Stationary digital breast tomosynthesis system with a multi-beam fleld emission x-ray source array

被引:22
作者
Yang, Guang [1 ]
Rajaram, Ramya [2 ]
Cao, Guohua [1 ]
Sultana, Shabana [2 ]
Liu, Zhijun [1 ]
Lalush, David [3 ,4 ]
Lu, Jianping [1 ]
Zhou, Otto [1 ,2 ,5 ]
机构
[1] Univ N Carolina, Dept Phys & Astron, Chapel Hill, NC 27599 USA
[2] Univ N Carolina, Curriculum Appl& Mat Sci, Chapel Hill, NC 27599 USA
[3] Univ N Carolina, Joint Dept Biomed Engn, Chapel Hill, NC 27599 USA
[4] North Carolina State Univ, Raleigh, NC 27695 USA
[5] Univ N Carolina, Lineberger Comprehens Canc Ctr, Chapel Hill, NC 27599 USA
来源
MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3 | 2008年 / 6913卷
关键词
digital breast tomosynthesis; carbon nanotubes; field emission;
D O I
10.1117/12.770622
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A stationary digital breast tomosynthesis (DBT) system using a carbon nanotube based multi-beam field emission x-ray (MBFEX) source has been designed. The purpose is to investigate the feasibility of reducing the total imaging time, simplifying the system design, and potentially improving the image quality comparing to the conventional DBT scanners. The MBFEX source consists of 25 individually programmable x-ray pixels which are evenly angular spaced covering a 48 degrees field of view. The device acquires the projection images by electronically switching on and off the individual X-ray pixels without mechanical motion of either the x-ray source or the detector. The designs of the x-ray source and the imaging system are presented. Some preliminary results are discussed.
引用
收藏
页数:10
相关论文
共 15 条
[1]  
[Anonymous], 1961, FIELD EMISSION FIELD
[2]  
Bassett LW, 2005, DIAGNOSIS DIS BREAST
[3]  
Beutel J., 2000, HDB MED IMAGING
[4]   Digital breast tomosynthesis using an amorphous selenium flat panel detector [J].
Bissonnette, M ;
Hansroul, M ;
Masson, E ;
Savard, S ;
Cadieux, S ;
Warmoes, P ;
Gravel, D ;
Agopyan, J ;
Polischuk, B ;
Haerer, W ;
Mertelmeier, T ;
Lo, JY ;
Chen, Y ;
Dobbins, JT ;
Jesneck, JL ;
Singh, S .
Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 :529-540
[5]   Digital x-ray tomosynthesis: current state of the art and clinical potential [J].
Dobbins, JT ;
Godfrey, DJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (19) :R65-R106
[6]  
*EUR COMM STAND, 1999, 125435 EN EUR COMM S
[7]   Accelerated iterative transmission CT reconstruction using an ordered subsets convex algorithm [J].
Kamphuis, C ;
Beekman, FJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (06) :1101-1105
[8]  
Lalush DS, 2006, I S BIOMED IMAGING, P1180
[9]   Carbon nanotube based microfocus field emission x-ray source for microcomputed tomography [J].
Liu, Zejian ;
Yang, Guang ;
Lee, Yueh Z. ;
Bordelon, David ;
Lu, Jianping ;
Zhou, Otto .
APPLIED PHYSICS LETTERS, 2006, 89 (10)
[10]   Analytic method based on identification of ellipse parameters for scanner calibration in cone-beam tomography [J].
Noo, F ;
Clackdoyle, R ;
Mennessier, C ;
White, TA ;
Roney, TJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11) :3489-3508