Pseudo-CT Generation for MRI-Only Radiation Therapy Treatment Planning: Comparison Among Patch-Based, Atlas-Based, and Bulk Density Methods

被引:37
作者
Largent, Axel [1 ]
Barateau, Anais [1 ]
Nunes, Jean-Claude [1 ]
Lafond, Caroline [1 ]
Greer, Peter B. [2 ]
Dowling, Jason A. [3 ]
Saint-Jalmes, Herve [1 ]
Acosta, Oscar [1 ]
de Crevoisier, Renaud [1 ]
机构
[1] Univ Rennes, CLCC Eugene Marquis, Rennes, France
[2] Univ Newcastle, Sch Math & Phys Sci, Newcastle, NSW, Australia
[3] CSIRO, Australian E Hlth Res Ctr, Herston, Qld, Australia
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2019年 / 103卷 / 02期
关键词
COMPUTED-TOMOGRAPHY GENERATION; RADIOTHERAPY; IMAGE; VERIFICATION; HEAD;
D O I
10.1016/j.ijrobp.2018.10.002
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Methods have been recently developed to generate pseudo-computed to-mography (pCT) for dose calculation in magnetic resonance imaging (MRI)-only radiation therapy. This study aimed to propose an original nonlocal mean patch-based method (PBM) and to compare this PBM to an atlas-based method (ABM) and to a bulk density method (BDM) for prostate MRI-only radiation therapy. Materials and Methods: Thirty-nine patients received a volumetric modulated arc therapy for prostate cancer. In addition to the planning computed tomography (CT) scans, T2-weighted MRI scans were acquired. pCTs were generated from MRIs using 3 methods: an original nonlocal mean PBM, ABM, and BDM. The PBM was performed using feature extraction and approximate nearest neighbor search in a training cohort. The PBM accuracy was evaluated in a validation cohort by using imaging and dosimetric endpoints. Imaging endpoints included mean absolute error and mean error between Hounsfield units of the pCT and the reference CT (CTref). Dosimetric end-points were based on dose-volume histograms calculated from the CTref and the pCTs for various volumes of interest and on 3-dimensional gamma analyses. The PBM uncertainties were compared with those of the ABM and BDM. Results: The mean absolute error and mean error obtained from the PBM were 41.1 and -1.1 Hounsfield units. The PBM dose-volume histogram differences were 0.7% for prostate planning target volume V-95%, 0.5% for rectum V-70Gy, and 0.2% for bladder V-50Gy. Compared with ABM and BDM, PBM provided significantly lower dose uncertainties for the prostate planning target volume (70-78 Gy), the rectum (8.5-29 Gy, 40-48 Gy, and 61-73 Gy), and the bladder (12-78 Gy). The PBM mean gamma pass rate (99.5%) was significantly higher than that of ABM (94.9%) or BDM (96.1%). Conclusions: The proposed PBM provides low uncertainties with dose planned on CTref. These uncertainties were smaller than those of ABM and BDM and are unlikely to be clinically significant. (C) 2018 Elsevier Inc. All rights reserved.
引用
收藏
页码:479 / 490
页数:12
相关论文
共 48 条
[1]   A patch-based pseudo-CT approach for MRI-only radiotherapy in the pelvis [J].
Andreasen, Daniel ;
Van Leemput, Koen ;
Edmund, Jens M. .
MEDICAL PHYSICS, 2016, 43 (08) :4742-4752
[2]   Patch-based generation of a pseudo CT from conventional MRI sequences for MRI-only radiotherapy of the brain [J].
Andreasen, Daniel ;
Van Leemput, Koen ;
Hansen, Rasmus H. ;
Andersen, Jon A. L. ;
Edmund, Jens M. .
MEDICAL PHYSICS, 2015, 42 (04) :1596-1605
[3]   Generation of synthetic CT using multi-scale and dual-contrast patches for brain MRI-only external beam radiotherapy [J].
Aouadi, Souha ;
Vasic, Ana ;
Paloor, Satheesh ;
Torfeh, Tarraf ;
McGarry, Maeve ;
Petric, Primoz ;
Riyas, Mohamed ;
Hammoud, Rabih ;
Al-Hammadi, Noora .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 42 :174-184
[4]   70 GY VERSUS 80 GY IN LOCALIZED PROSTATE CANCER: 5-YEAR RESULTS OF GETUG 06 RANDOMIZED TRIAL [J].
Beckendorf, Veronique ;
Guerif, Stephane ;
Le Prise, Elisabeth ;
Cosset, Jean-Marc ;
Bougnoux, Agnes ;
Chauvet, Bruno ;
Salem, Naji ;
Chapet, Olivier ;
Bourdain, Sylvain ;
Bachaud, Jean-Marc ;
Maingon, Philippe ;
Hannoun-Levi, Jean-Michel ;
Malissard, Luc ;
Simon, Jean-Marc ;
Pommier, Pascal ;
Hay, Men ;
Dubray, Bernard ;
Lagrange, Jean-Leon ;
Luporsi, Elisabeth ;
Bey, Pierre .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 80 (04) :1056-1063
[5]  
Burgos Ninon, 2016, Medical Image Computing and Computer-Assisted Intervention - MICCAI 2016. 19th International Conference. Proceedings: LNCS 9901, P547, DOI 10.1007/978-3-319-46723-8_63
[6]   Robust CT Synthesis for Radiotherapy Planning: Application to the Head and Neck Region [J].
Burgos, Ninon ;
Cardoso, M. Jorge ;
Guerreiro, Filipa ;
Veiga, Catarina ;
Modat, Marc ;
McClelland, Jamie ;
Knopf, Antje-Christin ;
Punwani, Shonit ;
Atkinson, David ;
Arridge, Simon R. ;
Hutton, Brian F. ;
Ourselin, Sebastien .
MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2015, PT II, 2015, 9350 :476-484
[7]   MRI-based treatment planning for radiotherapy: Dosimetric verification for prostate IMRT [J].
Chen, LL ;
Price, RA ;
Wang, L ;
Li, JS ;
Qin, LH ;
McNeeley, S ;
Ma, CMC ;
Freedman, GM ;
Pollack, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (02) :636-647
[8]   Feasibility and limitations of bulk density assignment in MRI for head and neck IMRT treatment planning [J].
Chin, Alexander L. ;
Lin, Alexander ;
Anamalayil, Shibu ;
Teo, Boon-Keng Kevin .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (05) :100-111
[9]  
de Crevoisier R, 2016, Cancer Radiother, V20 Suppl, pS200, DOI 10.1016/j.canrad.2016.07.037
[10]   Dosimetric characterization of MRI-only treatment planning for brain tumors in atlas-based pseudo-CT images generated from standard T1-weighted MR images [J].
Demol, Benjamin ;
Boydev, Christine ;
Korhonen, Juha ;
Reynaert, Nick .
MEDICAL PHYSICS, 2016, 43 (12) :6557-6568