Application of the two-dose-rate method for general recombination correction for liquid ionization chambers in continuous beams

被引:22
作者
Andersson, Jonas [1 ]
Tolli, Heikki [1 ]
机构
[1] Umea Univ, Dept Radiat Sci, SE-90185 Umea, Sweden
关键词
COLLECTION EFFICIENCY;
D O I
10.1088/0031-9155/56/2/001
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A method to correct for the general recombination losses for liquid ionization chambers in continuous beams has been developed. The proposed method has been derived from Greening's theory for continuous beams and is based on measuring the signal from a liquid ionization chamber and an air filled monitor ionization chamber at two different dose rates. The method has been tested with two plane parallel liquid ionization chambers in a continuous radiation x-ray beam with a tube voltage of 120 kV and with dose rates between 2 and 13 Gy min(-1). The liquids used as sensitive media in the chambers were isooctane (C(8)H(18)) and tetramethylsilane (Si(CH(3))(4)). The general recombination effect was studied using chamber polarizing voltages of 100, 300, 500, 700 and 900 V for both liquids. The relative standard deviation of the results for the collection efficiency with respect to general recombination was found to be a maximum of 0.7% for isooctane and 2.4% for tetramethylsilane. The results are in excellent agreement with Greening's theory for collection efficiencies over 90%. The measured and corrected signals from the liquid ionization chambers used in this work are in very good agreement with the air filled monitor chamber with respect to signal to dose linearity.
引用
收藏
页码:299 / 314
页数:16
相关论文
共 17 条
[2]   ACCURACY OF APPROXIMATE SOLUTIONS FOR CURRENTS IN A PLANE PARALLEL ION CHAMBER [J].
ARMSTRON.W ;
TATE, PA .
PHYSICS IN MEDICINE AND BIOLOGY, 1965, 10 (02) :229-&
[4]   CURRENT COLLECTION AND IONIC RECOMBINATION IN SMALL CYLINDRICAL IONIZATION CHAMBERS EXPOSED TO PULSED RADIATION [J].
BOAG, JW ;
CURRANT, J .
BRITISH JOURNAL OF RADIOLOGY, 1980, 53 (629) :471-478
[5]   Volume recombination parameter in ionization chambers [J].
Boutillon, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (08) :2061-2072
[6]  
DEALMEIDA CE, 1986, BIPM8612, P1
[7]   SATURATION CHARACTERISTICS OF PARALLEL-PLATE IONIZATION CHAMBERS [J].
GREENING, JR .
PHYSICS IN MEDICINE AND BIOLOGY, 1964, 9 (02) :143-&
[8]  
International Atomic Energy Agency, 2000, IAEA TRS-398
[9]   General collection efficiency for liquid isooctane and tetramethylsilane used as sensitive media in a parallel-plate ionization chamber [J].
Johansson, B ;
Wickman, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (01) :133-145
[10]  
Kara-Michailova E, 1940, P CAMB PHILOS SOC, V36, P101