SAR characteristics of the Sigma-60-Ellipse applicator

被引:21
作者
Fatehi, Daryoush [1 ,2 ]
Van Rhoon, Gerard C. [1 ]
机构
[1] Shahrekord Univ Med Sci, Dept Med Phys, Shahrekord, Iran
[2] Erasmus MC, Dept Radiat Oncol, Unit Hyperthermia, Dr Daniel Den Hoed Canc Ctr, Rotterdam, Netherlands
关键词
SAR; Sigma-60-Ellipse applicator; Schottky diode sheet; phantom; quality assurance; hyperthermia;
D O I
10.1080/02656730701832326
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To characterize the basic performance of the Sigma-60-Ellipse applicator. Materials and methods: The E-field distributions were measured using Schottky diode sheets in a cylindrical phantom (diameter 26 cm, length 50 cm), filled with saline-water (2 g NaCl/L). The phantom was positioned symmetrically in the Sigma-60-Ellipse applicator. The stability of the SAR distribution was assessed as a function of power and frequency. Furthermore, the accuracy of target steering was evaluated at various frequencies. Finally, the SAR characteristics were compared with those of the Sigma-60 and the Sigma-Eye applicators. Results: The average 50% iso-SAR area increased from 241 to 296 cm(2) when the RF power increased from 100 to 1600 W. The SAR maximum was located in the centre of the applicator for the frequencies of 75-80 MHz and it moves towards the feet for higher frequencies (up to 3.5 cm at 120 MHz). The average 50% iso-SAR area decreased from 268 to 161 cm(2) With increasing frequency from 75 to 120 MHz. The 50% iso-SAR longitudinal length was almost stable (mean 21.3 cm) at 75-120 MHz for both power outputs of 400 and 800 W. As expected the 50% iso-SAR radial length decreased with frequency from 14.9 cm at 75 MHz to 8.4 cm at 120 MHz. There was a fair agreement between requested and measured target settings. At the lower frequencies of 75-90 MHz and at 100 MHz the SAR characteristics were almost identical to those of the Sigma-60 and Sigma-Eye applicators, respectively. Conclusion: At the frequency range of 75-90 MHz the Sigma-60 and at 100 MHz the Sigma-Eye can safely replaced by the Sigma-60-Ellipse applicator.
引用
收藏
页码:347 / 356
页数:10
相关论文
共 21 条
[1]   A method for the quantitative evaluation of SAR distribution in deep regional hyperthermia [J].
Baroni, C ;
Giri, MG ;
Meliadó, G ;
Maluta, S ;
Chierego, G .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2001, 17 (05) :369-381
[2]  
*BSD 2000 FUNCT OP, 2004, BSD MED CORP REP, P22
[3]   DEEP REGIONAL HYPERTHERMIA - COMPARISON BETWEEN THE ANNULAR PHASED-ARRAY AND THE SIGMA-60 APPLICATOR IN THE SAME PATIENTS [J].
FELDMANN, HJ ;
MOLLS, M ;
KRUMPELMANN, S ;
STUSCHKE, M ;
SACK, H .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 26 (01) :111-116
[4]   FINITE-ELEMENT SIMULATION OF SIGMA-60 HEATING IN THE UTAH PHANTOM - COMPUTED AND MEASURED DATA COMPARED [J].
JIA, X ;
PAULSEN, KD ;
BUECHLER, DN ;
GIBBS, FA ;
MEANEY, PM .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1994, 10 (06) :755-774
[5]  
Kaatee RSJP, 1999, INT J HYPERTHER, V15, P441, DOI 10.1080/026567399285620
[6]   EVALUATION OF THE SIGMA-60 APPLICATOR FOR REGIONAL HYPERTHERMIA IN TERMS OF SCATTERING PARAMETERS [J].
LEYBOVICH, LB ;
MYERSON, RJ ;
EMAMI, B ;
STRAUBE, WL .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1991, 7 (06) :917-935
[8]  
TURNER P, 2007, COMMUNICATION
[9]  
Turner P F, 1988, Recent Results Cancer Res, V107, P249
[10]   Comparison of radiotherapy alone with radiotherapy plus hyperthermia in locally advanced pelvic tumours:: a prospective, randomised, multicentre trial [J].
van der Zee, J ;
González, DG ;
van Rhoon, GC ;
van Dijk, JDP ;
van Putten, WLJ ;
Hart, AAM .
LANCET, 2000, 355 (9210) :1119-1125