Fast Concomitant Gradient Field and Field Inhomogeneity Correction for Spiral Cardiac Imaging

被引:13
作者
Cheng, Joseph Y. [1 ]
Santos, Juan M. [1 ,2 ]
Pauly, John M. [1 ]
机构
[1] Stanford Univ, Dept Elect Engn, Magnet Resonance Syst Res Lab, Stanford, CA 94305 USA
[2] HeartVista Inc, Los Altos, CA USA
关键词
concomitant gradient field; field inhomogeneity; spiral; non-stationary convolution; OFF-RESONANCE CORRECTION; MAGNETIC-FIELD; MR-IMAGES; GEOMETRIC DISTORTION; RECONSTRUCTION; ARTIFACTS; PHANTOM; ARRAY;
D O I
10.1002/mrm.22802
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Non-Cartesian imaging provides many advantages in terms of flexibility, functionality, and speed. However, a major drawback to these imaging methods is off-resonance distortion artifacts. These artifacts manifest as blurring in spiral imaging. Common techniques that remove the off-resonance field inhomogeneity distortion effects are not sufficient, because the high order concomitant gradient fields are nontrivial for common imaging conditions, such as imaging 5 cm off isocenter in an 1.5T scanner. Previous correction algorithms are either slow or do not take into account the known effects of concomitant gradient fields along with the field inhomogeneities. To ease the correction, the distortion effects are modeled as a non-stationary convolution problem. In this work, two fast and accurate postgridding algorithms are presented and analyzed. These methods account for both the concomitant field effects and the field inhomogeneities. One algorithm operates in the frequency domain and the other in the spatial domain. To take advantage of their speed and accuracy, the algorithms are applied to a real-time cardiac study and a high-resolution cardiac study. Both of the presented algorithms provide for a practical solution to the off-resonance problem in spiral imaging. Magn Reson Med 66:390-401,2011. (C) 2011 Wiley-Liss, Inc.
引用
收藏
页码:390 / 401
页数:12
相关论文
共 34 条
[1]  
Ahunbay E, 2000, MAGNET RESON MED, V44, P491, DOI 10.1002/1522-2594(200009)44:3<491::AID-MRM22>3.0.CO
[2]  
2-Z
[3]   Characterization, prediction, and correction of geometric distortion in 3 T MR images [J].
Baldwin, Lesley N. ;
Wachowicz, Keith ;
Thomas, Steven D. ;
Rivest, Ryan ;
Fallone, B. Gino .
MEDICAL PHYSICS, 2007, 34 (02) :388-399
[4]   Concomitant gradient terms in phase contrast MR: Analysis and correction [J].
Bernstein, MA ;
Zhou, XHJ ;
Polzin, JA ;
King, KF ;
Ganin, A ;
Pelc, NJ ;
Glover, GH .
MAGNETIC RESONANCE IN MEDICINE, 1998, 39 (02) :300-308
[5]   Combination of signals from array coils using image-based estimation of coil sensitivity profiles [J].
Bydder, M ;
Larkman, DJ ;
Hajnal, JV .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (03) :539-548
[6]   Fast Conjugate Phase Image Reconstruction Based on a Chebyshev Approximation to Correct for B0 Field Inhomogeneity and Concomitant Gradients [J].
Chen, Weitian ;
Sica, Christopher T. ;
Meyer, Craig H. .
MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (05) :1104-1111
[7]   Semiautomatic off-resonance correction in spiral imaging [J].
Chen, Weitian ;
Meyer, Craig H. .
MAGNETIC RESONANCE IN MEDICINE, 2008, 59 (05) :1212-1219
[8]   Fast automatic linear off-resonance correction method for spiral imaging [J].
Chen, Weitian ;
Meyer, Craig H. .
MAGNETIC RESONANCE IN MEDICINE, 2006, 56 (02) :457-462
[9]  
Cheng J Y., 2009, Proceedings of the 17th Annual Meeting of ISMRM, Honolulu, Hawaii, USA, P2844
[10]  
Cheng J Y., 2010, Proceedings of the 18th Annual Meeting of ISMRM, Stockholm, Sweden, P3108