Comparison of the Beacon and Quark indirect calorimetry devices to measure resting energy expenditure in ventilated ICU patients

被引:3
作者
Slingerland-Boot, H. [1 ]
Adhikari, S. [2 ]
Mensink, M. R. [2 ]
van Zanten, A. R. H. [1 ,2 ]
机构
[1] Gelderse Vallei Hosp, Dept Intens Care Med, Willy Brandtlaan 10, NL-6716 RP Ede, Netherlands
[2] Wageningen Univ & Res, Div Human Nutr & Hlth, Wageningen, Netherlands
关键词
Critically ill patients; Intensive care unit (ICU); Mechanical ventilation; Indirect calorimetry (IC); Breath-by-breath analysis; Resting energy expenditure (REE); METABOLIC-RATE; PREDICTIVE EQUATIONS; VALIDATION;
D O I
10.1016/j.clnesp.2022.01.015
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Introduction: Critically ill patients in the Intensive Care Unit (ICU) should receive nutritional support matched to their metabolic needs as both under- and overfeeding energy has been shown to increase mortality. Critical illness can significantly affect metabolism. Consequently, resting energy expenditure (REE) can vary markedly during critical illness. Therefore, indirect calorimetry to estimate REE is recommended to determine energy requirements in individual ICU patients and to guide optimal nutritional support. Currently, the Quark metabolic monitor is considered the gold standard in our ICU, but novel mechanical support devices are also equipped with indirect calorimetry functionalities. This study aimed to evaluate the performance of a currently unevaluated device. Methods: A cross-sectional analysis in mechanically ventilated patients was conducted in a mixed medical-surgical ICU. The primary outcome was a numerical and visual comparison of the performance of the Beacon indirect calorimeter to calculate REE compared to the Quark device using Bland Altman plots. Performance was evaluated using bias, precision, accuracy, and reliability. Secondary analysis included a comparison with REE estimated by predictive equations. Results: Seventy-one measurements were obtained in 27 mechanically ventilated subjects. An underestimation by the Beacon device in calculated REE of -96.2 kcal/day (4.5%) was found. There was a bias towards higher VCO2 and lower VO2 values with Beacon as compared to Quark. The reliability of the Beacon was good, with an absolute intraclass correlation coefficient of 0.897 (95%CI 0.751-0.955; p = 0.000). There was a poor correlation (<0.40) between the separate indirect calorimetry devices and most predictive equations. Only the Faisy predictive equations had good reliability (ICC 0.687, p = 0.002). Conclusions: Beacon indirect calorimetry accurately determined REE in mechanically ventilated critically ill patients compared to the gold standard in our ICU (Quark indirect calorimeter), although confidence intervals were wide. There was low bias and good reliability. On the other hand, predictive equations performed poorly compared to both devices, underestimating the true metabolic needs of mechanically ventilated ICU patients. (C) 2022 The Author(s). Published by Elsevier Ltd on behalf of European Society for Clinical Nutrition and Metabolism.
引用
收藏
页码:370 / 377
页数:8
相关论文
共 34 条
  • [1] BRANSON R D, 1990, Respiratory Care, V35, P640
  • [2] Care Mermaid, BEAC CAR ADV SYST IC
  • [3] Cosmed the Metabolic Company Quark RMR, GOLD STAND MET MEAS
  • [4] De Waele E, 2015, MINERVA ANESTESIOL, V81, P272
  • [5] Assessment of resting energy expenditure in mechanically ventilated patients
    Faisy, C
    Guerot, E
    Diehl, JL
    Labrousse, J
    Fagon, JY
    [J]. AMERICAN JOURNAL OF CLINICAL NUTRITION, 2003, 78 (02) : 241 - 249
  • [6] FAO/WHO/UNU Expert Consultation, EQ PRED BAS MET RAT
  • [7] Cross-Validation of Resting Metabolic Rate Prediction Equations
    Flack, Kyle D.
    Siders, William A.
    Johnson, LuAnn
    Roemmich, James N.
    [J]. JOURNAL OF THE ACADEMY OF NUTRITION AND DIETETICS, 2016, 116 (09) : 1413 - 1422
  • [8] Validation of 2 approaches to predicting resting metabolic rate in critically ill patients
    Frankenfield, D
    Smith, JS
    Cooney, RN
    [J]. JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 2004, 28 (04) : 259 - 264
  • [9] Evaluation of three indirect calorimetry devices in mechanically ventilated patients: Which device compares best with the Deltatrac II®? A prospective observational study
    Graf, Severine
    Karsegard, Veronique Laurie
    Viatte, Valerie
    Heidegger, Claudia Paula
    Fleury, Yvan
    Pichard, Claude
    Genton, Laurence
    [J]. CLINICAL NUTRITION, 2015, 34 (01) : 60 - 65
  • [10] A biometric study of human basal metabolism
    Harris, JA
    Benedict, FG
    [J]. PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1918, 4 : 370 - 373