Different CAD/CAM-processing routes for zirconia restorations: influence on fitting accuracy

被引:63
作者
Kohorst, Philipp [1 ]
Junghanns, Janet [1 ]
Dittmer, Marc P. [2 ]
Borchers, Lothar [1 ]
Stiesch, Meike [1 ]
机构
[1] Hannover Med Sch, Dept Prosthet Dent & Biomed Mat Sci, D-30625 Hannover, Germany
[2] Hannover Med Sch, Dept Orthodont, D-30625 Hannover, Germany
关键词
CAD/CAM; Processing route; Zirconia; FDP; Fitting accuracy; FIXED PARTIAL DENTURES; LOAD-BEARING CAPACITY; MARGINAL ADAPTATION; CLINICAL FIT; INTERNAL FIT; FINISH LINE; CROWNS; FRAMEWORKS;
D O I
10.1007/s00784-010-0415-9
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The aim of the present in vitro study was to evaluate the influence of different processing routes on the fitting accuracy of four-unit zirconia fixed dental prostheses (FDPs) fabricated by computer-aided design/computer-aided manufacturing (CAD/CAM). Three groups of zirconia frameworks with ten specimens each were fabricated. Frameworks of one group (CerconCAM) were produced by means of a laboratory CAM-only system. The other frameworks were made with different CAD/CAM systems; on the one hand by in-laboratory production (CerconCAD/CAM) and on the other hand by centralized production in a milling center (Compartis) after forwarding geometrical data. Frameworks were then veneered with the recommended ceramics, and marginal accuracy was determined using a replica technique. Horizontal marginal discrepancy, vertical marginal discrepancy, absolute marginal discrepancy, and marginal gap were evaluated. Statistical analyses were performed by one-way analysis of variance (ANOVA), with the level of significance chosen at 0.05. Mean horizontal discrepancies ranged between 22 mu m (CerconCAM) and 58 mu m (Compartis), vertical discrepancies ranged between 63 mu m (CerconCAD/CAM) and 162 mu m (CerconCAM), and absolute marginal discrepancies ranged between 94 mu m (CerconCAD/CAM) and 181 mu m (CerconCAM). The marginal gap varied between 72 mu m (CerconCAD/CAM) and 112 mu m (CerconCAM, Compartis). Statistical analysis revealed that, with all measurements, the marginal accuracy of the zirconia FDPs was significantly influenced by the processing route used (p < 0.05). Within the limitations of this study, all restorations showed a clinically acceptable marginal accuracy; however, the results suggest that the CAD/CAM systems are more precise than the CAM-only system for the manufacture of four-unit FDPs.
引用
收藏
页码:527 / 536
页数:10
相关论文
共 44 条
[1]   MARGINAL ADAPTATION OF THREE DIFFERENT ZIRCONIUM DIOXIDE THREE-UNIT FIXED DENTAL PROSTHESES [J].
Att, Wael ;
Komine, Futoshii ;
Gerds, Thomas ;
Strub, Joerg Rudolf .
JOURNAL OF PROSTHETIC DENTISTRY, 2009, 101 (04) :239-247
[2]   Marginal and internal fits of fixed dental prostheses zirconia retainers [J].
Beuer, Florian ;
Aggstaller, Hans ;
Edelhoff, Daniel ;
Gernet, Wolfgang ;
Sorensen, John .
DENTAL MATERIALS, 2009, 25 (01) :94-102
[3]   Precision of fit: zirconia three-unit fixed dental prostheses [J].
Beuer, Florian ;
Naumann, Michael ;
Gernet, Wolfgang ;
Sorensen, John A. .
CLINICAL ORAL INVESTIGATIONS, 2009, 13 (03) :343-349
[4]  
Bindl A, 2007, INT J PERIODONT REST, V27, P567
[5]   Clinical fit of Procera AllCeram crowns [J].
Boening, KW ;
Wolf, BH ;
Schmidt, AE ;
Kästner, K ;
Walter, MH .
JOURNAL OF PROSTHETIC DENTISTRY, 2000, 84 (04) :419-424
[6]  
Bornemann G, 2002, Int J Comput Dent, V5, P177
[7]  
de Oyagüe RC, 2009, AM J DENT, V22, P79
[8]  
Coli P, 2004, INT J PROSTHODONT, V17, P577
[9]  
Coli P, 2004, INT J PROSTHODONT, V17, P59
[10]   Influence of Cervical Finish Line Type on the Marginal Adaptation of Zirconia Ceramic Crowns [J].
Comlekoglu, M. ;
Dundar, M. ;
Ozcan, M. ;
Gungor, M. ;
Gokce, B. ;
Artunc, C. .
OPERATIVE DENTISTRY, 2009, 34 (05) :586-592