Distal-end force prediction of tendon-sheath mechanisms for flexible endoscopic surgical robots using deep learning

被引:64
作者
Li, Xiaoguo [1 ]
Cao, Lin [1 ]
Tiong, Anthony Meng Huat [1 ]
Phuoc Thien Phan [1 ]
Phee, Soo Jay [1 ]
机构
[1] Nanyang Technol Univ, Robot Res Ctr, Sch Mech & Aerosp Engn, 50 Nanyang Ave, Singapore 639798, Singapore
基金
新加坡国家研究基金会;
关键词
Haptic feedback; Surgical robot; Tendon sheath mechanisms; Deep learning; Hysteresis; MINIMALLY INVASIVE SURGERY; HAPTIC FEEDBACK; FRICTION MODEL; HYSTERESIS; ELONGATION; SENSOR;
D O I
10.1016/j.mechmachtheory.2018.12.035
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Accurate haptic feedback is highly challenging for flexible endoscopic surgical robots due to space limitation for sensors on small end-effectors and critical force hysteresis of their tendon-sheath mechanisms (TSMs). This paper proposes a deep learning approach to predicting the distal force of TSMs when manipulating a biological tissue based on only proximal-end measurements. Both Multilayer Perceptron (MLP) and Recurrent Neural Network (RNN) were investigated to study their capabilities of making sequential distal force predictions. The results were compared with those of the conventional modelling approach. It was observed that, when sufficient data was provided for training, RNN achieved the most accurate prediction (RMSE = 0.0219 N) in experiments with constant system velocity. The effects of insufficient training data, varying system velocity and irregular motion trajectories on the performance of RNN were further studied. Notably, RNN could precisely identify the current system phase in the force hysteresis profile and can be applied to TSMs with realistic non-periodic movement such as manual manipulation trajectory (RSME = 0.2287 N). The proposed approach can be applied to any TSM-driven robotic systems for accurate haptic feedback without requiring sensors at the distal ends of the robots. (C) 2019 Elsevier Ltd. All rights reserved.
引用
收藏
页码:323 / 337
页数:15
相关论文
共 42 条
[1]  
Akinbiyi Takintope, 2006, Conf Proc IEEE Eng Med Biol Soc, V2006, P567
[2]   Characterization of frictional hysteresis in ball-bearing guideways [J].
Al-Bender, F ;
Symens, W .
WEAR, 2005, 258 (11-12) :1630-1642
[3]   Theoretical analysis of the dynamic behavior of hysteresis elements in mechanical systems [J].
Al-Bender, F ;
Symens, W ;
Swevers, J ;
Van Brussel, H .
INTERNATIONAL JOURNAL OF NON-LINEAR MECHANICS, 2004, 39 (10) :1721-1735
[4]   A SURVEY OF MODELS, ANALYSIS TOOLS AND COMPENSATION METHODS FOR THE CONTROL OF MACHINES WITH FRICTION [J].
ARMSTRONGHELOUVRY, B ;
DUPONT, P ;
DEWIT, CC .
AUTOMATICA, 1994, 30 (07) :1083-1138
[5]  
Bergstra J, 2012, J MACH LEARN RES, V13, P281
[6]  
Dinh BK, 2016, 2016 IEEE/RSJ INTERNATIONAL CONFERENCE ON INTELLIGENT ROBOTS AND SYSTEMS (IROS 2016), P5670, DOI 10.1109/IROS.2016.7759834
[7]   Human-robot skills transfer interfaces for a flexible surgical robot [J].
Calinon, Sylvain ;
Bruno, Danilo ;
Malekzadeh, Milad S. ;
Nanayakkara, Thrishantha ;
Caldwell, Darwin G. .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2014, 116 (02) :81-96
[8]  
Connor J., 1991, IJCNN-91-Seattle: International Joint Conference on Neural Networks (Cat. No.91CH3049-4), P301, DOI 10.1109/IJCNN.1991.155194
[9]   Nonlinear friction modelling and compensation control of hysteresis phenomena for a pair of tendon-sheath actuated surgical robots [J].
Do, T. N. ;
Tjahjowidodo, T. ;
Lau, M. W. S. ;
Phee, S. J. .
MECHANICAL SYSTEMS AND SIGNAL PROCESSING, 2015, 60-61 :770-784
[10]   A new approach of friction model for tendon-sheath actuated surgical systems: Nonlinear modelling and parameter identification [J].
Do, T. N. ;
Tjahjowidodo, T. ;
Lau, M. W. S. ;
Phee, S. J. .
MECHANISM AND MACHINE THEORY, 2015, 85 :14-24