Development of an extrapolation chamber for the calibration of beta-ray applicators

被引:14
作者
Dias, SK [1 ]
Caldas, LVE [1 ]
机构
[1] IPEN CNEN SP, Sao Paulo, Brazil
关键词
beta dosimetry; calibration; extrapolation chamber;
D O I
10.1109/23.685289
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
An extrapolation chamber was designed and constructed, and its performance was studied in beta radiation fields. The characteristics of this chamber and the measurement procedures used for the calibration of beta-ray applicators are discussed in this work. The surface dose rate of a planar Sr-90 + Y-90 applicator was also obtained using the extrapolation chamber, and the results were compared with the Amersham calibration certificate, An agreement within 2.7% was verified.
引用
收藏
页码:1666 / 1669
页数:4