INTENSIVE CARE, 2021, 9 (01)
  • [17] The clinical evaluation of the new indirect calorimeter developed by the ICALIC project
    Oshima, Taku
    Delsoglio, Marta
    Dupertuis, Yves M.
    Singer, Pierre
    De Waele, Elisabeth
    Veraar, Cecilia
    Heidegger, Claudia-Paula
    Wernermann, Jan
    Wischmeyer, Paul E.
    Berger, Mette M.
    Pichard, Claude
    [J]. CLINICAL NUTRITION, 2020, 39 (10) : 3105 - 3111
  • [18] Reliability of, and Agreement Between, two Breath-by-Breath Indirect Calorimeters at Varying Levels of Inspiratory Oxygen
    Poulsen, Mathias Krogh
    Thomsen, Lars Pilegaard
    Kjaergaard, Soren
    Rees, Stephen Edward
    Karbing, Dan Stieper
    [J]. NUTRITION IN CLINICAL PRACTICE, 2019, 34 (05) : 767 - 774
  • [19] Acute Respiratory Distress Syndrome The Berlin Definition
    Ranieri, V. Marco
    Rubenfeld, Gordon D.
    Thompson, B. Taylor
    Ferguson, Niall D.
    Caldwell, Ellen
    Fan, Eddy
    Camporota, Luigi
    Slutsky, Arthur S.
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2012, 307 (23): : 2526 - 2533
  • [20] Measuring energy expenditure in the intensive care unit: a comparison of indirect calorimetry by E-sCOVX and Quark RMR with Deltatrac II in mechanically ventilated critically ill patients
    Rehal, Martin Sundstrom
    Fiskaare, Erik
    Tjader, Inga
    Norberg, Ake
    Rooyackers, Olav
    Wernerman, Jan
    [J]. CRITICAL CARE, 2016, 20