Fast enantioseparations of basic analytes by high-performance liquid chromatography using cellulose tris(3,5-dimethylphenylcarbamate)-coated zirconia stationary phases

被引:35
作者
Castells, CB [1 ]
Carr, PW [1 ]
机构
[1] Univ Minnesota, Dept Chem, Minneapolis, MN 55455 USA
基金
美国国家卫生研究院;
关键词
chiral stationary phases; LC; enantiomer separation; mobile-phase composition; beta-blockers; basic compounds; cellulose tris(3,5-dimethylphenylcarbamate);
D O I
10.1016/S0021-9673(00)00883-9
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In this work, we study the influence of the mobile phase and column temperature on the enantioresolution of basic compounds on microparticulate porous zirconia coated with cellulose tris(3,5-dimethylphenylcarbamate) (CDMPC). The chiral analytes are amino compounds, including a number of P-blockers. Analytes are eluted with hexane-alcohol mobile phases. We investigated the effect of alcohol (type and concentration), basic eluent additives, and column temperature on the parameters that control resolution (column efficiency, retention and selectivity). Conditions for achieving an adequate separation in the least time have been determined for numerous racemic mixtures. For most solutes, baseline resolution of the enantiomeric pair was achieved in less than 1 min; 12 of 13 pairs were separated in less than 2 min. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:17 / 33
页数:17
相关论文
共 32 条