A method for improving the accuracy of calibration-free laser-induced breakdown spectroscopy (CF-LIBS) using determined plasma temperature by genetic algorithm (GA)

被引:42
|
作者
Dong, Juan [1 ]
Liang, Long [1 ]
Wei, Jiao [1 ]
Tang, Hongsheng [1 ]
Zhang, Tianlong [1 ]
Yang, Xiaofeng [1 ]
Wang, Kang [2 ]
Li, Hua [1 ]
机构
[1] NW Univ Xian, Coll Chem & Mat Sci, Inst Analyt Sci, Xian 710069, Peoples R China
[2] Changan Univ, Coll Sci, Xian 710064, Peoples R China
基金
高等学校博士学科点专项科研基金; 中国国家自然科学基金;
关键词
LOCAL THERMODYNAMIC-EQUILIBRIUM; SELF-ABSORPTION; ELEMENTAL ANALYSIS; EMISSION-LINES; SPECTRAL-LINES; CLASSIFICATION; SELECTION;
D O I
10.1039/c4ja00470a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Accuracy is still a challenge for classical calibration-free laser-induced breakdown spectroscopic (CF-LIBS) quantitative analysis since absolute theoretical calculation and mathematical models cannot compensate for the self-absorption effect and plasma temperature variability. The aim of this research is to obtain a more accurate plasma temperature which contributes to the precise determination of the elemental composition of unknown samples for CF-LIBS. Herein, an internal reference-external standard with iteration correction (IRESIC) method is proposed to correct for the self-absorption effect and plasma temperature in CF-LIBS based on an internal reference line and one standard sample. The spectral intensities of each species are corrected by an internal reference line via the iteration procedure, and the internal reference line suffers from a negligible self-absorption effect for self-absorption correction. Furthermore, one standard sample matrix-matched with the unknown samples along with the genetic algorithm (GA) is utilized to simulate the accurate plasma temperature of the unknown samples. It is worth mentioning that the GA is used for plasma temperature correction through iteration correction for the first time in CF-LIBS. The proposed method demonstrates a significant improvement in accuracy compared with the classical CF-LIBS in the quantitative analysis of aluminum-bronze alloy samples due to the integrated merits of internal reference line usage and accurate plasma temperature evaluation.
引用
收藏
页码:1336 / 1344
页数:9
相关论文
共 50 条
  • [1] Calibration-Free Laser-Induced Breakdown Spectroscopy (CF-LIBS) with Standard Reference Line for the Analysis of Stainless Steel
    Fu, Hongbo
    Dong, Fengzhong
    Wang, Huadong
    Jia, Junwei
    Ni, Zhibo
    APPLIED SPECTROSCOPY, 2017, 71 (08) : 1982 - 1989
  • [2] Correction of self-absorption effect in calibration-free laser-induced breakdown spectroscopy(CF-LIBS) by considering plasma temperature and electron density
    Yang, Yanwei
    Hao, Xiaojian
    Ren, Long
    OPTIK, 2020, 208
  • [3] Accuracy improvement of calibration-free laser-induced breakdown spectroscopy
    Fu, Hongbo
    Ni, Zhibo
    Wang, Huadong
    Jia, Junwei
    Dong, Fengzhong
    PLASMA SCIENCE & TECHNOLOGY, 2019, 21 (03)
  • [4] Correction of self-absorption effect in calibration-free laser-induced breakdown spectroscopy (CF-LIBS) with blackbody radiation reference
    Li, Tianqi
    Hou, Zongyu
    Fu, Yangting
    Yu, Jianlong
    Gu, Weilun
    Wang, Zhe
    ANALYTICA CHIMICA ACTA, 2019, 1058 : 39 - 47
  • [5] Accuracy improvement of calibration-free laser-induced breakdown spectroscopy
    付洪波
    倪志波
    王华东
    贾军伟
    董凤忠
    Plasma Science and Technology, 2019, (03) : 10 - 19
  • [6] A Calibration-Free Laser-Induced Breakdown Spectroscopy (CF-LIBS) Quantitative Analysis Method Based on the Auto-Selection of an Internal Reference Line and Optimized Estimation of Plasma Temperature
    Yang, Jianhong
    Li, Xiaomeng
    Xu, Jinwu
    Ma, Xianghong
    APPLIED SPECTROSCOPY, 2018, 72 (01) : 129 - 140
  • [7] Progress in calibration-free laser-induced breakdown spectroscopy
    Hermann, Jorg
    Gerhard, Christoph
    Burger, Milos
    Craciun, Valentin
    Pelascini, Frederic
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2023, 200
  • [8] Laser-induced breakdown spectroscopy of archaeological findings with calibration-free inverse method: Comparison with classical laser-induced breakdown spectroscopy and conventional techniques
    Gaudiuso, R.
    Dell'Aglio, M.
    De Pascale, O.
    Loperfido, S.
    Mangone, A.
    De Giacomo, A.
    ANALYTICA CHIMICA ACTA, 2014, 813 : 15 - 24
  • [9] A method for improving the accuracy of calibration-free laser-induced breakdown spectroscopy by exploiting self-absorption
    Hu, Zhenlin
    Chen, Feng
    Zhang, Deng
    Chu, Yanwu
    Wang, Weiliang
    Tang, Yun
    Guo, Lianbo
    ANALYTICA CHIMICA ACTA, 2021, 1183
  • [10] Analysis of an Iron-Copper Alloy by Calibration-Free Laser-Induced Breakdown Spectroscopy (CF-LIBS) and Inductively Coupled Plasma - Mass Spectrometry (ICP-MS)
    Ahmed, Nasar
    Shahida, Shabnam
    Kiani, Saba Mujeed
    Razzaq, Muhammad Inzamam
    Hameed, Muhammad Usman
    Iqbal, S. M. Z.
    Abbasi, S. A.
    Rafique, Muhammad
    Baig, Muhammad Aslam
    ANALYTICAL LETTERS, 2022, 55 (14) : 2239 - 2250